PSF VS FoV position

Centrale TN150719

TN151237 X=+0.15mm

TN151319 X=-0.15mm

TN151432 Y=+0.15mm

TN151513 Y=-0.15mm

TN151615 X=-0.15mm Y=-0.15mm

TN151717 X=+0.15mm Y=+0.15mm

TN151819 X=-0.15mm Y=+0.15mm
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nphSorted ascending AntiDrift SR filter exp #frames disturb SN notes
20150603_150719 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 255 ON 44.5   1 1 ON ncpa.fits  
20150603_151432 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 256 ON 43.6   1 1 ON ncpa.fits  
20150603_151717 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 256 ON 43.0   1 1 ON ncpa.fits  
20150603_151201 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 258 ON -1.0 ? -1 -1 ON ncpa.fits  
20150603_151237 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 258 ON 43.8   1 1 ON ncpa.fits  
20150603_151615 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 261 ON 49.8   1 1 ON ncpa.fits  
20150603_151319 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 262 ON 45.6   1 1 ON ncpa.fits  
20150603_151819 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 263 ON 49.1   1 1 ON ncpa.fits  
20150603_151513 ??? 8.4 324000 -2147483648 NaN -1.00 113212 1 500 990 0.3 0.2 -1.0 3 271 ON 45.6   1 1 ON ncpa.fits  
IDL string: ['20150603_150719','20150603_151201','20150603_151237','20150603_151319','20150603_151432','20150603_151513','20150603_151615','20150603_151717','20150603_151819']

NCPA con procedura automatica

Astigm Z5 automatico da -40 +40nm su specchio con Christine attiva TN154624-> TN154822 SR=16 - 36 - 39 - 23 - 8

Diminuiamo lo step (ora a 10nm) primo -30nm (TN155052) -> +10nm (TN155248) SR=24 - 34 - 41 - 40 - 32 L'algoritmo fitta +6nm

Scan di fuoco -20nm +20nm SR= 23 35 42 39 30 -> FITTATO +2.nm

Scan l'altro astigmatismo Z6 -20nm +20nm SR 25 35 41 40 33 -> FIT 3.5nm

Scan Z7 -10nm +10nm FITTATO -1nm

Scan Z8 -10nm +10nm FITTATO =+2nm

Scan Z9 trif -10nm +10nm FITTATO =-1.5nm

Scan Z10 trif -10nm +10nm FITTATO = -0.3nm

Scan Z11 Sfer -10nm +10nm FITTATO = 0.9nm

Rifacciamo Z5 passo fino -10nm +10nm FITTATO= -5nm

Rifacciamo Z6 passo fino -10nm +10nm FITTATO= +2.8nm

Salviamo SHARK_Optimized.

Acquisiamo PSF con NCPA ON, no atmosfera Christine on 1000 frames TN163018

Acquisiamo PSF con NCPA OFF, no atmosfera Christine on 1000 frames TN163105

Acquisiamo PSF con NCPA ON, atm 0.8 Christine on 1000 framesTN163453 TN163820
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150603_163453 ??? 8.4 324000 10 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 252 ON 42.8   1 1 ON ncpa.fits - long stop of 68 frames at frame 809 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150603_163536 ??? 8.4 324000 11 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 257 ON 39.1   1 1 ON

ncpa.fits

TUTTE =0!!!

- long stop of 60 frames at frame 2413 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150603_163608 ??? 8.4 324000 9 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 258 ON 41.7   1 1 ON ncpa.fits - long stop of 112 frames at frame 3579 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150603_163639 ??? 8.4 324000 11 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 256 ON 42.3   1 1 ON ncpa.fits - long stop of 116 frames at frame 1526 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150603_163711 ??? 8.4 324000 11 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 253 ON 41.4   1 1 ON ncpa.fits - long stop of 88 frames at frame 2297 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150603_163750 ??? 8.4 324000 11 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 257 ON 41.9   1 1 ON ncpa.fits - long stop of 117 frames at frame 727 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150603_163820 ??? 8.4 324000 9 NaN -1.00 113212 1 500 990 0.5 0.4 0.3 3 259 ON 42.0   1 1 ON ncpa.fits - long stop of 87 frames at frame 2209 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
IDL string: ['20150603_163453','20150603_163536','20150603_163608','20150603_163639','20150603_163711','20150603_163750','20150603_163820']

SCOPRIAMO ALLA FINE CHE ABBIAMO LE SLOEP OFFSET SUL W (15nm di astigm)! Quindi i valori delle NCPA hanno un bias

Tolte Slopeoffset Scan Z5 -20nm +20 per errore con la turbolenza FITTATO -16nm

Rifacciamo scan senza turbolenza.... FINE DAYTIME!!!!