You are here: Foswiki>FLAO Web>TowerLab>TowerTest20090701 (02 Jul 2009, EnricoPinna)Edit Attach
Francesca,JCarlos,Alfio

Failure start up WFS system. WFS first start up failure, some process didn't come up; ccd39 controller,slope computer,bayside X stage,pupil check.

Solution; power off all the systems, close the WFS GUI and stop the WFS arbitrator.

Pupils a little binietting, using the cube rot to have the right pupils dimension ......position: 0.1 relative

TEST OF REC MATRIX ACQUIRED FROM A SINUSOIDAL PROCEDURE.

torre10
tt: 21.3 (lambda/D)
ccd:400Hz
FIBRA PICCOLA....

TEST close loop of the rec, the first try rip at gain: 0.1, acquire loop diagnostic to identify the bad modes.

CLOSE LOOP WITH REC.

rec: rec_sin_600m_30Hz_Noseq_new.fits
gain:-0.1
frame:4000
NO_dist
cl:20090701_102535

rec: rec_sin_600m_30Hz_Noseq.fits
gain:0.4
frame:4000
NO_dist
cl:20090701_111446

**********

presi i dati per:

sinus_m408_30Hz
sinus_m408_30Hz

*************

set di intmat in openloop:

600m_Seq_30Hz

***********************

si risalva un cl a 600m sia sin che pp per farne il confronto


rec: rec_sin_600m_30Hz_Noseq.fits
gain:0--> oloop
frame:4000
NO_dist
cl:20090701_121321

psf1.fits
average frame:30

--> ricarichiamo la REC 600m_30Hz_Noseq che prima (cl:20090701_111446) funzionava fino a gain=0.4 !!

rec: rec_sin_600m_30Hz_Noseq.fits
gain:0.01 poi skippa il secondario!
frame:4000
NO_dist
cl:20090701_122559

AGW unmount

TT calibration

Written procedure "calibra_tt.py" (aoserver: $ADOPT_SOURCE/scripts) to automatize measurements.

Data saved in /home/labot/testw2/measurements/CalibTT/ <tracknum>

Analysis: analyse.pro in the same directory as the measurements.

Acquisitions:

20090701_185755

Fiber: 5.8 um
Lamp: halogen
FW1: T=99.6%
Freq: 100hz - 1000hz step 100hz
Amp: 1V-8V step1V

Notes: good measurements except low amplitudes (1V-2V) at freq >600Hz

-- JuanCarlosGuerra - 2009-07-01
Topic revision: r3 - 02 Jul 2009, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.