Hexapod

[-524.151, 2156.92, -146.337, 157.452, -41.9523, -0.0782142]

WFS

frame rate 25Hz

Mirror reflattening

reflattened mirror using the optical interaction matrix data in meas/flat/2009_08_10

print, opt_int_mat_reduce(600,/MULT,NMEAS=5)

peak force: -0.44N in the inner ring

residual rms (2*surface): 36nm rms (tip-tilt,focus,coma removed).

intrerferogram (average of 100 frames, 1fr/s) saved in: D:\4d\data\2009_08_27_flattened600modes_36nmRMS

Forces are too high!

Flattening with less modes to reduce the forces.

400 modes: peak force: -0.36N, interferogram saved> 2009_08_27_flattened400modes_40nnRMS

18:00 system shut down due to unfavourable weather conditions

20:00 system restart

several failures during startup (set&flat procedure). in the end the shell was set using the GUI set_flat command, then the ICC was reset using

print, set_icc(/reset)

reflattened mirror using the optical interaction matrix data in meas/flat/2009_08_10

print, opt_int_mat_reduce(300,/MULT,NMEAS=5)

peak force=0.3N, surface error=44nm RMS (interf image saved: 2009_08_27_flattened300modes_44nmRMS.h5)

That will be the flattening configuration for the following measurements

Basename directory name: ~/Zcopy/2009_08_27_...

21:00 system shut down due to the a possible failure in the W lext box communication

--