You are here: Foswiki>FLAO Web>TowerLab>TowerTest20090709 (10 Jul 2009, EnricoPinna)Edit Attach
PSF1.
directory:adsec_data/20090709/

OPEN LOOP. high flux.
tt:2.6 volt
CCD39:400.
REC_ 142233.
ccd49: (avg300).

OPtical loop diagnostics...

NOTA: disturbance file has an ro different, measurements of the open loop image with disturbance on, the ro measured is 0.47, so the original file has to change the disturbance to have a r0 of 0.8.
Analysis: open loop 65 pixel... which means a 0.47 arcsec (65*13/1800).

PSF2.
OPEN LOOP. high flux.
tt:2.6 volt
CCD39:1000. 600 ph per subaperture.
REC
142233.
ccd49: (avg300).
disturbance ON: modified to have 0.8 arcasec.
MIRROR IN



CALIBRAZIONE BIN2

Pupille: 20090709-181434
Camera lens: Y=0.0mm, X=-32.9mm

Loop params:
BIN=2
FREQ=625
Mod: 3 lambda/D

DISTURBO: 20090710_100155

nmodes = 200
Frames= 5
cycles=2
amp file: pp_351_mod4ok.fits

ACQ: 20090710_100556

1x4000 frames.

INTMAT: 20090710_101752

verifica ampiezze: sono tutte un po' basse di un fattore 2 circa.

NUOVO DISTURBO: 20090710_103508

nmodes = 200
Frames = 5
cycles =2
amp file: pp_200_bin2_625hz_mod3

INTMAT ACQ: 20090710_103637
OPTICAL LOOP DIAGNOSTIC SAVED: data_20090709_150940

PSF3.
CLOSE LOOP. high flux.
tt:2.6 volt
CCD39:1000. 600 ph per subaperture.
REC
142233.
ccd49: (avg300).
GAIN:0.6
disturbance ON: modified to have 0.8 arcasec.
FW1:600
FW2:850-10

OPTICAL LOOP DIAGNOSTIC SAVED: data_20090709_151855
NOTA: the image of the psf doen't appear.

PSF4.
CLOSE LOOP. high flux.
tt:0.67 volt
CCD39:196.... 600 ph per subaperture.
REC
142233.
ccd49: (avg300).
GAIN:0.6
disturbance ON: modified to have 0.8 arcasec.
FW1:600
FW2:850-10


NOTA: the image of the psf doen't appear.

PSF5.
CLOSE LOOP. high flux.
tt:0.67 volt
CCD39:196Hz.... 2000 ph per subaperture.
REC
142233.
ccd49: (avg300).
GAIN:0.6
disturbance OFF: modified to have 0.8 arcasec.
FW1:600
FW2:850-10


NOTA: the secondary skipping, the lamp off, the images analysis give a statistical: 1.8 1.8 4.27 pixel,.. peak/energy:0.048
RIP the secondary. loosing 30'

PSF6......SLOPE NULL WITH DEFOCUS ..
CLOSE LOOP. high flux. LAMP flower pot WS
tt:0.67 volt
CCD39:196Hz.... 1700 ph per subaperture.
REC
142233.
ccd49: 0.4Hz (avg50).
GAIN:0.5
disturbance ON: modified to have 0.8 arcasec.
FW1:600
FW2:850-10.

Optical Loop diagnostic :Data_20090709_162534.
NOTA: 4.5L/D statistics: 6.7 4.46, peak/energy:0.005.

PSF7 ............... SLOPE NULL ZERO..
CLOSE LOOP. high flux. LAMP flower pot WS
tt:0.67 volt
CCD39:196Hz.... 1700 ph per subaperture.
REC
142233.
ccd49: 0.4Hz (avg50).
GAIN:0.5
disturbance ON: modified to have 0.8 arcasec.
FW1:600
FW2:850-10.

NOTA: 3.5L/D statiswtics: 5.2 3.6, peak/energy:0.008.


PSF8 ............... SLOPE NULL ZERO..
CLOSE LOOP. high flux. LAMP flower pot WS
tt:0.67 volt
CCD39:196Hz.... 1700 ph per subaperture.
REC
142233.
ccd49: 0.4Hz (avg50).
GAIN:0.5
disturbance ON: 0.47 arcsec. (enrico original)
FW1:600
FW2:850-10.

Optical Loop diagnostic :Data_20090709_163353.
NOTA: 1.4L/D statiswtics: 1.8. 1 .8, peak/energy:0.045.


PSF9 ............... SLOPE NULL WITH DEFOCUS..
CLOSE LOOP. high flux. LAMP flower pot WS
tt:0.67 volt
CCD39:196Hz.... 1700 ph per subaperture.
REC
_142233.
ccd49: 0.4Hz (avg50).
GAIN:0.5
disturbance ON: 0.47 arcsec. . (enrico original)
FW1:600
FW2:850-10.

Optical Loop diagnostic :Data_20090709_163755.
NOTA: 1.48L/D statistics: 1.9. 1.7, peak/energy:0.047.

-----------------------------------BIN 2.

TEST OF THE HOT PROCEDURE TO RECORD THE PUPILS. --------------WORK AT DIFFERENT BINNING...
- flat the secondary at bin 1 200Hz and TT.10, close loop small gain.
- camera lens at mid-range, pupils center: 29,29,11,11.Diameter:aprox15. POSITION OF THE CAMARA LENS:

ADQUIRE INTMAT AND REC MAT. FOR THE BINNING 2.





-- EnricoPinna - 2009-07-10