MODS Overview

MODS1 and MODS2 are UV to nearIR Multi-Object Double Spectrgraphs that are mounted at the Direct Gregorian focal stations of the LBT. They were designed and built by The Ohio State University Department of Astronomy in Columbus Ohio.

Each MODS is a low- and medium-resolution spectrograph (R=100-2300) and imager working across the entire 330-1100nm band with a 6x6-arcminute field of view. Multi-object spectroscopy is accomplished using laser-machined focal-plane slit masks fed into the beam by a 24-position mask cassette. A beam selector below the slit carries a dichroic that splits the incoming beam into separate red- and blue-optimized channels (the dichroic cross-over wavelength is 575nm), each with its own collimator, grating, camera and detector, allowing simultaneous operation across the entire CCD band. The beam selector can also direct light into the red or blue channels alone, providing blue-/red-only modes to extend wavelength coverage across the dichroic cross-over wavelength as required. The MODS CCDs are custom-built 3x8K monolithic e2v CCDs; blue-coated standard silicon on the blue channel and extended-red coated 40-micron deep depletion silicon on the red channel.

The following image shows both MODS 1 and 2 mounted on the LBT (looking from the back of LBT). LBT <a class="foswikiCurrentTopicLink" href="/bin/view/Software/MODS">MODS</a> 1 and 2 mounted.jpg

MODS machines and how to login

The following table lists the main machines associated with the MODS system. The usual login account can be obtained from the shared Google drive document "Anonymous Accounts" (https://docs.google.com/spreadsheets/d/1fYW8KX5Wf86jkDvDjfyHWt-hsAxljN9WaeykR24IrDY/edit?ts=5c0eaa87#gid=0).

MODS1 Computers
mods1 Owns the MODS processes (CentOS Linux)
mods1data Stores data (FC3 Linux)
blue ccd DOS machine for CCD
red ccd DOS machine for CCD

MODS2 Computers
mods2 Owns the MODS processes (CentOS Linux)
mods2data Stores data (FC3 Linux)
blue ccd DOS machine for CCD
red ccd DOS machine for CCD

Others
APC Connected power management (192.168.52.38 & 39)
Raritan/KVM Remote access (192.168.139.19)

Note the APC unit is a network connected power management unit (similar to the dataprobe units on the LBC). You can login to the APC unit and then remotely restart or shutdown instruments (see section on PDU #APC_PDU). The MODS CCD machines are running DOS and are usually logged in via the KVM.

MODS Software Overview

The figure below shows the software components and architecture for a single MODS instance (substitute the "#" in the figure with either 1 or 2). Both MODS have identical underlying systems that operate independently. The dashed boxes enclose programs running on particular host computers, with the exception of "LBTO Systems" which is a blanket container for all LBTO subsystems that the MODS data-taking systems interact with.

Interface protocols are shown in the key at the lower left corner. "Ethernet" is a mix of TCP/IP and UDP/IP depending on the system. Optical fibers are either Fiber Ethernet (the one into the IUB) or dedicated data I/O fibers for the MODS science CCDs and the AGw Guide and WFS cameras. "Virtual C&C" is a catch-all indicating "other" virtual command-and-control protocols, for example the modsTCS to LBT IIF processes communicate via the Zero-C ICE interface, whereas the LBT GCS and MODS AGw server and Az Cam servers communicate using a TCP/IP socket interfaces.

The MODS GUI program is shown running on an obsN workstation, but note that this is just a window into the "modsUI" program running on the mods1/mods2 linux host, with the GUI portion instantiated on the obsN workstation via X tunneling through an ssh session between the observer account on obsN and the relevant mods# host computer.

In binocular MODS mode, the acqMODS and execMODS programs are run via a wrapper script that makes sure that the appropriate binocular context flag is set, and which permits both sides to be running either the same acquisition or observation script (called the "identical twin mode") or different acq/obs scripts for each MODS (called the "fraternal twin mode").

LBT <a class="foswikiCurrentTopicLink" href="/bin/view/Software/MODS">MODS</a> 1 and 2 mounted.jpg

ISIS = Integrated Science Instrument Server, see http://www.astronomy.ohio-state.edu/~pogge/Software/ISIS/server/. The ISIS server is the message passing server that mediates all interprocess communications between all MODS instrument subsystems via the IMPv2.5 messaging protocol. It also records all command and control messaging traffic in runtime communications log stored on this machine.

The following is a list of abbreviated named used in the figure. We do not give the details of them in this document as they are more related to hardware.
  • IEB: Instrument Electronics Box, mainly for power connection and management.
  • IUB: Instrument Utility Box.
  • LLB: Lamp/Laser Box.
  • HEB: CCD Head Electronics Box.

Software Builds

Currently the only MODS software built by LBTO is the library libagwutils.a used by GCS. In the repository, the top-level makefile ( Makefile.build ) builds only the API and app directories - the app directory builds the libagwutils.a used by GCS.

Log in to a 64-bit CentOS machine in Tucson (for example: tcs@tcs-test ), export/checkout to the version-number you are building, make and install.
For instance, for the build version 1.2.3:
ssh tcs@tcs-test
cd modsagw
git clone git@github.com:LBTO/mods.git
cd mods/agw
edit Makefile.build to set the VERSION, SUBLEVEL, PATCHLEVEL numbers for what you want (top of file suppose you change them to 1.2.3)
make
make install   (copies to /lbt/tcs_devel)

ssh tcs@tcs1.mountain.lbto.org
cd /lbt/tcs_devel
scp -pr tcs@tcs-test.tucson.lbto.org:/lbt/tcs_devel/mods-1.2.3 .
  • Note that the GCS links in the static library, so the TCS file .build/environment has to be modified for the path of the latest MODS library.

Software/System Use and Management

Power Up and Down

Normally support astronomers will do the power up. In case you are wondering about the steps, you can follow the instructions at: https://wiki.lbto.org/Instrumentation/MODSStartUp.

If an emergency (or unanticipated) power outage is impending, the automatic UPS script will take care of putting the instrument into a safe configuration and powering off the mechanisms and computers (see https://wiki.lbto.org/Instrumentation/InstrumentsShutdownForPowerOutage). Since this procedure does not home all of the mechanisms, a full reset of the MODS system is required when the power is returned - home all mechanisms, then send to nominal configuration positions.

If however you want to manually power down the MODS system for some reason, then you need to follow the instructions at: https://wiki.lbto.org/Instrumentation/MODSSupportUtils#Power_Outage_45_modsShutdown. Note that the scripts and operations mentioned in the instructions are located/performed on any obsN machine (e.g., obs2, obs3 etc.).

Also see notes from Olga on post-shutdown startup https://wiki.lbto.org/Instrumentation/MODSPostShutdownStartup.

Start/stop status checking scripts

On the obs3 machine (note: it must be obs3), there are scripts that periodically query the MODS status and send out email alerts if status problems are detected. These scripts are located at the directory: /home/telescope/bin/. There are two scripts there emailMODSErrors.sh and emailMODS2Errors.sh. They are run by the cron daemon. So to disable them (e.g., after MODS have been turned off or taken off from the telescope), just login to obs3 as user telescope, and do "crontab -e" from the command line and then comment out the lines corresponding to these jobs (by putting "#" in front of that line). To re-enable them, do "crontab -e" from the command line and then uncomment the lines.

APC PDU

MODS hardware are connected to PDU (Power Distribution Unit) manufactured by APC. They are similar to the dataprobe units used by the LBC. The PDUs have internal addresses: http://192.168.52.38 and http://192.168.52.39 (one for each MODS). You can access these PDUs in two ways:
  • Use telnet to login to a PDU (user name is mods and password is the same). Then you can issue commands there (much the same as the dataprobe units used by LBC).
  • You can also use your web browser to login to the PDUs (use the above address, though you need to be in the LBTO intranet). Use mods to login and then you will be presented with an interface screen similar to the figure shown below. You can then click to operate the PDU.

MODS PDU web interface

Software Tools and Utilities

All the essential MODS related processes are running on the MODS machines (mods{1,2}, mods{1,2}data, and the DOS CCD control machines). See the above notes on power up (#Power_Up_and_Down) to know these processes and services. Regular users and observers will however use the observation machines (obs{2,3,4} for example) to launch GUI and execute scripts, which will communicate with the MODS processes and services on the MODS machines. The follow list is a summary of the common commands and scripts.

  • The main user interface is to launch mods1 start gui on an obs machine (e.g. obs3). This will launch the MODS control panel GUI. Refer to the OSU MODS manual (http://www.astronomy.ohio-state.edu/MODS/Manuals/MODSManual.pdf).
  • On any obs machine, the directory /home/MODSeng/modsScripts/Support/ contains various support scripts that an observer will likely use. The OSU MODS script reference (http://www.astronomy.ohio-state.edu/MODS/Manuals/MODSScripts.pdf) is a detailed manual for everything about MODS scripts (in particular acqMODS and execMODS).
  • modsView is a visualization and guide star selection tool. Refer to http://www.astronomy.ohio-state.edu/MODS/ObsTools/modsView/ for help.
  • modsDisplay for raw data display (see the MODS manual for details).
  • modsAlign is an interactive mask alignment tool (see the MODS manual for details).
  • mods1 status and mods2 status, this will bring up a summary of the status of the major MODS services. You can either run the command on an obs machine, or you can also run it on the mods{1,2}data machine. The following screen shows an example of what you'd expect to see.
lbto@obs3:3 % mods1 status
MODS1 Status at 2018 Dec 19 23:24:29

Instrument Server:
  Service       State           Owner
  --------------------------------------
  IE            Running         mods        (IE = Instrument Electronics Command & Control)
  AGw           Running         mods        (Acquisition, Guiding, and Wavefront Sensing)
  redIMCS       Running         mods        (IMCS = Image Motion Compensation System)
  blueIMCS      Running         mods
  modsenv       Running         mods        (Environmental Monitor)
  lbttcs        Running         mods        (TCS IIF Interface)
  modsUI        Stopped                     (MODS Control Panel GUI)
  --------------------------------------

Data Server:
  Service       State           Owner
  --------------------------------------
  isis          Running         mods        (Integrated Science Instrument Server)
  Red Caliban   Running         mods        (Caliban Data Transfering Service)
  Blue Caliban  Running         mods
  --------------------------------------
  • isisCmd sends command string to the ISIS client (detailed description and documents missing, mainly seen used to power cycle instruments).
  • execISIS executes an ISIS script (detailed description and documents missing).

A quick reference for a usual night operation of MODS can be found at: https://wiki.lbto.org/PartnerObserving/MODSQuickRefWiki.

Data Management

MODS uses the "Caliban" data handling program developed at OSU to transfer raw FITS images from the CCD control computers to the /lhome/data data storage drive on host machines mods{1,2}data. The OSU team has produced a help page for the "Caliban" program (http://www.astronomy.ohio-state.edu/~prospero/Caliban/).

The details of the data transfer occur in three steps:

Step 1: CCD Controller to MODS Data Server

A MODS CCD image is first read off the CCD into memory on the CCD control computer (e.g., M1.RC for the MODS1 Red Channel, M1.BC for the MODS1 Blue Channel). From memory it is written onto a transfer disk shared between the DOS-based CCD control computer and the MODS data server machine (the Linux workstation mods1data for MODS1, and mods2data for MODS2). Once on the transfer disk, a dedicated instance of the Caliban data-transfer daemon running on the data server copies it from the transfer disk and writes it onto the /lhome/data staging disk in FITS format. Data from the red and blue channels are written to the same data server staging disk. Once written, its FITS header is checked and augmented with additional archive and engineering keywords, its header is scanned and the running data log in /Logs/ is updated , and the image is ready to be copied to the LBTO data archive.

The step of copying the raw image from the transfer disk onto the data server’s staging disk is between 4 and 12 seconds, depending on the size of the image (this is a known bottleneck in the system and is on the list of improvement with future hardware upgrades). Post-processing of the image header and logging take less than a second. This final transfer-and-process step is asynchronous: if a sequence of images is being acquired, the next image in the sequence will be started as soon as the last file is written to the inter-machine transfer disk, and the final transfer-and-process step will occur while the next image starts.

However there is an IMPORTANT NOTE: It is at the data-transfer step between the CCD control computer and the MODS data server that the data-transfer queue can stall. The symptom is that the LastFile counter will fail to increment and images will stop appearing in the modsDisp windows, despite the fact that the NextFile counter is incrementing. If the difference between LastFile and NextFile grows larger, the data-transfer queue has stalled. If you notice that the transfer queue has stalled, type the command fitsflush in the Command window on the MODS dashboard GUI. This should restart (“flush”) the FITS data transfer queue, and you’ll start seeing the LastImage counter increment and images appearing on the disk.

Step 2: Data Server to the LBTO Archive Staging Disk (/newdata)

After a MODS FITS image MODS data server’s staging disk has been processed for archiving, it is then copied across the network to the LBTO “new data” archive staging disk named, appropriately enough, /newdata. The transfer from the MODS data server to the LBTO /newdata disk usually takes around 1 second for unbinned 8K×3K images. Data written to /newdata are read-only, but may be copied onto the observing workstation disks for further analysis. The /newdata disk is where new MODS images first become available to the observers on the observing workstations. This disk is where modsDisp will watch for new raw-images to display, and where modsAlign will get thru-slit and field images for mask alignment.

Step 3: /newdata to LBTO archive (/Repository and beyond)

Each image that appears on /newdata will be immediately ingested by the LBTO data archive software, triggering a sequence of events that typically take no more than a couple minutes to complete. These steps include (in approximately this order):
  1. Image FITS header keywords are read and the archive database is updated.
  2. The image is copied to /Repository/ for access on subsequent days.
  3. The image is gzip compressed and filed on the /archive disk (no user access).
  4. The image copied to the Tucson archive machine.
  5. The Tucson archive repeats steps 1 through 4 above.
  6. The Tucson archive sends copies as needed to the Germany & Italy archives.

Within a few minutes multiple copies of each image will be distributed across a network of RAID6 data arrays in the observatory archive machines. The images will stay on the /newdata disk until noon the following day when they are deleted to make room for the next night's data. Images copied to the /Repository disk will be available (read-only) for a months, stored in subdirectories organized by date. For example, data from UTC 2018 Dec 24 will be stored in /Repository/20181224/. Guider and WFS images taken during that same night will be stored in a GCS subdirectory of this same folder. The /Repository disk is kept organized by the archive software; older data are automatically deleted to make room for the newest data. Both /newdata and /Repository are available (read-only) to observers logged into the observing workstations at the LBT.

Fits Headers

The PARTNER, PROPID, and PI_NAME FITS header keywords are used to assign ownership of the data, primarily by the PARTNER keyword. The PARTNER is defined by the observatory to be one or more of these reserved values:

Partner Code
LBT Observatory Staff LBTO
LBTB (Germany) Partner Observing LBTB
INAF (Italy) Partner Observing INAF
Ohio State & Research Corp Partners OSURC
Arizona Partner Observing AZ

For regular science operations. Additional PARTNER IDs (e.g., COMMISSIONING and CALIBRATION) are used for technical observing and special applications. The PROPID and PI_NAME are used differently by different partners and are user definable (or at least defined within a partner group, for example the OSURC partner block has internal conventions for how to assign PROPID and PI_NAME values for its observing queue). If multiple partners are sharing data, PARTNER can be a comma-separated list, for example PARTNER AZ,LBTB with no spaces before or after the comma. This will allow observers from either AZ or LBTB to access the data from the archive. Note that CALIBRATION partner ID is a special flag that should be used for all calibration data (biases, flats, comparison lamps, etc.) that allows all partners access to calibration images.

Log Files

Log Machine Description
/home/mods/Logs/mmc.log mods{1,2} communication between devices
/home/mods/Logs/agw.log mods{1,2} all communication between AGw interface and LBT GCS instance
/home/mods/Logs/Env/mods1.YYYYMMDD.log
/home/mods/Logs/Env/mods2.YYYYMMDD.log
mods{1,2} environmental sensor data
/Logs/Caliban/cb_blue.log
/Logs/Caliban/cb_red.log
mods{1,2}data  
/Logs/Data/YYYYMMDD.log mods{1,2}data image header data
/Logs/Env mods{1,2}data old MODS1 env logs
/Logs/CCD/ mods{1,2}data CCD environmental monitor logs
/Logs/ISIS/isis.YYYYMMDD mods{1,2}data also communication between devices

Config Files

Most of the config files are located on the mods1 and mods2 machine under the directory /home/mods/Config/. The config files for mods data server are located under the directory /lhome/dts/Config/ on machine mods1data and mods2data ("dts" is a user that stands for "data-taking system").

Troubleshooting

So far most of the troubleshooting procedures are for the support astronomers, operators, and the observers. Refer to: As usual, many of the first steps are to try to power cycle the various instruments (via the isisCmd command line).

ISorted descending Attachment Action Size Date Who Comment
LBT_MODS1and2.jpgjpg LBT_MODS1and2.jpg manage 59 K 07 Dec 2018 - 22:23 YangZhang MODS 1 and 2 mounted on LBT
This topic: Software > MODS
Topic revision: 17 Jan 2019, YangZhang
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback