You are here: Foswiki>FLAO Web>TowerLab>LabTest20110601 (01 Jun 2011, EnricoPinna)Edit Attach

Visita sarta

Installazione dello styrodur del lineare

Functiotnality check

Fatto OPERATE, funziona tutto, solo la slitta X e ADC2 hanno fallito il setup alla prima, poi ha fuinzionato
Board setup 03 Jan 11 - W-only_2

Alziamo la lampada e abbiamo luce sul 39 e il 47

CCD39 OK
CCD47 OK
Stages XYZ OK
Flowerpot OK
Temperatures OK (1 sensirion ko e LM35 piramide ko)
FW1 e 2 OK
Lamp OK
TT ok anche il synch
Camera lens OK
PUP rot OK
Cub rot OK
Cub stage OK

RON TEST no cooling con AGW su carrello
OFF = cube stag, cube tras, lamp

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.11 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.56 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.00 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.00 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.31 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.58 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.65 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.00 (estimated)

RON TEST no cooling con AGW su carrello
OFF = cube stag, cube tras, lamp, stages XYX

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.89 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.52 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.92 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.91 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.29 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.51 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.63 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.99 (estimated)
>>>

N TEST no cooling con AGW su carrello
OFF = cube stag, cube tras, lamp, stages XYX. MONITOR

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.90 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.45 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.85 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.88 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.11 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.34 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.36 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.87 (estimated)
>>>