You are here: Foswiki>FLAO Web>TowerLab>LabTest20110429 (29 Apr 2011, EnricoPinna)Edit Attach
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.60 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.30 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.34 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.32 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.50 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.57 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.59 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.74 (estimated)

Mag12.5 Seeing 1.2

Primo giro, lui dice che non c'e' il disturbo nella elablib, ma nell'header del W il disturbo e' corretto, l'header del secondario no, ma dovrebbe contare quella del W

Header del W dice disturbo @625 in synch

Corretta la elablib: ora per il disturbo considera solamente l'header del W e ignora quello del adsec
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_110246 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 40 ON 33.5 H 300 50 ON  
20110429_110321 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 40 ON 32.1 H 300 50 ON  
20110429_110509 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 41 ON 32.4 H 300 50 ON  
20110429_110544 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 41 ON 33.1 H 300 50 ON  
20110429_110618 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.6 0.7 3 43 ON 33.4 H 300 50 ON  

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_111501 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 1.0 3 43 ON 30.3 H 300 50 ON  
Dopo il primo loop va in skipframe e poi stoploop automatico

Auto gain iterativo refurbished

Ora l'iterazine avviene solo su TT o su HO separatamente.

Labmode del W fallisce: subisce un colpo di derotatore dal TCS e ci ruota le pup.

Mag14.5 Seeing 1.2

Incertezza sull'oversampling veramente applicato....
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_114510 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.7 0.7 6 46 ON 3.9 H 600 25 ON - oversampling freq not a multiple of loop freq
20110429_114542 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.7 0.7 6 48 ON 3.7 H 600 25 ON - oversampling freq not a multiple of loop freq
20110429_114614 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.7 0.7 6 49 ON 3.5 H 600 25 ON - oversampling freq not a multiple of loop freq
20110429_114722 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.7 0.7 6 48 ON 3.4 H 600 25 ON - oversampling freq not a multiple of loop freq
20110429_114754 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.7 0.7 6 48 ON 3.4 H 600 25 ON - oversampling freq not a multiple of loop freq
20110429_114826 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.7 0.7 6 46 ON 3.2 H 600 25 ON - oversampling freq not a multiple of loop freq

Ripetiamo aggiustando lo script di acceptance che opra sceglie il disturbo da 625 quando ne ha bisogno
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_115521 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.7 0.8 6 41 ON 7.0 H 600 25 ON  
20110429_115600 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.7 0.8 6 43 ON 6.6 H 600 25 ON  
20110429_115638 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.7 0.8 6 41 ON 6.9 H 600 25 ON  
20110429_115752 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.7 0.8 6 41 ON 6.2 H 600 25 ON  
20110429_115830 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.7 0.8 6 40 ON 6.7 H 600 25 ON  
20110429_115908 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.7 0.8 6 41 ON 6.6 H 600 25 ON  

Mag 8.5 Seeing 0.4

Primo giro interrotto da uno skipframe che fa in stoploop. Un po' inspiegabile perche' lo specchio era rilassato dal seeing basso. Secondo giro liscio.

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_121802 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.5 3 259 ON 90.4 H 50 300 ON  
20110429_121921 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.5 3 258 ON 89.0 H 50 300 ON  
20110429_122040 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.5 3 255 ON 90.2 H 50 300 ON  
20110429_122319 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.5 3 249 ON 92.1 H 50 300 ON  
20110429_122439 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.5 3 259 ON 89.5 H 50 300 ON  
20110429_122558 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.5 3 258 ON 90.3 H 50 300 ON  

Mag 12.5 Seeing 0.4

Tutto liscio fino al secondo loop, fra il secondo e il terzo abbiamo skipframe e stoploop. Trovato il problema, e' il rotatore che viene messo in tracking erroneamente: il labmode del W ha qualche buco

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_124630 ??? 12.6 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.2 0.3 3 29 ON 70.0 H 100 150 ON  
20110429_124722 ??? 12.6 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.2 0.3 3 28 ON 72.0 H 100 150 ON  
20110429_124814 ??? 12.6 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.2 0.3 3 27 ON 74.4 H 100 150 ON  
20110429_125002 ??? 12.8 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.2 0.3 3 23 ON 74.8 H 100 150 ON  
20110429_125054 ??? 12.7 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.2 0.3 3 27 ON 71.3 H 100 150 ON  
20110429_125146 ??? 12.7 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.2 0.3 3 26 ON 70.5 H 100 150 ON  

Mag 14.5 Seeing 0.4

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_130225 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.6 6 45 ON 41.2 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_130308 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.6 6 47 ON 43.0 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_130351 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.6 6 46 ON 42.0 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_130523 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.6 6 43 ON 42.4 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_130606 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.6 6 49 ON 41.6 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_130649 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.6 6 45 ON 40.8 H 150 100 ON - oversampling freq not a multiple of loop freq

Mag 8.5 Seeing 0.6

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb n otes
20110429_143756 ??? 8.7 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 204 ON 89.3 H 50 300 ON  
20110429_143917 ??? 8.7 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 203 ON 91.8 H 50 300 ON  
20110429_144039 ??? 8.7 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 202 ON 88.1 H 50 300 ON  
20110429_144407 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 255 ON 90.8 H 50 300 ON  
20110429_144531 ??? 8.5 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.3 0.4 3 257 ON 89.2 H 50 300 ON  

Mag 12.5 Seeing 0.6"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_145734 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.3 0.3 3 32 ON 61.9 H 100 150 ON  
20110429_145826 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.3 0.3 3 31 ON 62.5 H 100 150 ON  
20110429_145918 ??? 12.6 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 801 0.3 0.3 3 29 ON 63.6 H 100 150 ON  
Blocco della BCU39 dopo il terzo punto...

Mag 14.5 Seeing 0.6"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_154424 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.7 6 44 ON 24.1 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_154507 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.7 6 47 ON 23.5 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_154550 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.7 6 43 ON 23.6 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_154721 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.7 6 40 ON 22.7 H 150 100 ON - oversampling freq not a multiple of loop freq
20110429_154804 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 312 0.6 0.7 6 41 ON 23.3 H 150 100 ON - oversampling freq not a multiple of loop freq

SEEING 1.5"

Mag 8.5 Seeing 1.5"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_155614 ??? 8.6 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 222 ON 67.9 H 200 75 ON - Warning: Selected IRTC dark +502h old!
20110429_155654 ??? 8.6 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 221 ON 67.1 H 100 75 ON  
20110429_155733 ??? 8.6 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 218 ON 68.6 H 100 75 ON  
20110429_155849 ??? 8.7 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 215 ON 64.7 H 100 75 ON  
20110429_155928 ??? 8.6 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 218 ON 67.1 H 100 75 ON  
20110429_160007 ??? 8.6 -2147483648 -2147483648 -1.00 -1.00 110422 1 400 990 0.4 0.5 3 218 ON 66.6 H 100 75 ON  
STRANISSIMI VALORI MOLTO ALTI Sono migliori di quelli del seeing 1.2", boh!!

Mag 12.5 Seeing 1.5"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_160605 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 0.9 3 40 ON 19.5 H 500 30 ON  
20110429_160637 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 0.9 3 40 ON 19.0 H 500 30 ON  
20110429_160710 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 0.9 3 41 ON 20.2 H 500 30 ON  
20110429_160819 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 0.9 3 39 ON 20.1 H 500 30 ON  
20110429_160851 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 0.9 3 41 ON 20.5 H 500 30 ON  
20110429_160923 ??? 12.4 -2147483648 -2147483648 -1.00 -1.00 154434 2 120 625 0.7 0.9 3 42 ON 21.2 H 500 30 ON  

Mag 14.5 Seeing 1.5"

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110429_161516 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.6 6 62 ON 2.0 H 600 25 ON  
20110429_161553 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.6 6 63 ON 1.7 H 600 25 ON  
20110429_161631 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.6 6 61 ON 1.8 H 600 25 ON  
20110429_161744 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.6 6 59 ON 1.7 H 600 25 ON  
20110429_161822 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.6 6 61 ON 1.7 H 600 25 ON  
20110429_161859 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.6 6 61 ON 1.8 H 600 25 ON  

NCP aberration procedure test

Primissimo test con la procedura di NCP aberration
S0_4_mag8_5 = ['20110429_121802','20110429_121921','20110429_122040','20110429_122319','20110429_122439','20110429_122558']
S0_4_mag12_5 = ['20110429_124630', '20110429_124722', '20110429_124814', '20110429_125002', '20110429_125054', '20110429_125146']
S0_4_mag14_5 = ['20110429_130225','20110429_130308','20110429_130351','20110429_130523','20110429_130606','20110429_130649']
flao2_S0_4 = obj_new('aodataset',[S0_4_mag8_5, S0_4_mag12_5, S0_4_mag14_5])
sr_S0_4 = flao2_S0_4->value('irtc.sr_se')
mag_S0_4 =flao2_S0_4->value('mag')


S0_6_mag8_5 = ['20110429_143756','20110429_143917','20110429_144039','20110429_144407','20110429_144531']
S0_6_mag12_5 = ['20110429_145734','20110429_145826','20110429_145918']
S0_6_mag14_5 = [ '20110429_154424','20110429_154507','20110429_154550','20110429_154721','20110429_154804']
flao2_S0_6 = obj_new('aodataset',[S0_6_mag8_5, S0_6_mag12_5, S0_6_mag14_5])
sr_S0_6 = flao2_S0_6->value('irtc.sr_se')
mag_S0_6 =flao2_S0_6->value('mag')

S0_8_mag9_5 = ['20110407_142718','20110407_143346','20110407_143428','20110407_144422','20110407_144459','20110407_144537','20110407_144659','20110407_144737','20110407_144815']
S0_8_mag12_5 = ['20110408_151120','20110408_151152','20110408_151225','20110408_151339','20110408_151412','20110408_151445','20110408_151608','20110408_151641','20110408_151714','20110408_151829','20110408_151902','20110408_151934']
S0_8_mag14_5 = ['20110411_114618','20110411_114649','20110411_114719','20110411_115056','20110411_115127','20110411_115232']
flao2_S0_8 = obj_new('aodataset',[S0_8_mag9_5,S0_8_mag12_5 ,S0_8_mag14_5])
sr_S0_8 = flao2_S0_8->value('irtc.sr_se')
mag_S0_8 =flao2_S0_8->value('mag')

S1_2_mag8_5 = ['20110428_172751', '20110428_172844','20110428_173033','20110428_173127','20110428_173220']
S1_2_mag12_5 = ['20110429_110246','20110429_110321','20110429_110509','20110429_110544','20110429_110618','20110429_111501']
S1_2_mag14_5 = ['20110429_115521','20110429_115600','20110429_115638','20110429_115752','20110429_115830','20110429_115908']
flao2_S1_2 = obj_new('aodataset',[S1_2_mag8_5, S1_2_mag12_5,S1_2_mag14_5])
sr_S1_2 = flao2_S1_2->value('irtc.sr_se')
mag_S1_2 =flao2_S1_2->value('mag')

S1_5_mag8_5 = ['20110429_155654','20110429_155733','20110429_155849','20110429_155928','20110429_160007']
S1_5_mag12_5 = ['20110429_160605', '20110429_160637', '20110429_160710', '20110429_160819', '20110429_160851', '20110429_160923']
S1_5_mag14_5 = ['20110429_161516','20110429_161553','20110429_161631','20110429_161744','20110429_161822','20110429_161859']
flao2_S1_5 = obj_new('aodataset',[S1_5_mag8_5, S1_5_mag12_5, S1_5_mag14_5])
sr_S1_5 = flao2_S1_5->value('irtc.sr_se')
mag_S1_5 =flao2_S1_5->value('mag')



; Simulation data (baseline and goal)
simul_table=fltarr(9,3)
simul_table[*,0]=[8.5,9.5,10.5,11.5,12.5,13.5,14.5,15.5,16.5]
simul_table[*,1]=[0.62,0.59,0.50,0.48,0.42,0.27,0.19,0.071,0.026]
simul_table[*,2]=[0.80,0.77,0.62,0.59,0.54,0.37,0.27,0.12,0.052]

; Simul SR vs Seeing
simul_mag = [8.5,10.5,12.5,14.5,16.5]
simul_nmag = n_elements(simul_mag)
; different seeings (v=15m/s)
simul_seeings = [0.6,0.8,1.0,1.2,1.5]
simul_nseeings = n_elements(simul_seeings)
                       ;                         -----> mag
flao_srH_v15 = [[87., 73., 67., 43., 15.], $
                           [80., 61., 52., 27., 05.], $
                           [68., 47., 35., 13., 02.], $
                           [44., 33., 20., 05., 01.], $
                           [21., 17.,  7.,  1., !VALUES.F_NAN]]/100.

plot,  simul_mag, flao_srH_v15[*,1], title='SR FLAO#2 @ Arcetri - Seeing 0.4" ; 0.8"; 1.2" ', xtitle='R+I mag',ytitle='SR value (H band)', charsize=2, yrange=[0,1]
oplot, simul_mag, flao_srH_v15[*,0], col='808000'x ; 0.6"
oplot, simul_mag, flao_srH_v15[*,1], col='00FF00'x ; 0.8"
oplot, simul_mag, flao_srH_v15[*,2], col='008080'x ; 1.0"
oplot, simul_mag, flao_srH_v15[*,3], col='0000FF'x ; 1.2"
oplot, simul_mag, flao_srH_v15[*,4], col='000080'x ; 1.5"
;oplot, simul_table[*,0], simul_table[*,2],col='00ff00'x
;oplot, simul_table[*,0], simul_table[*,1],col='0000ff'x
oplot, mag_S0_4, sr_S0_4, psym=4, col='FF0000'x
oplot, mag_S0_6, sr_S0_6, psym=4, col='808000'x
oplot, mag_S0_8, sr_S0_8, psym=4, col='00FF00'x
oplot, mag_S1_2, sr_S1_2, psym=4, col='0000FF'x
oplot, mag_S1_5, sr_S1_5, psym=4, col='000080'x


;oplot, mag_b4, sr_b4, psym=2 

-- AlfioPuglisi - 29 Apr 2011

  • Screenshot-adoptadsecsx.png:
    Screenshot-adoptadsecsx.png
Topic revision: r5 - 29 Apr 2011, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback