UT 20150331 LUCI2 AO Nighttime Commissioning

Observer LUCI-Team: JHeidt, WSeifert, APramskiy, DThompson (LBTO)
Observer AO-Team: SEsposito, GAgapito, EPinna, APuglisi
Telescope Operator: SAllenson
LBTO Support: TSargent

Plan for the night

- Since LUCi2/AGw are in the lab, no AOCom today

- Arcetri people work on the AGw issue(s)

- Calibration of the new gratings

- If weather is good, LBTO will use the telescope

Summary

New gratings G40/G210 calibrated

Preparations

LUCI2/AGw were brought to the lab today

Details

All times are UT
Files are luci2.20150331.NNNN.fits

01:43 Sunset, sky mostly covered, T = +6deg, low wind, Hum = 32%

We are still running on generator power

  • Grating calibration

The G210 grating was calibrated using the N1.8 camera+0.5" slit,
centered at xpix=1028, the G40 grating was calibrated using the N3.75
camera+0.5"slit centered at xpix=1034. Walter put an Argon-lamp in front of LUCI2.
The table below gives the summary. The numbers for the Ne/Xe-lamps
are interpolated (but should be good).

Arc-lamp Lambda Voltage Image-number/comment
G210z
Ar 0.8427 4.712 #36
Ne 0.8471 4.541 interpol
Ar 0.8569 4.350 #35
Xe 0.8822 3.596 interpol
  0.8955 3.24 interpol
  0.9048 2.997 interpol
Ar 0.9125 2.735 #31
  0.9227 2.391 #32
  0.9660 0.792 #33
  0.9787 0.302 #34
Xe 0.9923 -0.243 interpol
  1.0032 -0.563 interpol
  1.0124 -0.97 interpol
G210J
Ar 1.1516 2.50 #30
  1.1684 2.00 #29
Ne 1.1770 1.680 interpol
Ar 1.2116 0.675 #20
  1.2143 0.592 #21
  1.2347 -0.05  
  1.2406 -0.245 #23
  1.2491 -0.52 #24
  1.2706 -1.19 #25
  1.2806 -1.50  
  1.2937 -1.90 #26
  1.3012 -2.110 #27
  1.3242 -2.759 #28
G210H
Ar 1.5085 3.031 #19
  1.5199 2.792 #18
Ne 1.5251 2.650 interpol
Ar 1.5346 2.478 #17
Xe 1.5425 2.30 interpol
Ar 1.5904 1.252 #16
  1.5994 1.042 #15
Xe 1.6058 0.896 interpol
Ar 1.6184 0.60 #9
  1.6441 0.00 #10
  1.6524 -0.211 #11
  1.6745 -0.72 #12
  1.6945 -1.181 #13
Ne 1.7119 -1.675 interpol
Xe 1.7330 -2.05 interpol
Ar 1.7475 -2.324 #14
G210K
Ar 2.0622 4.980  
  2.0992 4.675 #5
Ne 2.1047 4.635 interpol
Ar 2.1339 4.369 #4
  2.1540 4.160 #6
Ne 2.1714 3.985 interpol
Ar 2.2083 3.622 #3
Ne 2.2580 3.070 interpol
Ar 2.3096 2.293 #7
Ne 2.3519 1.96 interpol
Ar 2.3827 1.260 #8
G40z
Ar 0.8427 0.065 #71
Ne 0.8471 0.035 interpol
Ar 0.8569 -0.026 #70
Xe 0.8822 -0.11 interpol
  0.8955 -0.165 interpol
  0.9048 -0.250 interpol
Ar 0.9125 -0.346 #67
  0.9227 -0.409 #66
  0.9660 -0.660 #68
  0.9787 -0.738 #69
Xe 0.9923 -0.880 interpol
  1.0032 -0.960 interpol
  1.0124 -1.040 interpol
G40J
Ar 1.1516 -0.388 #65
  1.1684 -0.473 #64
Ne 1.1770 -0.520 interpol
Ar 1.2116 -0.676 #62
  1.2143 -0.690 #63
  1.2347 -0.783 #61
  1.2406 -0.811 #63
  1.2491 -0.850 #60
  1.2706 -0.957 #59
  1.2806 -1.007 #58
  1.2937 -1.067 #57
  1.3012 -1.10 #55
  1.3242 -1.199 #56
G40H
Ar 1.5085 -0.30 #44
  1.5199 -0.339 #45
Ne 1.5251 -0.356 interpol
Ar 1.5346 -0.392 #46
Xe 1.5425 -0.432 interpol
Ar 1.5904 -0.592 #47
  1.5994 -0.622 #48
Xe 1.6058 -0.640 interpol
Ar 1.6184 -0.688 #49
  1.6441 -0.777 #50
  1.6524 -0.808 #51
  1.6745 -0.888 #52
  1.6945 -0.958 #53
Ne 1.7119 -1.103 interpol
Xe 1.7330 -1.107 interpol
Ar 1.7475 -1.135 #54
G40K
Ar 2.0622 0.139 #37
  2.0992 0.055 #38
Ne 2.1047 0.040 interpol
Ar 2.1339 -0.021 #39
  2.1540 -0.061 #40
Ne 2.1714 -0.105 interpol
Ar 2.2083 -0.185  
Ne 2.2580 -0.219 interpol
Ar 2.3096 -0.423 #42
Ne 2.3519 -0.518 interpol
Ar 2.3827 -0.59 #43

11:30 Sky mostly clear, Steve opens, Olga & Barry did some observations with LUCI1
during the last hour of the night.

13:00 Telescope closed
Topic revision: r5 - 01 Apr 2015, JochenHeidt