W start up. Fault.
Bayside stage Z down.

Secondary start up. Ok.
Start OSS. ok
StartAOS. ok
Start PSFL ok.

Hexapod in position using the PSFGUI left. ok
The focus global parameter has to be changed.

Hexapod position :[0.79 -1.72 -0.180 -193.4 37.2 0]

PUPIL CALIBRATION PROCEDURE USING INTERNAL SOURCE AND MOVING THE CAMERA LENS. "PUPIL CHECK OPTION"
A NEW PUPIL IS SAVED AS:20090000-105150.

BIN 1 _111601 1000HZ HIGH FLUX, 2L/d, rec:

no disturbance. Simone SR:91%
_______________________-----------------------------------------------------------------------------------------------------------------
mag:11.5 BIN 2 625Hz, 3/D


_113035,_113200 ,_113622 dark:20091201_113358

SimoneSR: aprox. 54%
dist:0.8_ov625 syncWFS

---------------------------------------------------------------------------------------------------------------------------------------------------------
mag:13.5 BIN 2 400Hz, 3/D.


_115239,_115919,_120111,_120301. dark:_20091201_115616.
dist:0.8_ov800

Simone SR:aprox. 32%

----------------------------------------------------------------------------------------------------------------------------------------------------------
mag:15.5 BIN 4 200Hz, 6/D.


_122256._122601,_122750 dark:_123156.
dist:0.8_ov800

Simone SR:aprox. aprox 13%

--------------------------------------------------------------------------------------------------------------------------------------------------------------------
mag:16.5 BIN 4 100Hz, 6/D.


_123701,_123851,_125413,dark:_123156.
dist:0.8_ov800 Simone SR:aprox. aprox11 %

Applying new integrator: IntegratorforgetFactor0.95.

_125610,_125809 .dark:(0.5sec,150frames)_130203.


simone SR: aprox8%.

Applying new integrator: IntegratorForgetFactor0.85

_131355,_131542 dark:_131932.

TAKEN MORE IMAGES CHECK RED BOOK


mag:15.5 BIN 4 200Hz, 6/D. FF0.95


_153834,_154351, 154611,dark:(0.5sec,150frames)_154239
dist:0.8_ov800 OVERSAMPLING.

Simone SR:aprox.5 %.

Reflatened Secondary. Acquire new CCD38 dark.

_155325,_155521, simone SR:aprox7.5%
rip the secondary
_160204,_160350,_160547 dark:(0.5sec,150frames)_160917 Simone SR: aprox 6.2%

reflatened.
_161208(wrong),_161815 ,_162236, _162630, _162851 (offloading applied),_163218(tilt about 2um applied)
_163814(offloading applied), _164217(tilt about 3um applied),_164542(offloading applied)
dark:(0.5sec,150frames)_163603
simone SR: aprox

Dist:1.5 arcsec.
_165156(possible wrongSR:0.8%),
RIP THE SECONDARY

_170335 seeing limited.

CHECK RED BOOK more images taken after.

-------------------------------------------------------------------------------------------------------------

MAG:13.5 BIN 2 400Hz 3l/D pureIntegrator.

dist:0.6"
_173050,_173130,_173544
dark(50000us,150 frames)_173323
simone SR: aprox:48%

dist:1.5"
_173943,_174030, _174105 (irtc image 50000um,150 frames)
irtc:0.5 sec,150 frames) _174649
_174259(not irtc file),_175135,_175418

Simone SR: aprox 5%.


RUNNING AOS. TEST USING THE AOS FAILED
PROBLEMS WITH THE USE OF THE AOS. AOARBITRATOR COMPILED AND THE MANY MODULES OF THE TCS.

"AOSYSTEM mode: ACE-AO
freq : 1000.000000
nModes : -1219430068
binning : 1
TTmod : 2.000000
Fw1Pos : 0
Fw2Pos : 0
gain : 0.500000
pupils : 20090925-113454
base : KL
rec : 20091119_124749
decimation: 2
freq : 1000.000000
nModes : -1219430068
binning : 1
TTmod : 2.000000
Fw1Pos : 0
Fw2Pos : 0
gain : 0.500000
pupils : 20090925-113454
base : KL
rec : 20091119_124749
decimation: 2