Simone,alfio,Lorenzo,Juan Carlos

OPTICAL ALIGNMENT OF THE W2

CCD39 looks like in lower position than we expected, for this reason it was necessary to lift a bit the ccd39 base.
Pupil rerotator seems to work well and the system is aligned to the rotation axe.

CCD47 focused a psf images taken. Saved in:
/home/labot/testw2/settember/
psf_47_f800.40.fits
psf_47_f800.40_focus.fits
FW1:600-1000
FW2: 800-40

SOLVED PROBLEM OF THE CAMERA LENS SENSE. the problem was the connector of the camare lens sense, one of the cable was off, after solding the cable and plug the connector again the device moved correctly in both axes.

Start the pupil dimension alignments, the system is close to the final alignment. CCD 39 in focus. At the end of the alignment procedure, the ccd39 viewer started to shut, this was done several times. We decided to left it and to check with Alfio the fault.