remove from control loop 552: with 0.15 force about 150 modes corrected

/home/aosup/adopt/measures/adsec_calib/adsec_save/20120529_152836/adsec_save.sav

Test slew 1 time: ok

Tue May 29 15:39:55 CEST 2012 shell set @ 25 c.a: with removed act 552. Going up to zenith to verify forces decrease and not increase /home/aosup/adopt/measures/adsec_calib/status_save/20120529_154201/ss.sav (28 deg) and /home/aosup/adopt/measures/adsec_calib/status_save/20120529_154502/ss.sav (90 deg)

Test FF 5 times for test: 20120529_154709 Tue May 29 15:47:57 CEST 2012 - CP temp 19.3 C deg Tue May 29 16:11:59 CEST 2012 - CP temp 19.8 C deg found noise high on DSP#381

Test FF 20 times: 20120529_161919 => BAD DATA: with ICC and HIGH GAINS frown, sad smile Tue May 29 16:20:42 CEST 2012 - CP temp 20.0 C deg Tue May 29 16:38:20 CEST 2012 - CP temp 20.4 C deg Tue May 29 16:46:13 CEST 2012 - CP temp 20.6 C deg - Added ICE Tue May 29 17:02:00 CEST 2012 - CP temp 20.8 C deg Tue May 29 17:10:30 CEST 2012 - CP temp 21.0 C deg Tue May 29 17:37:12 CEST 2012 - CP temp 21.6 C deg Tue May 29 17:47:01 CEST 2012 - CP temp 21.8 C deg

flux 46.6 - 46.7 = 15 L/min

flux -25.68 = 0L/min

PB serial number: 22 Old calibration: 6.506000e-003; -2.578785e+001 new calibrations: 4.620000e-003; 0.0

FF calibration 20 iterations: 20120529_194752 Problem on settings with FF: too much current on 670 (ab 0.14) -> removed check on ramp gain of fsm_set_flat() Tue May 29 19:47:57 CEST 2012 - CP temp: 21.6 Tue May 29 20:02:31 CEST 2012 - CP temp :21.6 Tue May 29 20:29:25 CEST 2012 - CP temp: 21.7 Tue May 29 20:52:58 CEST 2012 - CP temp: 22.00 Tue May 29 21:03:22 CEST 2012 - CP temp: 22.00 Tue May 29 21:13:39 CEST 2012 - CP temp : 22.2

Ok. good ff matrix fitting and good residuals. High noise verified on: DSP#381 Bad news for KL: the slave will be no more slave inserting the new ff matrix...

- TODO: calibrare flussimetro + test coi KL + step final configuration + turbulence final configuration + report + procedura + test set meno attuatori sullo slot

-- MarcoXompero - 29 May 2012