UT 20131105 LUCI2 Daytime Commissioning

Observer: DThompson (Tucson)
Telescope Operator: GBechetti
LBTO Support:

Summary

A solid set of flexure data with g210/K-filter could be derived.
Except for a problem with the ReadOut Manager at the beginning
the system was stable.

Details

20131105 LUCI2 Flexure Comp. Calibrations
Have to do them by hand, not scripted.

Starting READ2 GUI from LMC hung the LMC panel!
Open LMC on second desktop
Stop/start cheduler
Restart LUCI2 from top of LMC

Data in: /dataB/readout2/DATA/20131104/
Filenames: luci2.20131104.NNNN.fits (incorrect date)

0001 1x5.0s Dark - spectro sieve + blind filters

Calib Unit IN
N1.80 G210@2.20um clear+K filters

0002 test 1x5.0s lamp off
0003 test 1x5.0s AR lamp on
0004 test 1x10.0s AR lamp on
0005 test 1x10.0s lamp off
- use 10.0s

0006 1x10.0s lamps off for dark

N1.80 210_zJHK Ar lamp FC calib.
ROT EL90 EL55 EL20
-90 7 44 81
-60 8 45 82,83
-30 9 46 84
000 10 47 85
030 11 48 86
060 12 49 87
090 13 50 88
120 14 51 89
150 15 52 90
180 16 53 91
210 17 54 92
240 18 55 93
270 19 56 94
300 20 57 95
330 21 58 96
360 22 59 97
390 23 60 98
420 24 61 99
450 25 62 100
420 26 63 101
390 27 64 102
360 28 65 103
330 29 66 104
300 30 67 105
270 31 68 106
240 32 69 107
210 33 70 108
180 34 71 109
150 35 72 110
120 36 73 111
090 37 74 112
060 38 75 113
030 39 76 114
000 40 77 115
-30 41 78 116
-60 42 79 117
-90 43 80 118

Ar lamp off

0119 lamp off

Calib unit out
mirror + N3.75 + blind filters

0120 blind filters in (dark)