system restarted with new configuration

this includes also the list of actuator working in clean room

451 remove dfrom position, it was jumping while rest

collect_ff_data(/no_flat)

tn =20150528_135602

we found a large residual on 539

new run with ff, tn =20150528_141824

last runvi ff 20150528_143548

New ff matrix, 10 cycles

20150528_145448 the reduction was ok

Many errors found

communication stopped with msgd

hexmoveby not working

it is because the reference was not update, so the relative movement did not work

marco fixed it by adding a hexgetpos in the hexmoveby in the wrapper

New ff installed

lots of error when restarting the sw. idl died, fast disconnected from msgdmn

2015-05-28 17:25:53

Starting the adsecarb we had jumpig actuators

Review of the procedure for the optical calibration

preliminary flattening

we corrected the astigmatism and the trefoils, and aligned to minimize coma, tilt and power

preliminaryflat: 20150528_195232/

if function sampling
Tracknum modes amp iter rersolution flattening modes flat rms flat notes
20150528_201734 0-49 100 3 1000x1000 20150528_203514 50 200nm  
20150528_203947 50-99 100 3 1000x1000 20150528_205731 100 100nm  
20150528_213015 0-249 100 3 500x500       bad intersection mask
the intersection mask of 250 modes is really poor, and thetest of flattening was diverging in the center

we tried to apply a command at the fisrt and second act ring:

act=adsec.mir_act2

IDL> c0=fltarr(672)
IDL> c0[act[0:8]]=1
IDL> display, c0
IDL> c1=fltarr(672)
IDL> c1[act[9:23]]=1
IDL> display, c1

applying a (approx) 600nm command on the firs and approx 300 on the second ring
Tracknum modes amp iter notes flattening modes rms note
20150528_223025 0-249 100 3   20150528_225646 250 30  
20150528_231311 250,499 100 3          
20150528_231527 500-662 100 3          

sampling for the modal basis

Tracknum modes amp iter
20150528_233025 0-9 500 51
20150528_233446 10-24 250 51
20150528_234040 25-49 100 31
20150528_234425 50-74 100 31
20150528_234703 75-99 100  
20150528_234947 100-124 100  
20150528_235155 125-149 100  
20150528_235357 150-174    
20150528_235549 175-199 80  
20150528_235730 200-224 80  
20150528_235920 225-249 80  
20150529_000112 250-274 80  
20150529_000304 275-299 60  
20150529_000443 300-324 60  
20150529_000553 325-349 60  
20150529_000701 350-374 60  
20150529_000811 375-399 60  
20150529_000924 400-424 60  
20150529_001038 425-449 60  
20150529_001145 450-474 60  
20150529_001257 475-499 50  
20150529_001403 500-524 50  
20150529_001511 525-549 50  
20150529_001619 550-574 50  
20150529_001726 575-599 50  
20150529_001833 600-624 50  
20150529_001940 625-649 50  
20150529_002051 650-661 50  
Topic revision: r4 - 29 May 2015, RunaBriguglio