You are here: Foswiki>FLAO Web>TowerLab>TowerTest20090921 (22 Sep 2009, EnricoPinna)Edit Attach
Enrico, Alfio e Simone

Flowerpot PIC problems and substitution with spare board

Flowerpot PIC does not work at first startup. Works at second startup, then resets by itself after a few minutes and does not come back. The board suffered a glycole spill some days ago and has been working erratically since then.

Flowerpot PIC board removed. At first inspection, some of the PIC contacts seem slightly damaged and a soldered wire to a capacitor is displaced when touched. Wire re-soldered and board remounted (electronic technician says that the contacts probably work anyway). Works for a minute or two and then stops responding to serial communication.

Spare board mounted.
This board did not have originally the soldered wire (reason unknown), and it was added before mounting it on the W unit.

Firmware seems identical. Board works for two hours straight apart from a few timeouts on commands (the Pic controller automatically repeats the command, which then works). Declared OK for now.

FOCUSING CCD47

AGW#2 appoggiato sul pozzo della torre.
Findstar funziona se si mediano nframes>=2

Fuocheggiamo la pira con l'autofocus a modulazione 1V
FW1 = 600-1000
FW2 = 800-40
L'autofocus del w trova Z=50.61

Ora fuocheggiamo il 47 prima a occhio
a 52.0 appare una psf ellittica allungata orizzontalmente di roundness 0.45 (findstar)
a 53.0 la barrettina si orienta a 90 perche' sfuocata dall'altra parte.7

L'astigmatismo che si vede sul 47 sembra essere dovuto ai filtri della FW1 perche' cambiano intensita' e segno cambiando fra dicroico 600-1000 e 700-1000 e 4%.

Usiamol'autofocus47.py per acquisire le immaigni ma l'analisi del finstar sembra non troppo chiara per ora e preferiamo analizzare le immagini con il nostro programma di fit gauss
psf_fwhm.pro

Le immagini sono in home/labot/testw2/measurements/autoFocus47/20090921_144832/ zpos_#.fits (posizione reale della z nell'header)

Piccolo upgrade della psf_fwhm.pro
Centro specificabile e plot attraferso il picco

Creata psf_focus.pro che usa psf_fwhm.pro per stimare il miglior fuoco. Fa il plot di picco e FWHM fittati dalla psf_FWHM.pro in funzione della Z.

Trova la posizione z=52.6 che ha miglior picco e FWHM

La differenza e' 2.0mm all F15 spostandosi verso la fioriera per mettere a fuoco il 47. Dunque il 47 va portato verso la fioriera di 18mm.MEMENTO fattore 9 in fuoco fra F15 e F45.
Ora la basetta del CCD47 e' tornata a filo con il bordo della board.

Rimisuriamo con autofocus47.py,immagini in 20090921_161045
si analizza con psf_focus. Questo dice che il miglior fuoco e' a 50.65 invece di 50.60 trovato dall'autofocus CCD39.

50um sulla slitta z corrispondono a 50/(8*15^2) = 0.027um in fronte d'onda, quindi ci fermiamo qui.

Inoltre anche l'incertezza dell'autofocus del 39 ha una riproducibilita' comparabile ai 50um.

Ci prendiamo un PSF47 con la sorgente DL
20090921_162449
Fiber = sm58
Lamp = interna
FW1 4%
FW2 800-40
Frames mediati = 30
Note: sorgente con LPF600

Stesse condizioni con 6um di sisplacement
20090921_163315

RIMONTATO IL W al tubone

CLOOP 20090921_182812 CL no dist gain basso
CCD47PSF = stesso
Bin = 1
LoopHz = 400
Pup = 20090918_183032
Dist =off
Gain = 0.2
Rec = 20090918_184605
SN = zero
FW1 = 4%
FW2 = 800-40
Phot = 1800 phot/subap
Nota = <................>

CLOOP 20090921_183125 OL senza disturbo
CCD47PSF = stesso
Bin = 1
LoopHz = 400
Pup = 20090918_183032
Dist =off
Gain = 0.0 (OL)
Rec = 20090918_184605
SN = zero
FW1 = 4%
FW2 = 800-40
Phot = 1800 phot/subap
Nota = <................>

CLOOP 20090921_183320 OL con disturbo
CCD47PSF = stesso
Bin = 1
LoopHz = 400
Pup = 20090918_183032
Dist =80pix + Over Sampling
Gain = 0.0 (OL)
Rec = 20090918_184605
SN = zero
FW1 = 4%
FW2 = 800-40
Phot = 1800 phot/subap
Nota = <................>

CLOOP 20090921_183703 CL con disturbo
CCD47PSF = stesso
Bin = 1
LoopHz = 400
Pup = 20090918_183032
Dist =80pix + Over Sampling
Gain = 0.7
Rec = 20090918_184605
SN = zero
FW1 = 4%
FW2 = 800-40
Phot = 1800 phot/subap
Nota = Sul 47 la PSF appare a volte buona a volte completamente allungata dalle vibrazioni probabilmente introdotte dal vento sulla torre

Proviamo a prendere le SN con la pennina direttamente sul W

Ripristiniamo il Boardsetup = Final setup
ricentriamo con autocenter e autoficus poi loop di camera

Rirotatore centrato a 174deg


SN 20090921_184511 sn su fibra
PUP=20090921_183032
BIN = 1
FW1 = 4%

CLOOP 20090921_185004 CL no dist gain basso
CCD47PSF = stesso
Bin = 1
LoopHz = 400
Pup = 20090918_183032
Dist =NO
Gain = 0.1
Rec = 20090918_184605
SN = 20090921_184511 quelle con la pennina diretta su W
FW1 = 4%
FW2 = 800-40
Phot = 1800 phot/subap
Nota =

CLOOP 20090921_185112 CL no dist gain a 0.2
CCD47PSF = stesso
Bin = 1
LoopHz = 400
Pup = 20090918_183032
Dist =NO
Gain = 0.2
Rec = 20090918_184605
SN = 20090921_184511 quelle con la pennina diretta su W
FW1 = 4%
FW2 = 800-40
Phot = 1800 phot/subap
Nota =

-- AlfioPuglisi - 2009-09-21