You are here: Foswiki>FLAO Web>TowerLab>TowerTest20090806 (06 Aug 2009, EnricoPinna)Edit Attach
RON test W#1

Enrico JCarlos

W#1 in camera pulita (sporca)
NO COOLING!!!!

Piastrina CCD47 di PVC

****************************************

RON TEST 1
Basetta = metallica tradizionale
Massa = treccia dalla testa al frame AGW ma fissata con nastro e non viti + cavo massa nero testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 63.84 - RON (e-): 31.92 (estimated)
Speed: 890 kpix/sec - RON (ADU): 50.67 - RON (e-): 25.34 (estimated)
Speed: 400 kpix/sec - RON (ADU): 12.27 - RON (e-): 6.13 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.18 - RON (e-): 6.09 (estimated)
******************************************

RON TEST 2
Basetta = metallica tradizionale
Massa = togliamo la treccia dalla testa + cavo massa nero testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 79.24 - RON (e-): 39.62 (estimated)
Speed: 890 kpix/sec - RON (ADU): 61.58 - RON (e-): 30.79 (estimated)
Speed: 400 kpix/sec - RON (ADU): 40.56 - RON (e-): 20.28 (estimated)
Speed: 150 kpix/sec - RON (ADU): 61.92 - RON (e-): 30.96 (estimated)

****************************************

RON TEST 3
Basetta = testa sospesa sopra la boardtenuta in mano con 2cm di gommapiuma fra mano e testaMassa =nessuna massa + cavo massa nero testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 71.82 - RON (e-): 35.91 (estimated)
Speed: 890 kpix/sec - RON (ADU): 58.97 - RON (e-): 29.48 (estimated)
Speed: 400 kpix/sec - RON (ADU): 13.78 - RON (e-): 6.89 (estimated)
Speed: 150 kpix/sec - RON (ADU): 14.73 - RON (e-): 7.37 (estimated)

****************************************
RON TEST 4
Basetta = testa sospesa sopra la boardtenuta in mano con 2cm di gommapiuma fra mano e testaMassa =treccia testa-AGW frame come in test 1 + cavo massa nero testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 76.59 - RON (e-): 38.29 (estimated)
Speed: 890 kpix/sec - RON (ADU): 59.98 - RON (e-): 29.99 (estimated)
Speed: 400 kpix/sec - RON (ADU): 13.46 - RON (e-): 6.73 (estimated)
Speed: 150 kpix/sec - RON (ADU): 13.51 - RON (e-): 6.75 (estimated)

****************************************
SPENTO IL SISTEMA PER TROPPO CLADO POI SI RIPARTE
****************************************
RON TEST 5
Basetta = in vetronite
testaMassa = treccia testa-AGW frame RIMOSSA + cavo massa nero testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 61.53 - RON (e-): 30.76 (estimated)
Speed: 890 kpix/sec - RON (ADU): 51.82 - RON (e-): 25.91 (estimated)
Speed: 400 kpix/sec - RON (ADU): 12.05 - RON (e-): 6.02 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.42 - RON (e-): 5.71 (estimated)
****************************************
RON TEST 6 Basetta = in vetronite
testaMassa = treccia testa-AGW frame connessa + cavo massa nero testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 76.94 - RON (e-): 38.47 (estimated)
Speed: 890 kpix/sec - RON (ADU): 59.29 - RON (e-): 29.64 (estimated)
Speed: 400 kpix/sec - RON (ADU): 12.89 - RON (e-): 6.44 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.36 - RON (e-): 5.68 (estimated)
****************************************
RON TEST 7 Basetta = in vetronite
testaMassa = treccia testa-AGW frame connessa TOLTO CAVO NERO testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 73.40 - RON (e-): 36.70 (estimated)
Speed: 890 kpix/sec - RON (ADU): 59.41 - RON (e-): 29.71 (estimated)
Speed: 400 kpix/sec - RON (ADU): 12.64 - RON (e-): 6.32 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.55 - RON (e-): 5.77 (estimated)

****************************************
RON TEST 8
Basetta = in vetronite
testaMassa = treccia testa-AGW frame connessa CORTA e TOLTO CAVO NERO testa-vuoto
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 85.46 - RON (e-): 42.73 (estimated)
Speed: 890 kpix/sec - RON (ADU): 61.00 - RON (e-): 30.50 (estimated)
Speed: 400 kpix/sec - RON (ADU): 12.94 - RON (e-): 6.47 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.24 - RON (e-): 6.12 (estimated)

-- EnricoPinna - 2009-08-06
Topic revision: r1 - 06 Aug 2009, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback