Test Software

Alfio, Juan Carlos

Acquisition and reduction software test. Diagnostic
New INTMAT and RECMAT.

Disturbance:

20090306_142424.fits
Generato disturbo PUSH-PULL.
nmodes=650
FR=5
Cycle=1
Ampl = pp_650.fits

NOTE: to apply the disturbance open/close.

Acquisition: 20090306_150246.fits
Disturbance: sopra.
Iteration:5.
Gain:0.1.
TT amp:2.
binning:1.

NOTA: VERY GOOD!!!

Calculation of the INTMAT and RECMAT.

Intmat_20090306_151453.fits
Rec_20090306_151453.fits

acq:20090306_150246.fits.
nmodes=600.
Cut=1 mode.

Intmat_20090306_152627.fits
Rec_20090306_152527.fits

acq:20090306_150246.fits.
nmodes=500.
Cut=1 mode.

NOTE: During the close loop the secondary mirror went in failure. Recover the secondary was needed.
OPEN AND CLOSE LOOP TEST.

CLOSE LOOP.

INTMAT:
RECMAT:Rec_20090603_151453.fits
Disturbance: atm_comm02_kolm_42m_s5_v2.fits
NMODES: 600
GAIN:0.1
FILTER1:50/50
FILTER2:850/10.
SlopeNull: N-A

CCD39:
Speed:800Hz
Dark:161705.fits

slopes from secundary: towertest/disturbances/20090603/cl_gain0.1.
slopes_1_switchbcu.sav
slopes_2_switchbcu.sav

OPEN LOOP.

GAIN:0


slopes from secundary: towertest/disturbances/20090603/cl_gain0/.
slopes_1_switchbcu.sav
slopes_2_switchbcu.sav
OPEN AND CLOSE LOOP PSF.

CLOSE LOOP.

INTMAT: Intmat_20090603_151453.fits
RECMAT:Rec_20090603_151453.fits
Disturbance: atm_comm02_kolm_42m_s5_v2.fits
NMODES: 600
GAIN:0.4
FILTER1:50/50
FILTER2:850/10.
SlopeNull: _164150.fits

CCD39:
Speed:200Hz
Dark:161705.fits

CCD47:
bin=1.
Dark:163859.fits.
frame saved (100): /home/labot/measurements/towertest/atmospheric disturbance/030609/speed200Hz/cl_gain0.4
psf_1.fits.
psf_2.fits

slopes saved:
/home/labot/measurements/towertest/atmospheric disturbance/030609/speed200Hz/cl_gain0.4
slopes_1_switchbcu.sav
slopes_s_switchbcu.sav

GAIN:0.1.

/home/labot/measurements/towertest/atmospheric disturbance/030609/speed200Hz/cl_gain0.1
psf_1.fits.
psf_2.fits

slopes saved:
/home/labot/measurements/towertest/atmospheric disturbance/030609/speed200Hz/cl_gain0.1
slopes_1_switchbcu.sav
slopes_s_switchbcu.sav

OPEN LOOP.

INTMAT: Intmat_20090603_151453.fits
RECMAT:Rec_20090603_151453.fits
Disturbance: atm_comm02_kolm_42m_s5_v2.fits
NMODES: 600
GAIN:0.0
FILTER1:50/50
FILTER2:850/10.
SlopeNull: _164150.fits

CCD39:
Speed:200Hz
Dark:161705.fits

CCD47:
bin=1.
Dark:163859.fits.
frame saved (100): /home/labot/measurements/towertest/atmospheric disturbance/030609/speed200Hz/cl_gain0
psf_1.fits.
psf_2.fits

slopes saved:
/home/labot/measurements/towertest/atmospheric disturbance/030609/speed200Hz/cl_gain0
slopes_1_switchbcu.sav
slopes_s_switchbcu.sav

NOTA: Few problems with the GUIs.
After a flat command a waffle apper , saved screenshot waffleccd39.png.
The command applied two times the waffle structure went off, secundary ok.
SINUSOIDAL DISTURBANCES:

OPEN LOOP.
Disturbance: sin_inj_20m_freq20_go04_ampl_ok.fits.
REC: Rec_20090603_151453.fits
GAIN=0

slopes saved: /towerdata/
sin_20m_freq20_switchbcu.sav
sin_20m_freq20_switchbcu2.sav

Disturbance: sin_inj_5m_freq30_go04_ampl_ok.fits.

slopes saved:/towerdata/
sin_5m_freq30_switchbcu.sav
sin_5m_freq30_switchbcu2.sav

Disturbance: sin_inj_5m_freq45_go04_ampl_ok.fits.

slopes saved:/towerdata/
sin_20m_freq45_switchbcu.sav
sin_20m_freq45_switchbcu2.sav

-- AlfioPuglisi - 2009-06-03
Topic revision: r2 - 08 Jun 2009, AlfioPuglisi
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.