NCPA measuerement senza avere delle slopenull inserite per sbaglio

Ripartiamo dal set di NCPA ottenute ieri.

Scan automatico con SR misurato su psf Shift&Add

Z4fuoco +1nm
Z5astig -11nm
Z6astig +6.5nm
Z7coma -0.5nm
Z8coma 2.nm
Z9trif -1.nm
Z10trif 0nm
Z11sfer +2nm

file NCPA_optimized3

INTMAT con modi >500

Intmat 3l/D con KL

NIENTE PERCHE' le KL dalla V14 in poi sono tutte con 500 modi!!!

Acquisizione PSF 500 modi daytime

NCPA ON; ATM OFF; Christine ON - TN095026, 095055, 095123, 095156
NCPA OFF; ATM OFF; Christine ON - TN095257, 095325, 095356, 095437

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150604_095026 ??? 8.3 323998 11 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 278 ON 47.4   1 1 ON zero.fits - long stop of 110 frames at frame 1071 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095055 ??? 8.3 323998 10 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 278 ON 46.9   1 1 ON zero.fits - long stop of 114 frames at frame 0 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095123 ??? 8.3 323998 9 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 277 ON 47.2   1 1 ON zero.fits - long stop of 61 frames at frame 2685 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095156 ??? 8.3 323998 9 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 276 ON 47.7   1 1 ON zero.fits - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095257 ??? 8.3 323998 7 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 277 ON 40.0   1 1 ON zero.fits - long stop of 108 frames at frame 3919 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095325 ??? 8.3 323998 9 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 280 ON 40.1   1 1 ON zero.fits - long stop of 69 frames at frame 2506 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095356 ??? 8.4 323998 8 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 274 ON 39.4   1 1 ON zero.fits - long stop of 112 frames at frame 436 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_095437 ??? 8.4 323998 8 NaN -1.00 113212 1 500 990 0.4 0.3 0.2 3 273 ON 39.5   1 1 ON zero.fits - long stop of 110 frames at frame 3149 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
IDL string: ['20150604_095026','20150604_095055','20150604_095123','20150604_095156','20150604_095257','20150604_095325','20150604_095356','20150604_095437']

NCPA ON; ATM 0.8"; Christine ON - TN100451, 100524, 100613 ,

NCPA OFF; ATM 0.8"; Christine ON - TN100808, 100839, 100910, 100938
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150604_100451 ??? 8.3 323998 8 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 282 ON 40.5   1 1 ON zero.fits - long stop of 110 frames at frame 411 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100524 ??? 8.3 323998 11 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 281 ON 40.5   1 1 ON zero.fits - long stop of 111 frames at frame 1623 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100613 ??? 8.3 323998 10 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 280 ON 41.1   1 1 ON zero.fits - long stop of 108 frames at frame 1317 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100642 ??? 8.3 323998 8 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 281 ON 39.5   1 1 ON zero.fits - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100808 ??? 8.3 323998 10 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 277 ON 33.4   1 1 ON zero.fits - long stop of 57 frames at frame 3177 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100839 ??? 8.3 323998 3 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 277 ON 34.5   1 1 ON zero.fits - long stop of 60 frames at frame 1335 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100910 ??? 8.3 323998 8 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 278 ON 32.0   1 1 ON zero.fits - long stop of 82 frames at frame 2961 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_100938 ??? 8.3 323998 2 NaN -1.00 113212 1 500 990 0.4 0.3 0.3 3 277 ON 33.4   1 1 ON zero.fits - long stop of 77 frames at frame 1603 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
IDL string: ['20150604_100451','20150604_100524','20150604_100613','20150604_100642','20150604_100808','20150604_100839','20150604_100910','20150604_100938']

Nuovo ricostruttore 500 modi

TrackNo RefStar Mag El WindSorted ascending DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20150604_110732 ??? 8.5 323998 3 NaN -1.00 104618 1 500 990 0.5 0.5 0.4 3 233 ON 43.6   1 1 ON zero.fits - long stop of 60 frames at frame 3536 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_110249 ??? 8.5 323998 4 NaN -1.00 104618 1 500 990 0.4 0.3 -1.0 3 231 ON 47.9   1 1 ON zero.fits - long stop of 104 frames at frame 2181 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_110352 ??? 8.6 323998 5 NaN -1.00 104618 1 500 990 0.4 0.3 -1.0 3 227 ON 47.2   1 1 ON zero.fits - long stop of 60 frames at frame 1585 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_110321 ??? 8.5 323998 6 NaN -1.00 104618 1 500 990 0.4 0.3 -1.0 3 232 ON 48.4   1 1 ON zero.fits - long stop of 77 frames at frame 3408 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_110830 ??? 8.5 323998 7 NaN -1.00 104618 1 500 990 0.5 0.5 0.4 3 235 ON 44.5   1 1 ON zero.fits - long stop of 101 frames at frame 1126 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_110801 ??? 8.5 323998 9 NaN -1.00 104618 1 500 990 0.5 0.5 0.4 3 233 ON 44.6   1 1 ON zero.fits - long stop of 112 frames at frame 2413 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_111054 ??? 8.5 323998 11 NaN -1.00 104618 1 500 990 0.4 0.3 0.3 3 244 ON 41.4   1 1 ON zero.fits - long stop of 93 frames at frame 2396 - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits
20150604_111009 ??? 8.5 323998 12 NaN -1.00 104618 1 500 990 0.4 0.4 0.4 3 241 ON 44.0   1 1 ON zero.fits - No shark darks found matching /local/towerdata/wfs_calib_LBTIDX/shark/backgrounds/bin1/*_cube.fits

-- EnricoPinna - 04 Jun 2015
  • psf_20150603_110801.png:
    psf_20150603_110801.png
Topic revision: r2 - 04 Jun 2015, AlfioPuglisi