RON measurements (OLD STYLE):


SUMMARY ------------------------------------

ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.24 - RON (e-): 9.62 (estimated
Speed: 890 kpix/sec - RON (ADU): 13.59 - RON (e-): 6.79 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.04 - RON (e-): 4.02 (estimated)

RON measurements(NEW STYLE):
------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.08 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.57 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 3.99 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 3.98 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.08 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.16 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.22 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.27 (estimated)

First set with camera NOT COOLED and first two images with wrong filter.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_101019 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 268 ON 39.8 ? 50 300 ON - SR may be in error - Warning: Selected IRTC dark +48h old!
20110511_101127 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 268 ON 39.9 ? 50 300 ON - SR may be in error - Warning: Selected IRTC dark +48h old!
20110511_101238 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 266 ON 103.1 H 50 300 ON - SR may be in error
20110511_101508 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 258 ON 96.6 H 50 300 ON  
20110511_101618 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 267 ON 100.5 H 50 300 ON  
20110511_101729 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 267 ON 102.6 H 50 300 ON  

Reference measure without disturbance

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_104147 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.2 0.4 3 261 ON 99.3 H 50 200 OFF  
20110511_104235 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.2 0.4 3 260 ON 99.0 H 50 200 OFF  
20110511_104322 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.2 0.4 3 261 ON 98.1 H 50 200 OFF  
After opening/closing the loop again:

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_104953 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.2 0.2 3 257 ON 93.2 H 50 200 OFF  
20110511_105041 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.2 0.2 3 258 ON 94.0 H 50 200 OFF  
20110511_105128 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.2 0.2 3 258 ON 94.5 H 50 200 OFF  

Performance measurement with automatic script

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_110138 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 1.1 1.2 6 25 ON 1.9 H 400 37 ON  
20110511_110230 ??? 16.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 1.1 1.2 6 27 ON 1.9 H 400 37 ON  
20110511_110322 ??? 16.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 1.1 1.2 6 28 ON 1.7 H 400 37 ON  
20110511_110452 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 1.1 1.2 6 24 ON 1.8 H 400 37 ON  
20110511_110543 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 1.1 1.2 6 24 ON 1.2 H 400 37 ON  
20110511_110633 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 1.1 1.2 6 25 ON 1.4 H 400 37 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_111152 ??? 15.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.6 0.8 6 37 ON 11.8 H 400 37 ON  
20110511_111221 ??? 15.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.6 0.8 6 35 ON 11.3 H 400 37 ON  
20110511_111250 ??? 15.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.6 0.8 6 36 ON 11.4 H 400 37 ON  
20110511_111357 ??? 15.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.6 0.8 6 34 ON -1.0 ? -1 -1 ON  
20110511_111425 ??? 15.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.6 0.8 6 35 ON 11.8 H 400 37 ON  
20110511_111456 ??? 15.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.6 0.8 6 36 ON 12.0 H 400 37 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_111905 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.1 1.3 6 40 ON 16.9 H 300 50 ON  
20110511_111934 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.1 1.3 6 41 ON 18.0 H 300 50 ON  
20110511_112005 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.1 1.3 6 43 ON 17.5 H 300 50 ON  
20110511_112115 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.1 1.3 6 37 ON 16.3 H 300 50 ON  
20110511_112145 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.1 1.3 6 38 ON 16.8 H 300 50 ON  
20110511_112215 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.1 1.3 6 40 ON 16.8 H 300 50 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_112719 ??? 13.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.0 1.5 6 113 ON 31.8 H 100 150 ON - long stop of 63 frames at frame 969
20110511_112801 ??? 13.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.0 1.5 6 112 ON 32.0 H 100 150 ON  
20110511_112844 ??? 13.5 -2147483648 -2147483648 NaN -1.00 171410 3 66 200 1.0 1.5 6 114 ON 31.8 H 100 150 ON  
20110511_113023 ??? 13.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.0 1.5 6 108 ON 30.3 H 100 150 ON  
20110511_113105 ??? 13.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.0 1.5 6 114 ON 30.7 H 100 150 ON  
20110511_113148 ??? 13.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 1.0 1.5 6 114 ON 30.7 H 100 150 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_114953 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.4 0 40 ON 56.3 H 150 100 ON - Pyramid modulation mismatch
20110511_115027 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.4 0 41 ON 54.9 H 150 100 ON - Pyramid modulation mismatch
20110511_115101 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.4 0 40 ON 55.6 H 150 100 ON - Pyramid modulation mismatch
20110511_115223 ??? 12.6 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.4 0 38 ON 38.5 H 150 100 ON - Pyramid modulation mismatch
20110511_115257 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.4 0 39 ON 52.1 H 150 100 ON - Pyramid modulation mismatch
20110511_115331 ??? 12.6 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.4 0 37 ON 50.1 H 150 100 ON - Pyramid modulation mismatch

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_115651 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.4 0.5 3 62 ON 58.9 H 100 150 ON  
20110511_115733 ??? 11.4 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.4 0.5 3 68 ON 57.8 H 100 150 ON  
20110511_115816 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.4 0.5 3 62 ON 57.5 H 100 150 ON  
20110511_115955 ??? 11.6 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.4 0.5 3 59 ON 59.0 H 100 150 ON  
20110511_120038 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.4 0.5 3 61 ON 59.4 H 100 150 ON  
20110511_120120 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.4 0.5 3 61 ON 58.6 H 100 150 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_120456 ??? 10.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.5 0.6 3 163 ON 59.8 H 70 214 ON  
20110511_120549 ??? 10.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.5 0.6 3 163 ON 60.4 H 70 214 ON  
20110511_120643 ??? 10.5 -2147483648 -2147483648 NaN -1.00 171951 2 153 991 0.5 0.6 3 163 ON 60.3 H 70 214 ON  
20110511_120843 ??? 10.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.5 0.6 3 158 ON 58.6 H 70 214 ON  
20110511_120937 ??? 10.5 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.5 0.6 3 161 ON 59.3 H 70 214 ON  
20110511_121031 ??? 10.4 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 991 0.5 0.6 3 166 ON 60.0 H 70 214 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_121458 ??? 9.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 104 ON 79.8 H 70 214 ON  
20110511_121553 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 109 ON 82.5 H 70 214 ON  
20110511_121647 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 109 ON 83.8 H 70 214 ON  
20110511_121848 ??? 9.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 103 ON 83.3 H 70 214 ON  
20110511_121943 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 110 ON 79.3 H 70 214 ON  
20110511_122038 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 112 ON 80.0 H 70 214 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_122948 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 260 ON 84.9 H 50 300 ON  
20110511_123058 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 260 ON 82.7 H 50 300 ON  
20110511_123208 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 263 ON 83.9 H 50 300 ON  
20110511_123437 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 253 ON 85.0 H 50 300 ON  
20110511_123547 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 261 ON 84.3 H 50 300 ON  
20110511_123657 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 261 ON 81.8 H 50 300 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_130545 ??? 16.3 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.9 1.5 6 32 ON 4.1 H 400 37 ON  
20110511_130636 ??? 16.3 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.9 1.5 6 33 ON 4.4 H 400 37 ON - SR may be in error
20110511_130726 ??? 16.3 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.9 1.5 6 34 ON 4.2 H 400 37 ON  

GOOD SEEING test

Seeing 0.6"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_163114 ??? 16.5 -2147483648 -2147483648 NaN -1.00 152928 4 36 100 0.7 0.9 6 28 ON 13.1 H 400 37 ON  
20110511_163204 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 0.9 6 24 ON 10.1 H 400 37 ON  
20110511_163253 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 0.9 6 26 ON 10.1 H 400 37 ON  
20110511_163403 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 0.9 6 26 ON 8.8 H 400 37 ON  
20110511_163453 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 0.9 6 25 ON 8.8 H 400 37 ON  
20110511_163543 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 0.9 6 29 ON 9.7 H 400 37 ON  

Seeing 0.6"
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_164031 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.5 3 40 ON 62.3 H 100 150 ON  
20110511_164114 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.5 3 40 ON 64.4 H 100 150 ON  
20110511_164156 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.5 3 39 ON 65.8 H 100 150 ON  
20110511_164335 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.5 3 39 ON 62.7 H 100 150 ON  
20110511_164419 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.5 3 41 ON 61.9 H 100 150 ON  
20110511_164502 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.4 0.5 3 42 ON 62.3 H 100 150 ON  

Seeing 0.6"
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_164907 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 258 ON 88.0 H 50 300 ON  
20110511_165017 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 261 ON 86.8 H 50 300 ON  
20110511_165127 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 263 ON 89.1 H 50 300 ON  
20110511_165357 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 248 ON 89.3 H 50 300 ON  
20110511_165507 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 260 ON 87.6 H 50 300 ON  
20110511_165617 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 261 ON 88.1 H 50 300 ON  

BAD SEEING test

Seeing 1.5"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_170527 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 42 ON 17.5 H 100 150 ON  
20110511_170610 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 43 ON 17.9 H 100 150 ON  
20110511_170652 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 42 ON 18.1 H 100 150 ON  
20110511_170831 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 40 ON 17.0 H 100 150 ON  
20110511_170914 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 44 ON 17.7 H 100 150 ON  
20110511_170956 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 41 ON 18.3 H 100 150 ON  

Seeing 1.5"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_171502 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 263 ON 68.9 H 50 300 ON  
20110511_171611 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 265 ON 67.3 H 50 300 ON  
20110511_171721 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 265 ON 68.4 H 50 300 ON  
20110511_171950 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 252 ON 68.5 H 50 300 ON  
20110511_172100 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 263 ON 66.9 H 50 300 ON  
20110511_172210 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.6 3 264 ON 67.6 H 50 300 ON  

Seeing 1.2"
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_182735 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.7 3 263 ON 79.5 H 50 300 ON  
20110511_182845 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.7 3 262 ON 79.4 H 50 300 ON  
20110511_182955 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.7 3 263 ON 78.3 H 50 300 ON  
20110511_183225 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.7 3 253 ON 79.8 H 50 300 ON  

500 modes test

Seeing 0.8"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_183739 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.3 0.4 3 106 ON 87.2 H 70 214 ON  
20110511_183833 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.3 0.4 3 107 ON 87.2 H 70 214 ON  
20110511_183927 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.3 0.4 3 104 ON 85.7 H 70 214 ON  
20110511_184128 ??? 9.5 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.3 0.4 3 99 ON 84.2 H 70 214 ON  
20110511_184223 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.3 0.4 3 105 ON 87.4 H 70 214 ON  
20110511_184317 ??? 9.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.3 0.4 3 106 ON 86.0 H 70 214 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110511_184824 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.4 0.4 3 269 ON 87.1 H 50 300 ON  
20110511_184935 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.4 0.4 3 265 ON 88.9 H 50 300 ON  
20110511_185045 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.4 0.4 3 264 ON 86.1 H 50 300 ON  
20110511_185315 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.4 0.4 3 254 ON 87.7 H 50 300 ON  
20110511_185425 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.4 0.4 3 265 ON 88.5 H 50 300 ON  
20110511_185535 ??? 8.4 -2147483648 -2147483648 -1.00 -1.00 151120 1 500 990 0.4 0.4 3 265 ON 87.1 H 50 300 ON  

; BIN1
m8_5 = ['20110511_122948','20110511_123058','20110511_123208','20110511_123437','20110511_123547','20110511_123657']
m9_5 = ['20110511_121458','20110511_121553','20110511_121647','20110511_121848','20110511_121943','20110511_122038']
Acctest_bin1 = obj_new('aodataset',[m9_5, m8_5])
sr_b1 =Acctest_bin1->value('irtc.sr_se')
mag_b1 = Acctest_bin1->value('mag')

rec500_m9_5 =  ['20110511_183739','20110511_183833','20110511_183927','20110511_184128','20110511_184223','20110511_184317']
rec500_m8_5 = ['20110511_184824','20110511_184935','20110511_185045','20110511_185315','20110511_185425','20110511_185535']
rec500_Acctest_bin1 = obj_new('aodataset',[rec500_m8_5, rec500_m9_5])
rec500_sr_b1 =rec500_Acctest_bin1->value('irtc.sr_se')
rec500_mag_b1 = rec500_Acctest_bin1->value('mag')


S15_m8_5 = ['20110511_171502','20110511_171611','20110511_171721','20110511_171950','20110511_172100','20110511_172210']
S15_Acctest_bin1 = obj_new('aodataset',[S15_m8_5])
S15_sr_b1 =S15_Acctest_bin1->value('irtc.sr_se')
S15_mag_b1 = S15_Acctest_bin1->value('mag')


s06_m8_5 = ['20110511_164907','20110511_165017','20110511_165127','20110511_165357','20110511_165507','20110511_165617']
s06_Acctest_bin1 = obj_new('aodataset',[s06_m8_5])
s06_sr_b1 =s06_Acctest_bin1->value('irtc.sr_se')
s06_mag_b1 = s06_Acctest_bin1->value('mag')

; BIN2
m10_5 =['20110511_120456','20110511_120549','20110511_120643','20110511_120843','20110511_120937','20110511_121031']
m11_5 = ['20110511_115651','20110511_115733','20110511_115816','20110511_115955','20110511_120038','20110511_120120']
m12_5 =  ['20110511_114953','20110511_115027','20110511_115101','20110511_115223','20110511_115257','20110511_115331']

S06_m12_5 =  ['20110511_164031','20110511_164114','20110511_164156','20110511_164335','20110511_164419','20110511_164502']
S15_m12_5 =  ['20110511_170527','20110511_170610','20110511_170652','20110511_170831','20110511_170914','20110511_170956']

Acctest_bin2 = obj_new('aodataset',[m12_5, m11_5, m10_5])
sr_b2 =Acctest_bin2->value('irtc.sr_se')
mag_b2 = Acctest_bin2->value('mag')

S06_Acctest_bin2 = obj_new('aodataset',[S06_m12_5])
S06_sr_b2 =S06_Acctest_bin2->value('irtc.sr_se')
S06_mag_b2 = S06_Acctest_bin2->value('mag')

S15_Acctest_bin2 = obj_new('aodataset',[S15_m12_5])
S15_sr_b2 =S15_Acctest_bin2->value('irtc.sr_se')
S15_mag_b2 = S15_Acctest_bin2->value('mag')


m13_5 = ['20110511_112719','20110511_112801','20110511_112844','20110511_113023','20110511_113105','20110511_113148']
m14_5 =  ['20110511_111905','20110511_111934','20110511_112005','20110511_112115','20110511_112145','20110511_112215']
Acctest_bin3 = obj_new('aodataset',[m14_5, m13_5])
sr_b3 =Acctest_bin3->value('irtc.sr_se')
mag_b3 = Acctest_bin3->value('mag')


m15_5 =['20110511_111152','20110511_111221','20110511_111250','20110511_111425','20110511_111456']
m16_5 = ['20110511_110138','20110511_110230','20110511_110322','20110511_110452','20110511_110543','20110511_110633']
m16_5_take2 =  ['20110511_130545','20110511_130636','20110511_130726']

s06_m16_5 = ['20110511_163114','20110511_163204','20110511_163253','20110511_163403','20110511_163453','20110511_163543']
Acctest_bin4 = obj_new('aodataset',[m15_5, m16_5, m16_5_take2])
sr_b4 =Acctest_bin4->value('irtc.sr_se')
mag_b4 = Acctest_bin4->value('mag')
s06_Acctest_bin4 = obj_new('aodataset',[s06_m16_5])
s06_sr_b4 =s06_Acctest_bin4->value('irtc.sr_se')
s06_mag_b4 = s06_Acctest_bin4->value('mag')


; Simulation data (baseline and goal)
simul_table=fltarr(9,3)
simul_table[*,0]=[8.5,9.5,10.5,11.5,12.5,13.5,14.5,15.5,16.5]
simul_table[*,1]=[0.62,0.59,0.50,0.48,0.42,0.27,0.19,0.071,0.026]
simul_table[*,2]=[0.80,0.77,0.62,0.59,0.54,0.37,0.27,0.12,0.052]

; Simul SR vs Seeing
simul_mag = [8.5,10.5,12.5,14.5,16.5]
simul_nmag = n_elements(simul_mag)
; different seeings (v=15m/s)
simul_seeings = [0.6,0.8,1.0,1.2,1.5]
simul_nseeings = n_elements(simul_seeings)
                       ;                         -----> mag
flao_srH_v15 = [[87., 73., 67., 43., 15.], $
                           [80., 61., 52., 27., 05.], $
                           [68., 47., 35., 13., 02.], $
                           [44., 33., 20., 05., 01.], $
                           [21., 17.,  7.,  1., !VALUES.F_NAN]]/100.




plot, simul_table[*,0], simul_table[*,2], title='SR FLAO#2 @ tower testbench in May 11th, 2011', xtitle='R+I mag',ytitle='SR value (H band)', charsize=2, yrange=[0,1]
oplot, simul_table[*,0], simul_table[*,2],col='00ff00'x
oplot, simul_table[*,0], simul_table[*,1],col='0000ff'x
oplot, mag_b1, sr_b1, psym=4
oplot, mag_b2, sr_b2, psym=6
oplot, mag_b3, sr_b3, psym=5
oplot, mag_b4, sr_b4, psym=2
oplot, rec500_mag_b1, rec500_sr_b1, psym=1

oplot, simul_mag, flao_srH_v15[*,0], col='808000'x ; 0.6"
oplot, s06_mag_b4, s06_sr_b4, psym=4, col='808000'x
oplot, s06_mag_b2, s06_sr_b2, psym=4, col='808000'x
oplot, s06_mag_b1, s06_sr_b1, psym=4, col='808000'x

oplot, simul_mag, flao_srH_v15[*,4], col='000080'x ; 1.5"
oplot, s15_mag_b2, s15_sr_b2, psym=4, col='000080'x
oplot, s15_mag_b1, s15_sr_b1, psym=4, col='000080'x

  • test_may11.png:
    test_may11.png
Topic revision: r3 - 11 May 2011, MarcoXompero
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback