Procedura per lo smontaggio shell:

a) configuration of mirrorctrl in $ADOPT_ROOT/conf/adsec/current/processconf/mirrorctrl.conf: set WatchDog _ms to 0, set AdamWatchdogEnable to 0. a) configuration of idlctrl in $ADOPT_ROOT/conf/adsec/current/processconf/idlctrl.conf: set use_rpc to 0. b) power up single phase and three phase generator c) test ADAM response (ping adamdx(adamsx)) d) adsc_start e) adsceng f) from adsceng, open the "System Process" and kill AOArbitrator and AdsecArbitrator processes.


install the foam for the shell mount/dismount
when the power and fiber cable are re-inserted you can continue

g) send command: set_plot, 'x' h) from adsceng, open "IDL terminal" i) send command: print, fsm_power_on() l) send command: print, fsm_load_program(/auto, /mount)

-- WARNING: proceed from here only if you want mount/dismount the shell

m) send command: print, umount_shell()

- Corretta procedura per far si che gli attuatori wo_curr siano disabilitati quando si spinge per levare la shell - shell smontata, rimossi 6 magneti - pulita shell dalla colla con cloroformio e lametta

Misure altezza coils attuatori (sveglia armando): 597: +53um 620: +172um 631: +178um 632: +203um 650: +84um 656: +38um

Installata la shell verificata l'assenza di polvere installato tennon stretta la membrana

set della shell allentata e ristretta la membrana

controllare flat 151

-- MarcoXompero - 24 Feb 2011
Topic revision: r3 - 24 Feb 2011, MarcoXompero