Problems on startup as reported on LogBook20110929Mag

May be a bad connection of flat cable??? to be checked.

Here the process dump analysis,
adsec.L            |INF|      2119|2011-10-14 10:34:02.137400|             MAIN > -- PROCESS DUMP START --
adsec.L            |INF|      2120|2011-10-14 10:34:02.137457|             MAIN > Fri Oct 14 10:34:02 2011
adsec.L            |INF|      2121|2011-10-14 10:34:02.852303|             MAIN > Get info from housekeeper.L variables...done
adsec.L            |INF|      2122|2011-10-14 10:34:03.185011|             MAIN > Get info from masterdiagnostic.L dump file: ADAPTIVE-SECONDARY_00001318581228_20111014_103348.log...dump loaded.
adsec.L            |INF|      2123|2011-10-14 10:34:03.433720|             MAIN > Get info from fastdiagn.L variables...done
adsec.L            |INF|      2124|2011-10-14 10:34:03.676798|             MAIN > Get info from fastdiagn.L variables...done
adsec.L            |INF|      2125|2011-10-14 10:34:05.006535|             MAIN > No variables in Alarm for family: CHCURRAVERAGE
adsec.L            |INF|      2126|2011-10-14 10:34:05.010107|             MAIN > No variables in Alarm for family: CHDISTAVERAGE
adsec.L            |INF|      2127|2011-10-14 10:34:05.013605|             MAIN > No variables in Alarm for family: CHINTCONTROLCURRENT
adsec.L            |INF|      2128|2011-10-14 10:34:05.017314|             MAIN > Fam: CHDISTRMS in Alarm over indexes: 656 658 659 660 661 662 663
adsec.L            |INF|      2129|2011-10-14 10:34:05.021077|             MAIN > No variables in Alarm for family: CHCURRRMS
adsec.L            |INF|      2130|2011-10-14 10:34:05.024510|             MAIN > No variables in Alarm for family: MODES
adsec.L            |INF|      2131|2011-10-14 10:34:05.027980|             MAIN > No variables in Alarm for family: CHNEWDELTACOMMAND
adsec.L            |INF|      2132|2011-10-14 10:34:05.031437|             MAIN > No variables in Alarm for family: CHFFCOMMAND
adsec.L            |INF|      2133|2011-10-14 10:34:05.034910|             MAIN > No variables in Alarm for family: CHFFPURECURRENT
adsec.L            |INF|      2134|2011-10-14 10:34:07.776673|             MAIN > Get info from housekeeper.L variables configuration...done
adsec.L            |INF|      2135|2011-10-14 10:34:07.777867|             MAIN > Get info from adamhousekeeper.L variables...done
adsec.L            |INF|      2136|2011-10-14 10:34:08.480556|             MAIN > Get info from housekeeper.L variables...done
adsec.L            |INF|      2137|2011-10-14 10:34:08.481892|             MAIN > Get info from adamhousekeeper.L variables...done
adsec.L            |INF|      2138|2011-10-14 10:34:08.652729|             MAIN > Adam watchdog not expired.
adsec.L            |INF|      2139|2011-10-14 10:34:08.679136|             MAIN > Ethernet connection watchdog on Switch BCU not expired.
adsec.L            |INF|      2140|2011-10-14 10:34:08.692699|             MAIN > Ethernet connection watchdog on Crate BCU not expired.
adsec.L            |INF|      2141|2011-10-14 10:34:08.694761|             MAIN > All global counters are progressing
adsec.L            |INF|      2142|2011-10-14 10:34:08.713259|             MAIN > DSP board watchdog not expired.
adsec.L            |INF|      2143|2011-10-14 10:34:08.726206|             MAIN > Distance threshold check analysis result:
adsec.L            |INF|      2144|2011-10-14 10:34:08.726272|             MAIN > Not stable capacitive sensor on CH ID: 657 (times: 46)
adsec.L            |INF|      2145|2011-10-14 10:34:08.726312|             MAIN > Consider to remove the actuators with highest fault frequency from the close loop list.
adsec.L            |INF|      2146|2011-10-14 10:34:08.726351|             MAIN > Consider to remove the other detected actuators from the integrator control loop list.
adsec.L            |INF|      2147|2011-10-14 10:34:08.761027|             MAIN > Temperature analysis result:
adsec.L            |INF|      2148|2011-10-14 10:34:08.761113|             MAIN > Warn temperature value of: EXTERNALTEMPERATURE - index: 0 - value [C deg]=20.105078
adsec.L            |INF|      2149|2011-10-14 10:34:08.761153|             MAIN > Consider to fix the temperature ranges in houskeeper process or substitute the PT sensors.
adsec.L            |INF|      2150|2011-10-14 10:34:08.856505|             MAIN > OVERCURRENT probability for TOTAL PUSH CURRENT:   0.0%
adsec.L            |INF|      2151|2011-10-14 10:34:08.856602|             MAIN > OVERCURRENT probability for TOTAL PULL CURRENT:   0.0%
adsec.L            |INF|      2152|2011-10-14 10:34:08.856678|             MAIN > OVERCURRENT probability for SINGLE CRATE TOTAL CURRENT:   0.0%
adsec.L            |INF|      2153|2011-10-14 10:34:08.856718|             MAIN > -- PROCESS DUMP END --

After 10 minutes the system went up full working, Tested ADAM Modbus communication for Fastdiagnostic/Housekeeper/Arbitrator/IDL in the previous days. Today checked with IDL Tested Housekeeper by disabling an item of the average for a derived variable: the housekeeper correctly use the remaining variable. System powerd on, shell set and rest, powered off and recovered from failure all ok. ADAM Modbus test positive.

-- MarcoXompero - 14 Oct 2011