Rimontata la board sugli stages

Riconnesso tutto

Test RON:
Alimentatore esterno 110 portatile, 110V non attraverso il pannello finale

Flowerpot OFF
LJ fan ON
LJ39 ON
LJ47 ON
FW OFF
ADC OFF
Bayside stages ON (frenate)
Pupil rerot ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.91 - RON (e-): 10.45 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.40 - RON (e-): 6.70 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.01 - RON (e-): 4.01 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.79 - RON (e-): 4.39 (estimated)

Stesse condizioni ma con ADC ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 22.03 - RON (e-): 11.02 (estimated)
Speed: 890 kpix/sec - RON (ADU): 16.57 - RON (e-): 8.29 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.30 - RON (e-): 4.15 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.60 - RON (e-): 4.80 (estimated)

Stesse condizioni ADC e FW ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 22.37 - RON (e-): 11.19 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.67 - RON (e-): 8.83 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.21 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.26 - RON (e-): 5.13 (estimated)

BOARD FUNCIOTNALITY CHECK

CCD39 OK
CCD47 OK
Lamp OK
FW1 OK
FW2 OK
Temp OK (tutte!)
TTM OK, ma problema fibre BOX47
Cube stage OK
Cube rotator OK
Stages OK
ADC1 OK
ADC2 OK
Camera lens OK

-- EnricoPinna - 09 Dec 2010