UT 20150325 LUCI2 AO Daytime Commissioning
Observer LUCI-Team: WSeifert
Observer AO-Team: SEsposito, GAgapito, EPinna, APuglisi
Telescope Operator:
LBTO Support: JPower
Plan for the day
- NCPA - tests
Summary
- pupil alignment done
- Image scale N30-camera found to be 0.015 +- 0.001 arcsec/pixel
- Problem with AO electroniccs: power supply for grabber CCD47 seems to be dead
Details
All times are UT
Files are luci2.20150325.NNNN.fits
- N30 T_Cam=73.5K
#0001 neglim, #0002 middle, #0003 poslim (best)
- Pupil alignment
#0004 FM1 M1: (y) 0steps M2 (x) -3000steps
-3000 only necessary to center in most illuminated area, otherwise NO movement necessary
FM4 to re-center field: M1 -250steps M2 +1600steps
#0006 N30 after re-align @ poslim
- Scale N30
from #0006 15.00 milli-arcsec/pix +-0.01, N = 30.0
- Input AO: N3.75 center = 1037 / 1057
- Preparation
init hexapod & center ARGOS cal unit in N375
FoV
focus before (January 2015: 52.8mm) now ~55mm
#0008 N30
FeII /cl subwindow 0.17s
-> image looks VERY similar to that in January!
- Shift & add cube images & automatics NCPA scan
Failure: power supply for grabber CCD47 dead (?)
-> no more working
-- %USERSIG{JochenHeidt - 2015-03-27}%