CMDDISTURBANCE - INTMAT ANALYSE
cMD_DIST_110351 INTMAT ANALYSE: _110820 Ampl file. _113324

AMPLITUDE FILE 672_0.1
KL
400 MODES.
cMD_DIST_111013 INTMAT ANALYSE: _111125 Ampl file. _113543

AMPLITUDE FILE 672_0.3
KL
400 MODES.
cMD_DIST_111302 INTMAT ANALYSE: _111337 Ampl file. _113633.

AMPLITUDE FILE 672_0.7
KL
400 MODES.
cMD_DIST_ 123303 INTMAT ANALYSE: _123359 Ampl file. _123607
AMPLITUDE FILE 672_0.5
KL
400 MODES.

ANALYSE INTMAT.
Generated a new amplitude file for BIN1 2L/D. 20011119_amp_opti_Bin1_mod2.fits.

GENERATED A NEW DISTURBANCE WITH THE NEW AMPLITUDE FILE.

DISTURBOIM <_114902>

type = <push-pull ..>
PP param 5,1
Modes = KL / .... >
#Modes = <400>
Amp = <2001119_amp_opti_bin1_mod2>
Note = <..............>

INTMATACQ <_115140> THIS SHOWS A WAFFLE ("GRATUIA")
Pup = <tracking#>
Dist = <tracking#>
Iteration = <4>
Loop speed = <400 Hz>
bin <1>
Modul =<2l/D at 400+HZ>
Rec.Close loop
<rec_2001005_111109>

Gain vector close loop = <gain0.05>

REC_2001119_115343 BIN 1 2 l/d 400 MODES.

RIP THE SECONDARY, LOAd itw default shape, correction of the tilòt with the hexapod.

INTMATACQ <_121009> THIS SHOWS A WAFFLE ("GRATUIA") less than the rec before.
Pup = <tracking#>
Dist = <_114902>
Iteration = <4>
Loop speed = <400 Hz>
bin <1>
Modul =<2l/D at 400+HZ>
Rec.Close loop
<rec_2001005_111109>

REC_20091119_122100 BIN 1 .....2L/D ..... 400 MODES. AMP.FILE: 2001119_amp_opti_bin1_mod2

test of the rec good up to 0.4

new aplitude file:

20011119_amp_opti_v2_Bin1_mod2.fits.

new disturbance with the new amplitude file:

CmdDist _124514



INTMATACQ <_124606> THIS SHOWS A WAFFLE ("GRATUIA") less than the rec before.
Pup = <tracking#>
Dist = <_124514>
Iteration = <4>
Loop speed = <400 Hz>
bin <1>
Modul =<2l/D at 400+HZ>
Rec.Close loop
<rec_2001005_111109>

REC_124749 BIN 1 ... 2l/d.....400 MODES. WITH OPTIMIZED FILE:

20011119_amp_opti_v2_Bin1_mod2.fits

test good up to 0.4, start to oscilaatre at 0.5

rec running at 400 Hz 200 phot. shows a good performances.
rec riunning at 1000Hz, 200 phot. shows a bad performances at gain 0.4 mode oscillation.

START BASELINE POINTS.

BIN1, 400 MODES, 1000Hz 2L/D.
REC_124749.

_134457, problems with the number of frames in the IRTC, with 500 frames .
_135317, _135504.


lamp: high flux.
No dist.
IR.25000us. nframes:300
Gain:0.3
--------------------------------------------------------------------------------------------------
_140907 (fern_SR:74%)
_141243(fern_SR:77%

Mag:8.5.
amp: 200phot/subap.
0.8_ovfreq_1000
IR.25000us. nframes:300 +dark(300 frames)
Gain:0.3
-----------------------------------------------------------------------------------------------------
_141815 (not all frame saved)

Mag:. 9.5
amp: 85phot/subap.
0.8_ovfreq_1000
IR.100000us. nframes:300 +dark(300 frames)
Gain:0.3.
Nota: system stuck during the close loop, ccd39viwer stop (possible problem with the bcu38?).

FAULT:

Nota: system stuck during the close loop, ccd39viwer stop (possible problem with the bcu38?).

BIN2 REC_20091005_175922 3l/d 153 MODES....

_ 182355(.
_

Freq:800
Mag:.
amp: 166phot/subap.
0.8_ovfreq_1000
IR.300000us. nframes:150 +dark(150 frames)
Gain:1.
--------------------------------------------------------------------------------------------
_190519
_190632(fer_SR:55%, simone_SR:63%)

Freq:800
Mag:.
amp: 166phot/subap.
0.8_ovfreq_1000
IR.100000us. nframes:150 +dark(150 frames _190847)
Gain:1.

**ANOTHER REC COMPUTED with an amplitude file 672_0.3. -> REC_20091119_125735 ...BIN1 ....2L/D,..... 400MODES

test rec with an osillation of the low modes at 0.4.

NOTE TO: Ottimizzazione ampiezze per acquisizione intmat

[15.20.26] Alfio Puglisi: 1) generi 10 disturbi con le ampiezze 672_0.1.fits, 672_0.2.fits eccetera
[15.20.38] Alfio Puglisi: 2) le acquisisci tutte
[15.21.04] Alfio Puglisi: 3) le analizzi con la checkbox "only check saturation", che salva i dati della saturazione e non fa l'svd.
[15.22.04] Alfio Puglisi: 4) prendi il file $ADOPT_SOURCE/idl/wfs_lib/elab_saturation.pro, lo modifichi mettendoci dentro i tracking number delle matrici analizzate, e lo esegui
[15.22.51] Alfio Puglisi: 5) la procedura si ferma su uno stop finale e hai un bel vettore "amp" con le ampiezze ottimizzate. Devi salvarlo con un writefits e ri-generare disturbo e intmat a partire da quello
[15.22.57] Alfio Puglisi: buona fortuna....

-- JuanCarlosGuerra - 19 Nov 2009