AdSec672 laboratory tests after boards replacement (session II)

8:40 unit is powered on, after verification of tap water temperature and dew point (the unit has been powered on for one minute without cooling, in order to read Housekeeper value of dewpoint). then the unit has been poowered off and the cooling water is injected into the pipes

power_on: Housekeeper fails reading from BCU. this is the same problem detected yesterday sometimes after poweron. adam_sys_reset, then everything is fine.

load program and set_flat_ao with no problems

series of 10 iterations

at 4th DSP fail

after power on the

RunaBriguglio - 2009-10-23
Topic revision: r2 - 26 Oct 2009, MarcoXompero