LBTO Techincal/D/Service Observing Night
OSA: D.G.Huerta
ISp: J.Morris
Manager: J.Urban
ISA: B.Rothberg, D.Thompson, J.Hill (Tucson)
================================================
UT 11:26 (Nov20) - scope configured for LUCI-2 at start
UT 12:48 - updated PCS with MODS2 0.6 degree rotation correction
UT 01:10 - moving to HD210501 - test of LUCI-2 throughput
UT 01:18 - doing a pointing check as guidestar acquisition failed twice
UT 01:19 - resending preset to HD210501 again
UT 01:40 - Error - voltage didn't stabilize at 2.2 micron, script aborted. - last image was 0012
UT 01:43 - Aborted, clicked skip to the 2.2micron @K 210 and started - it did an extra -20" dither
UT 01:44 - Aborted with voltage error again
UT 01:44 - manually putting in 2.2micron
UT 01:45 - error committing instrument setup with 2.2 micron @210 grating
UT 01:47 - again, another error commiting instrument setup - trying to set to 1.65 then 2.21, then 2.20
UT 01:49 -didn't like 2.21, trying 2.19
UT 01:50 - initializing grating, trying 2.20 - failed again
UT 01:52 - trying 2.22 - failed again
UT 01:55 - 2.3 worked, trying 2.25
UT 01:55 - 2.25 didn't work, trying 2.29, worked, trying 2.26
UT 01:57 - 2.26 failed, trying 2.27, worked
UT 01:58 - trying 2.24, failed trying again
UT 02:01 - i'm continuing script at 2.27 microns - note that everything is -20" off from where it should be in script - and we skipped one dither position in the K observing
UT 02:25 - Failed to set grating 210 to 2.2microns
UT 02:31 - still can't get anything at < 2.27 microns to stabilize
UT 02:33 - aborting script - reloading version with 2.27 micron for field stop, and redoing field stop part of script
UT 02:34 - sending preset again - going to line up again on LS2" slit, then skip in script down to field stop section and redo it
UT 02:47 - restarted N30 field stop portion
UT 03:03 - LUCI2 testing completed for throughput
========================================
UT 03:04 - LUCI-2 imaging of M31 - RMWagner
UT 03:33 - Completed M31 imaging field - moving to sky position
UT 03:34 - missed the guidestar - aborting script
UT 03:36 - resending M31-sky script
UT 04:03 - Script finished - some variable seeing towards the end
========================================
UT 04:07 - Moving to Stone A3 field for LUCI2 transform tests
UT 04:10 - N3.75 Camera, Pointing Mask, Br Gamma filter
UT 04:33 - some issues in getting transform running - seeing still highly variable
UT 04:37 - starting again
UT 04:39 - #0069 blank (blinds in)
UT 04:43 - starting again
UT 04:48 - 0070 first image
GS 8 = 0072
GS 2 = 0073 (binary?)
GS 5 = 0074
GS20 = star too far out
UT 05:01 - relodaing with start=4
GS 24 = stopped again, too far of a move
UT 05:07 - restarting again, start=3
GS 10 = 0075
GS 11 = 0076
GS 9 = 0077
GS 8 = stopped on this too
UT 05:17 - restarting again, start=7 - clouds visible across southern sky
GS 5 = 0078
GS 10 = 0079
GS 11 = 0080
GS9 = No guidestar found in box - could be too faint
UT 05:26 - restarting again, start=11
GS 4 = err = Resmsg[2] = error: isPosAccessible: couldn't move probe to position (out of patrol field) (SFP) (x,y): 134.88, -26.3228 mm.
UT 05:29 - restarting again, start=12
GS 0 = 0081
UT 05:39 - moving to
StoneD to do similar testing with field we've done before - using Stone_D_new_Oct27.list
GS0 = 0082
offset 30 = 0083
GS4 = 0084 PA=0
GS12 = 0085
GS23 = 0086
GS13 = 0087
GS22 = 0088
GS5 = 0089
GS2 = 0090
GS11 = 0091
GS29 = 0092
GS24 = 0093
GS30 = 0094
GS25 = 0095
GS4 = 0096 PA=15
GS12 = 0097
GS23 = 0098 --- seeing 1" or less
GS13 = 0099
GS22 = 0100
GS17 = 0101
GS5 = 0102
GS11 = 0103
GS29 = 0104
GS24 = 0105
GS30 = 0106 -seeing bloom
GS4 = 0107 PA=30
GS12 = 0108
GS23 = 0109
GS13 = 0110
GS22 = 0111
GS5 = 0112
GS11 = 0113
GS29 = 0114
GS24 = 0115
GS30 = failed preset - this happened Oct 27 too, so its ok
================================================
UT 06:56 - switching to MODS2 - putting LUCI2 in safe config
UT 07:13 - mods2r.201121.0002 - test exposure
UT 07:15 - put in ND1.5 filter - 1sec ~9300 counts
UT 07:18 - 4 sec ~22k peak
UT 07:21 - D.Miller remote test image -0005
UT 07:22 - starting collection
UT 07:27 - 006,007 guideprobe in way
UT 07:35 - exited on GS12, restarted with adjustments made to moving guideprobe
UT 07:43 - GS23 failed
UT 07:44 - restarting with start=3
UT 07:52 - changed exposure time to 3seconds to prevent saturation - GS5 Pa=180 is last 4sec exposure
UT 07:57 - failed on GS29 PA=180, restarting with restart=9
UT 08:00 - failed on GS24, restart=10
UT 08:02 - failed on GS30, restart=11
UT 08:04 - failed on GS25,restart=12
UT 08:07 - seeing subarcsecond
UT 08:08 - failed on GS12 PA=195 - changing exptime to 2sec
UT 08:10 - failed on GS23, restart=15
UT 08:22 -failed on GS29 PA=195, restart=21
UT 08:26 - failed on GS24 PA=195, restart=22
UT 08:28 - failed on GS30 PA=195, restart=23
UT 08:34 - failed on GS23 PA=195, restart = 26
UT 08:40 - seeing degraded a bit
UT 08:44 - failed on GS11 PA=210, restart=30
UT 08:50 - having trouble collimating - got an error after 7 iteratoins - continuing
UT 08:52 - failed on GS29 PA=210, restart=31
UT 08:55 - failed on GS24 PA=210, restart=32
UT 09:00 - failed on GS30 and GS20 PA=210
=================================
UT 09:01 - moving to
V959Mon (LST 05:30 script) - science field for RMWagner-OSURC
UT 09:04 - seeing 1" in guider
UT 09:11 - not clear which is the actual target - is it the bright star I centered on, or the slightly fainter blob to bottom left of it...
UT 09:15 - exposure started - will know if its correct target after first spectra reads out
UT 09:29 - i'm assuming this is it as I see bright emission lines
UT 09:50 - completed target - moving to
G191B2b spec-phot std
=================================
UT 09:59 - centering up g191b2b a bit in slit +-0.6" x
UT 10:00 - starting g191b2b spectra
UT 10:11 - finished std
==================================
UT 10:12 - moving to V380 Orionis - Rothberg target
UT 10:23 - need to take a shorter exp field image
UT 10:26 - starting spectra (r:0044-46, b:0008-0010)
UT 10:32 - had three IMCS red lock failures, elevation = 44
UT 10:38 - again no lock-on for red
UT 10:47 - ran through the procedure to power down and up lasers manually, still no joy, red imcs will not lock
UT 11:09 - went back to G191b2b to test, worked ok, came back to V380, acq worked ok, now trying spectroscopy again
UT 11:13 - MODS2 still having issues with RED IMCS lock, switching back to LUCI2 for end of night
UT 11:15 - IT 5767 for MODS2 Red IMCS lock
UT 11:33 - moving to WAS49
UT 11:48 - acq complete - taking K, then H G210 spectra
UT 12:00 - noticed flexure compensation appears to be off in GUI, but script says its on - header says Flexure comp is off - appears to have turned off after acquisition completed (on at 119, off 120-125)
UT 12:05 - flexure compensation went back ON after switch to H-band
UT 12:10 - bottom half of script vanished after I had put in a pause and removed it.
UT 12:14 - guiding paused on -60" offset, but I had already done this offset earlier - I suspect the "correction" when script vanished, might have done two dither moves instead of one
UT 12:22 - going to run through script again, manually adding back in +60" +X dither to get back on track
UT 12:23 - re-running K, flexure compensation turned itself off
UT 12:35 - removed a pause and script vanished again and can't continue. -Note flexure compensation is back on for last two K-G210 exposures in sequence (136-137)
UT 12:39 - 0.5-0.7" in guider
UT 12:57 - Moving to telluric
UT 01:16 - finished, closing enclosure then will do cals
UT 01:23 - LUCI2 - cals 0.5" slit K
UT 01:27 - commit error for cal setup - had to restart scheduler
UT 01:28- running slitflats 0.8" MODS2
UT 01:44 - LUCI2 instrument manager frozen?
UT 01:48 - Called Dave T. - unable to restart software or do anything, however GEIRS works
UT 01:49 - Running slitless pixel flats MODS2
UT 02:02 - Dave T. restarted instrument manager -all seems ok - running 0.5" G210 @K2.27 cals
UT 02:09 - MODS2 arc lamps
UT 02:19 - MODS2 8k biases
UT 02:27 - starting LUCI2 0.5"G210H cals
UT 02:28 - putting mods2 to sleep
UT 02:36 - doing h flats and just the argon arcs
UT 02:41 - putting LUCI2 into safe config, N375, blindsx2, mirror, flex off, cal unit out, leaving 0.5" slit in FPU
-- %USERSIG{BarryRothberg - 2015-11-20}%