-- RunaBriguglio - 03 Apr 2010

Preliminary operations

Environmental conditions will possibly be favourable during the night. John Hill will operate the telescope remotely.

01:20UT AOS stopped and restarted

01:28UT: preliminary inspection of the gap (nominal values, i.e. 24um<pos<39um); all processes on and no failure noticed. The telescope is at zenith.

01:29UT MirrorSet successfully

Maximum mode index used to apply flat -> # 416
COMM MAX [m] 7.76172e-05
FORCE MAX [N] 0.271323

Software preparation

According to John Hill's schedule, thin shell astigmatism will be sampled at changing elevation during the first part of the night. This operation requires that the automatic astigmatism offload is off: this operation is started and stopped with the command enable/disable_ho_offload, i.e. the offload of both high orders and astigmatism is executed by the same routine.

To disable the astigmatism correction without stopping the (real) high order correction, the routine get_zern_flat_lut is momentarily modified in $ADOPT_ROOT.

original code

Function get_zern_flat_lut, elevation

@adsec_common
zz = fltarr(22)
vv = linfit([20,35,90], [2.5e-6, 2e-6, 0])
zz[5] = vv[0]+vv[1]*elevation
return, zz

End

modified code

Function get_zern_flat_lut, elevation

@adsec_common
zz = fltarr(22)
vv = linfit([20,35,90], [2.5e-6, 2e-6, 0])
zz[5] = vv[0]+vv[1]*elevation

zz[5] = 0

return, zz

End

test successful.

the shell is put on its rest position to move telescope to 20°. a failure is detected when the command is sent

DOME OPENED

01:58UT approx

02:07UT the telescope is back at zenith, and shell set again

Maximum mode index used to apply flat -> # 413
COMM MAX [m] 7.76369e-05
FORCE MAX [N] 0.285740

the force range is however very large (-0.5, 0.5). the shell is then rested and set again

02:12:02UT shell rest

02:16:14 shell set

Maximum mode index used to apply flat -> # 413
COMM MAX [m] 7.76224e-05
FORCE MAX [N] 0.286672

again, at the first iteration of ho_offload, the forces are very large (-0.7,0.7)

the shell is then ripped and set again

02:21:47UT shell set again

Maximum mode index used to apply flat -> # 418
COMM MAX [m] 7.76234e-05
FORCE MAX [N] 0.264147

02:29:13 shell ripped

02:33:30 mirror set; disable_ho_offload

print, minmax(chcurraverage) (data from: ret=telemetry_file('20100403_023436','20100403_023442', /dosa)

-0.0699151 -0.0221951
-0.272972 0.182348
-0.270131 0.185035
-0.270197 0.184708
-0.270351 0.184801
-0.269861 0.185307
-0.500606 0.436494 ho_offload

for comparison:

print, minmax(chcurraverage) (data from:ret2=telemetry_file('20100403_013100','20100403_013117', /dosa)

-0.271279 0.196489
-0.271543 0.196282
-0.270923 0.196266
-0.271131 0.196460
-0.270639 0.196273
-0.271123 0.196320
-0.335034 0.282754
-0.334052 0.282826
-0.334768 0.282506
-0.334450 0.282523

02:51UT telescope moved to

03:11:35+1um astigm, wrong direction, in the end -2.5 um

03:25:37 shell ripped to restart again

high order modes correction disabled, modified fsm_set_flat

03:29:30UT shell set

Maximum mode index used to apply flat -> # 302
COMM MAX [m] 7.76375e-05
FORCE MAX [N] 0.249078

IDLCTRL00 |ALW| 14064|2010-04-03 03:30:38.085285| MAIN > HO_offload_disabled

Astigmatism correction:

20° v[5]=5.5e-6 03:53:07 on IDLCTRL.log

43.3° v[5]=2.4e-6 04:04:24

58.8° v[5]=1.23-6 04:18:32 wind:3 m/s. analyze data!!!

04:37:40 shell set again

04 :59:10 telescope moved, analyze telemetry

60.0° v[5]=2e-6 05:09:31 seeing is a little better

66.3° v[5]=1.5e-6 05:17:02

moving 5:17:30

71.5° v[5]=1.2e-3 05:23:01

05:24:10 approx moving

85.8° v[5]=0 05:33:17

60° v[5]=2e-6

53° v[5]=3e-6 05:49:19

27° v[5]=4.5e-6

55° v[5]=1.7e-6

54° v[5]=2.0e-6

37° v[5]=3.6e-6 06:17:58

43° v[5]=3.2e-6

FASTDGN00 |ERR| 1396587|2010-04-03 06:51:53.324063| FUNCTEMERGENCYST > CHCURRAVERAGE-0530 = -0.8

shell set again 07:17:17 enable_ho_offload

Maximum mode index used to apply flat -> # 260

COMM MAX [m] 7.76656e-05
FORCE MAX [N] 0.330535

ret4=telemetry_file('20100403_071717','20100403_071729', /dosa)

print, minmax(ret4.chcurraverage)

-0.324843 0.146047
-0.325293 0.148901
-0.327216 0.149484
-0.328306 0.143371
-0.350551 0.266387 ho_offload
-0.348361 0.269547
-0.349668 0.269109

new LUT for astigmatism, with 14 samples

astig.png

07:28 shell set again to install new LUT

Maximum mode index used to apply flat -> # 297
IDLCTRL00 |ALW| 20096|2010-04-03 07:29:29.456566| MAIN > step # 3
IDLCTRL00 |ALW| 20097|2010-04-03 07:29:29.681106| MAIN > Param Selector: 000041C0
IDLCTRL00 |ALW| 20098|2010-04-03 07:29:29.705455| MAIN > COMM MAX [m] 7.75985e-05
IDLCTRL00 |ALW| 20099|2010-04-03 07:29:29.706048| MAIN > FORCE MAX [N] 0.275228

same error on ho_offload

07:34:29 set again

W operation

pointed
  NSV 6945 2.29 V* 15 07 18.0659 +24 52 09.104 184.81 -163.51 50.70 -0.000024 -7.3 -7.30   5.36 4.93 4.7 4.5 F5V

board setup, and 2000 images acquired from CCD 39

Single frame example:

sky_pupils.png

Average of frames:

sky_pupils_average.png

The spider shadow is clearly visible: that could be a problem for the interaction matrix (taken with RR optics without spider shadow)

08:51UT dome closed due to wind and clouds
I Attachment Action Size Date Who Comment
astig.pngpng astig.png manage 2 K 04 Apr 2010 - 00:52 ArmandoRiccardi Astigmatism vs elevation
sky_pupils.pngpng sky_pupils.png manage 10 K 04 Apr 2010 - 00:55 ArmandoRiccardi Pupils on sky (single shot)