28-December on-sky testing of version 2.6

Had some on-sky testing of version 2.6 on the night of Friday, 28-December-2012. We wanted to check out:
  • ability to change the spiral search overlap size from the GUI
  • automatic offsetting during acquisition
  • sort of starlist in the GUI
  • focuser functionality

There's some junk in the starlist - why does the sun show up? "Sol" with mag of -27

we couldn't acquire stars with 2.6 initially. The first measurements we saw were 02:15UTC with version 2.5. Went back and forth at this point between versions to figure out if it was related to the new version. It appeared to be, even though I had disabled the offsetting.

12/29/12 02:08:13.725 - Build 2.6 DIMM Server is starting
12/29/2012 02:10:06.661 - Slew to RA: 2.064978, DEC: 42.329722,   dist from LBT:7.21.
12/29/2012 02:10:41.800 - Start measurement.
12/29/2012 02:10:47.021 - Warning: No stars found in field of view. Please try again or move on.
12/29/2012 02:11:35.009 - Start measurement.
12/29/2012 02:11:40.734 - Warning: No stars found in field of view. Please try again or move on.
12/29/2012 02:11:52.790 - Start measurement.
12/29/2012 02:11:57.328 - Warning: No stars found in field of view. Please try again or move on.
....
12/29/12 02:12:51.333 - Build 2.5 DIMM Server is starting
12/29/2012 02:13:27.216 - Slew to RA: 1.162166, DEC: 35.620556.
12/29/2012 02:13:33.870 - Warning: Not measuring - stop camera thread.
12/29/2012 02:14:32.472 - Start measurement.
12/29/2012 02:14:36.948 - DimmServiceI::measureDimm: Two stars found. Setting the region-of-interest.
12/29/2012 02:15:18.893 -       Seeing value: 0.788 at el 85.33 az 311.00,   LBT el 77.22  az 0.58
12/29/2012 02:15:23.147 -       Seeing value: 0.802 at el 85.32 az 310.88,   LBT el 77.22  az 0.52
12/29/2012 02:15:28.210 -       Seeing value: 0.802 at el 85.30 az 310.75,   LBT el 77.22  az 0.45
12/29/2012 02:15:33.155 -       Seeing value: 0.760 at el 85.29 az 310.63,   LBT el 77.22  az 0.39
12/29/2012 02:15:38.217 -       Seeing value: 0.837 at el 85.28 az 310.50,   LBT el 77.22  az 0.32
12/29/2012 02:15:38.972 - Stop measurement.

12/29/12 02:16:16.927 - Build 2.6 DIMM Server is starting
12/29/2012 02:16:46.527 - Slew to RA: 2.064978, DEC: 42.329722,   dist from LBT:7.21.
12/29/2012 02:17:23.513 - Start measurement.
12/29/2012 02:17:28.452 - Warning: No stars found in field of view. Please try again or move on.
12/29/2012 02:17:42.432 - Start measurement.
12/29/2012 02:17:47.173 - Warning: No stars found in field of view. Please try again or move on.
12/29/2012 02:18:24.631 - Warning: Not measuring - stop camera thread.
12/29/2012 02:18:32.105 - Start measurement.
12/29/2012 02:18:40.871 - Warning: No stars found in field of view. Please try again or move on.

12/29/12 02:19:12.895 - Build 2.5 DIMM Server is starting
12/29/2012 02:19:35.041 - Slew to RA: 1.162166, DEC: 35.620556.
12/29/2012 02:20:38.304 - Start measurement.
12/29/2012 02:20:42.483 - DimmServiceI::measureDimm: Two stars found. Setting the region-of-interest.
12/29/2012 02:21:34.237 -       Seeing value: 0.895 at el 84.28 az 303.03,   LBT el 77.18  az 355.61
12/29/2012 02:21:39.597 -       Seeing value: 0.872 at el 84.26 az 302.95,   LBT el 77.18  az 355.54
12/29/2012 02:21:44.187 -       Seeing value: 0.855 at el 84.25 az 302.87,   LBT el 77.18  az 355.48
12/29/2012 02:21:49.230 -       Seeing value: 0.899 at el 84.23 az 302.79,   LBT el 77.18  az 355.41
12/29/2012 02:21:54.170 -       Seeing value: 0.863 at el 84.22 az 302.70,   LBT el 77.18  az 355.35
....
12/29/2012 02:48:50.288 -       Seeing value: 1.060 at el 79.07 az 289.14,   LBT el 75.75  az 335.95

Then, I thought maybe I had not turned OFF the offsetting correctly and that maybe that was causing the problem. Ran 2.6 again, and we were able to acquire the star:

12/29/12 03:14:53.383 - Build 2.6 DIMM Server is starting
12/29/2012 03:15:27.329 - Slew to RA: 22.716703, DEC: 30.221111,   dist from LBT:12.71.
12/29/2012 03:16:07.040 - Start measurement.
12/29/2012 03:16:11.483 - DimmServiceI::measureDimm: Two stars found. Setting the region-of-interest.
12/29/2012 03:16:48.797 -       Seeing value: 1.321 at el 42.41 az 282.84,   LBT el 53.71  az 291.75
12/29/2012 03:16:53.094 -       Seeing value: 1.294 at el 42.39 az 282.85,   LBT el 53.69  az 291.76
12/29/2012 03:16:58.106 -       Seeing value: 1.358 at el 42.38 az 282.85,   LBT el 53.67  az 291.76
......
12/29/2012 03:42:20.084 -       Seeing value: 0.929 at el 37.22 az 285.20,   LBT el 48.74  az 292.68
12/29/2012 03:42:22.570 - Stop measurement.

After awhile, we could not get it to acquire again -- we could see the stars in the full-frame image, in the middle, but ... sometimes we'd get it, like this case, but then not see 2 stars after centering:
12/29/2012 04:25:41.361 - Error: Stars out of region-of-interest...stop measuring, stop tracking
12/29/2012 04:25:43.347 - Reset ROI to full frame x: 1388 y: 1024
12/29/2012 04:26:00.429 - MKSDriver::startMove: encoders -1775248, -3359330
12/29/2012 04:26:03.869 - MKSDriver::endMove: encoders -1771594, -3359330
12/29/2012 04:26:04.521 - DimmServiceI::endMoveMount: Mount has stopped moving...
12/29/2012 04:26:17.824 - Start measurement.
12/29/2012 04:26:21.584 - DimmServiceI::measureDimm: Two stars found. Setting the region-of-interest.
12/29/2012 04:26:21.613 - centerStar Current position: ud 19.357033 lr -4.883332
12/29/2012 04:26:21.613 - centerStar star X 601.69 Y 110.36  Distance [deg]: x  -0.008726 y -0.037966
12/29/2012 04:26:21.613 - centerStar New position [deg]: ud 19.348306 lr -4.921299
12/29/2012 04:26:22.240 - MKSDriver::startSlewRaw: startSlewRaw motor counts ra:-55.311752  dec:161.258966
12/29/2012 04:26:22.297 - MKSDriver::IsHomed2: encoders -1771177, -3359381
12/29/2012 04:26:22.773 - MKSDriver::endSlewRaw: new base ra: -1769987  dec: -3359557
12/29/2012 04:26:26.551 - Warning: Didn't find two stars after centerStar, not measuring

And other times, we'd not get it when we "start measuring". Even though we had handpaddled it in perfectly.

we saw some instances where we would handpaddle the star to the mark, and then it would scoot off immediately. It looked like it could be related to the offsetting. But, it shouldn't have been trying to do an offset

Focuser

We tested the move-in/move-out/stop functionality of the focuser. When we used the slider, we ended up getting an exception. We never got the connection to the focuser back.
In fact, we worked several hours on Saturday with Elliott and Kevin and never got it back until the telescope moved to zenith. ???
Here are the excepted log messages:

12/29/2012 02:05:35.018 - Stop Focuser.
12/29/2012 02:05:35.018 - DimmFocuserJmi: Stop
12/29/2012 02:05:35.018 - DimmFocuserJmi::sendCommand: Sending command to focuser...
12/29/2012 02:05:35.139 - DimmFocuserJmi::sendCommand: ok
12/29/2012 02:05:35.149 - DimmFocuserJmi: position 520
12/29/2012 02:05:35.158 - DimmFocuserJmi: status()  at_max=0  at_zero=0  motor_error=0  overrun_error=0  frame_error=0
12/29/2012 02:05:51.037 - Position Focuser 361.
12/29/2012 02:05:51.037 - DimmFocuserJmi: goTo 361
12/29/2012 02:05:51.047 - DimmFocuserJmi::writeValue: Processing...
12/29/2012 02:05:56.346 - Stop Focuser.
12/29/2012 02:05:56.346 - DimmFocuserJmi: Stop
12/29/2012 02:05:56.346 - DimmFocuserJmi::sendCommand: Sending command to focuser...
12/29/2012 02:07:51.052 - Error: DimmFocuserJmi::writeValue, CommException caught
12/29/2012 02:07:51.052 - Error: positionFocuser, Exception caught - CommSerial - Connection timeout. 

Changed the code (created a new version of 2.5) to not allow setting a numerical value or using the slider. Only can use the move-in/out/stop buttons.

Spiral still doesn't get the star. The percentage value is how big you move:
 udSize=width*scale*searchSize/3600 

So, moves about 1041 pixels in width when using 75% and about 768 in height.
I need to map the spiral search moves to handpaddle moves. Spiral search does the following moves:
      UD            LR 
     -udSize        0
      0            -lrSize
      udSize        0
      2*udSize      0
      0             lrSize
      0             2*lrSize
     -udSize        0
     -2*udSize      0
     -3*udSize      0
      0            -lrSize
      0            -2*lrSize
      0            -3*lrSize
.... 56 times 

We tried multiple sizes of the search overlap. 0.6 0.5 0.4 0.3

Guiding

We were able to handpaddle the mount and see that the guiding would keep it close to center.