AOS test

AOS upgraded to FLAO/LBTI version. After updating the AOArbitrator too, everything works as before.

Instrument must be set from IIF GUI, but the GUI crashes on startup. The instrument can be set using the "aostest" command line tool on tcstest instead
Acquire ref works with diffraction limited image on ccd47 - seeing simulation still to be done.

Complete test: power on/off WFS and Adsec, set/rest adsec, preset, checkref, acquireref, start, pause, resume, stop.

Some minor bugs found and put into bugzilla.

Ron test di oggi:

OFF: stages.

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.09 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.88 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.49 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.18 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.04 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.82 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.50 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.50 (estimated)

OFF: FW, ADC, REROT, stages, CUBE.

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.86 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.53 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.33 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.34 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.50 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.70 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.69 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.73 (estimated)

-- AlfioPuglisi - 12 Apr 2011