Compute PMZ:

/home/aosup/Source/Magellan/calib/adsec/585m/data/MMmatrix_MAG585_20110706.sav

make_offload_matrix_proj, imfile, c2z=c2z, nz=22, z2c=z2c, umat=umat, /ver, thr=2e-2, /sav

result: 20121119_182155

make_offload_matrix_proj, imfile, c2z=c2z, nz=22, z2c=z2c, umat=umat, /ver, thr=1e-2, /sav

result: 20121119_182827

Now selected and used the 20121119_182827

Test forze central obscuration:

histfile='/home/aosup/adopt/current/calib/adsec/585m/CMD/disturb/dist_mag585_atm_s0.8_L040.0_v15.0_ovfreq1000.00_sd1983.fits'

print, minmax((adsec.ff_matrix ## hist*2)[adsec.mir_act2act[indgen(total(adsec_shell.n_act_ring[0:3]))]])/max(abs(adsec.ff_matrix ## hist*2))
-0.163159 0.129334 ----> 12-16%

Trovati svariati errori: - slave non maneggiati bene nel calcolo dei centri e if - slave non maneggiati bene in generale - errore di sintassi

misure riprese: make_offload_matrix_proj, imfile, c2z=c2z, nz=22, z2c=z2c, /ver, thr=1e-2, /sav, slave=slm /home/aosup/adopt/measures/adsec_calib/optical-projection/20121120_022740/

DA UPDATARE A LBT la procedura MAKE_OFFLOAD_MATRIX_PROJ

-- MarcoXompero - 19 Nov 2012
Topic revision: r3 - 20 Nov 2012, MarcoXompero