Soft Power Cycle

If things are in a bad state (Readout issues, CCD Timeouts, or software in a funky state) re uploading the program is like a soft power cycle.
  • Close the Pepsi Interface
  • From a terminal on pepsi@alpha type
    pepsi
Often necessary after a power bump

Reboot PEPSI AZCam

  1. Click "Maintenance" in the lower left of the PEPSI LBT GUI.
  2. In the Mainentance Popup, select the Radio button for the side of interest where you want to reboot the azcam (SX FPU or DX FPU)
    Maintenance Popup:
    Maintenance.png
  3. Select the Reboot buttom. This will power cycle the Azcam. Access to the server will be unavailable during this.
  4. Restart GCS after AZCam returns (vnc for AZcam will be reachable again).
Alternately
  1. Click "Maintenance" in the lower left of the PEPSI LBT GUI.
  2. In the Mainentance Popup, select the Radio button for the side of interest where you want to reboot the azcam (SX FPU or DX FPU)
    Maintenance Popup.
  3. Select the VNC button to bring up a VNC window to the PEPSI AZCam.
  4. Through the start menu -> Restart
  5. Verify that you see in the terminal
    • /usr/bin/net rpc shutdown -r -I 192.168.37 (or 38) -u azcam%azcam
  6. Restart GCS after AZCam restart completes

Fringes in Cross Profile of Guider

If fringing is noted on the cross profile of the guider, the binning should be increased. Bin1 should only be used for the brightest of targets.

The guiding binning is adjusted from the PEPSI LBT GUI. Adjusting the binning on one side will automatically increase the binning for the guiding on the other telescope (they are linked together).
PEPSILBTGUI_changeBinning.png

Updates to binning are outlined in terminal window where the PEPSI interface was started. Users should watch there for any issues. If guiding is unable to resume after a pause when adjusting the binning then shifts in binning may fail.

Users cannot simultaneously preset and change binning. When changing binning, the guiding is paused, changed, and then resumed.

-- JenniferPower - 08 Sep 2021

*