-- RunaBriguglio - 13 Feb 2012

System powered on at 20120213_141800. temperature setpoint=13°, bench=11°

record of diagnostic data: store in /home/aosup/adopt/measures/adsec_calib/noise/ tracknum/noise.sav, variable =BUF
Event T (CP Out) tracknum  
Temperature Setpoint 13      
power on 14 20120213_144459  
#657, driver disabled 14 20120213_144649  
DSP reset 14 20120213_145023  
after setting (unsuccessfull) 14 20120213_150047  
Power off      
Temperature Setpoint 16      
power on 17 20120213_153321  
#657, driver disabled 17 20120213_153429  
DSP reset 17 20120213_153616  
after setting (unsuccessfull) 17 20120213_153807  
Power off   20120213_155200  
Temperature Setpoint 19 21    
Power on 21 20120213_163225  
#657, driver disabled 21 20120213_163736  
DSP reset 21 20120213_163956  
after setting (unsuccessfull) 21 20120213_164255  
DSP ok 21 20120213_173740  
       
!! power temperature on crate #5, dsp#12 (bad one) is 41°, while the other DSPs are at 35 (on averg.)

investigate thius point with RBIASI

Probe cooler out1 crate#1 is disconnected (that's why the temperature read is 77°).

for this reason the water cold plate outlet is about 50°

check DSP power temp fro crate#5, dsp#12 (the bad one): temperature was 42 (when the gap was 55), and then 38 like the others when the gap was 41. cause or consequence????

act#478 removed ( reading=2m)