Default system startup and more:
  • Launch "D" (dafult init script) on 4Sight GUI
  • Launch InterferometerCtrl in a shell
  • Setup the WAVE GENERATOR for triggering 10.76 Hz
  • Setup the "Camera Setting" in 4Sight GUI selecting TTL0 as "External Triggering"
  • Close and Start again 4Sight SW

IDL terminal command set final release for chopping test procedure:
  • print, fsm_set_flat(MS_CONF_FILE='mirror_set_conf_offload_70um.txt') ; Set the shell with correct conf file
  • print, hexmoveto([-591, 1836, -49, 179., -56., -0]) ; align the hexapod
  • print, idl_4d_init() ; init 4d communication throught MsgD-RTDB
  • print, get_m4d('dummy', img, mask) ; get a sample image
  • fit = pupilfit(mask, guess=[500,500,450, 0.01], /circle, /disp) ;fit the pupil (in this date fit=[496.149, 466.964, 459.007, -0.103177])
  • print, hextrack(/APPLY, PUPIL=fit) ; align the hexapod (more iteration can be done)
  • print, test_chopping_position(0, /DELTA) ; set the preshaper and accumulator settings and the sage skip settings
  • print, fsm_load_shape('/home/labot/idl/adsec672a/meas/flat/2009_07_30/chop_a.sav')
  • print, hexmoveby([0,0,0,-10,0,0])
  • print, hextrack(/APPLY, PUPIL=fit)
  • print, test_continuos_chop(HIST=hist)

For reducing chop data:
  • bb = read_opd(type='4d', 'Zcopy/chop2/hdf5/chop_0006.h5', bad = mask)
  • name = string(reverse(indgen(1500)), FORMAT='("chop_", I4.4, ".h5")')
  • for i=0, 699 do data[*,*,i] = read_opd(type='4d', 'Zcopy/chop2/hdf5/'+name[i])
  • data = fltarr(496, 500, 700)
  • fit = pupilfit(mask, guess=[250,250,220, 0.01], /circle, /disp) ;
  • .r tilt_psd

-- MarcoXompero - 2009-07-31
Topic revision: r3 - 31 Jul 2009, MarcoXompero
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.