SW test.

Settaggi filtri/lampda per le varie magnitudini:

Mag Lamp Thorlabs FW1 da tabella Count attesi sul ccd47 secondo tabella:
8.5 55 3 0 5000
9.5 32 3 0 2000
10.5 19 3 0 5000
11.5 49 1 0 1800
12.5 29 1 3 7000
13.5 19 1 3 2500
14.5 40 6 4 1700
15.5 23 6 4 600
16.5 17 6 4 250

Thorlabs Z pos
6 44.2
3 44.6
1 44.3

AO Arbitrator command Issued from Expected result Test result from AOSGUI with IRC Parameter validation
Power on Adsec Adopt Control GUI Adsec is powered on and initialized with default flat shape OK    
Power on WFS Adopt Control GUI WFS is powered on and initialized with default IRTC setup. OK    
Power off Adsec Adopt Control GUI Adsec mirror is put in "rest" position and safely powered of. OK    
Power off WFS Adopt Control GUI WFS is powered off. OK    
Set ENGINEERING mode Adopt Control GUI AOArbitrator in service state "ENGINEERING" OK    
Set STANDALONE mode Adopt Control GUI AOArbitrator in service state "STANDALONE" OK    
Start observation AOS GUI AOArbitrator in service state "OBSERVATION" and SeeingLmtd state. OK    
Terminate observation AOS GUI AOArbitrator in service state "STANDALONE" OK    
PresetFlatAO irc

Adsec selects the specified flat shape or, if no particular shape is specified, selects the default flat shape.
AOArbitrator in SeeingLmtd state

OK OK  
PresetAO irc

Reference star information is used to do a preliminary selection of AO parameters
(reference star is loaded with irc SetReference command)
AO parameters are shown in the AOS GUI.
WFS is prepared for source acquisition. A setup is loaded dependent on instrumen
AOArbitrator in ReadyToAcquire state.

OK. PCS bug in focal plane coordinates. OK SetReference and PresetAO compute coordinates correctly.  
CheckRefAO irc WFS checks reference star position in the technical camera and offsets are computed.
Offsets are shown in the AOS GUI
AOArbitrator remains in ReadyToAcquire state.
OK OK  
AcquireRefAO irc WFS acquires the reference star using the acquisiton camera
AO parameters are refined using the measured magnitude and seeing
AOS GUI shows thenew AO parameters.
Adsec is prepared for AO loop.
AOArbitrator in RefAcquired state.
OK OK  
ModifyAO irc AO parameters modified as specified in the command parameters.
WFS and Adsec are re-initialized as needed
AOS GUI shows the new AO parameters.
AOArbitrator in RefAcquired state.
Partially tested. OK - iTime must be changed to Freq.  
StartAO irc AO loop is closed
AOArbitrator in LoopClosed state
OK OK  
PauseAO irc AO loop is paused
AOArbitrator in LoopPaused state
OK OK  
ResumeAO irc Light levels are checked in the WFS camera.
AO loop is resumed
AOArbitrator in LoopClosed state
OK OK  
StopAO irc AO loop is opened.
AOArbitrator in SeeingLmtd state.
OK OK  
OffsetXY irc WFS applies XY offset.
If in closed loop, PSF on technical camera and/or IRTC shows the same displacement.
AOArbitrator remains in current state (LoopClosed or LoopSuspended).
OK OK OK
OffsetZ irc

WFS applies Z offset
If in closed loop, PSF on technical camera and/or IRTC shows the same displacement.
AOArbitrator remains in current state ( LoopClosed or LoopSuspended).

OK OK OK