Accendiamo il tutto dopo un po' di casino col VNC della camera infrarossa.

OFF: ADC, REROT, stages, CUBE

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.01 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.69 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.11 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.19 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.36 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.48 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.51 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.54 (estimated)

OFF: ADC, REROT, stages, CUBE, FW

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.87 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.52 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.46 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.46 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.64 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.86 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.84 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.85 (estimated)



Bin 3 mag 14.5

Aux FW pos=2 (T=1%) Autogain fuoco IRTC sbagliato
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_113123 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 44 ON 16.6 H 300 50 ON fuoco IRTC sbagliato
20110411_113154 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 45 ON 16.0 H 300 50 ON SR may be in error fuoco IRTC sbagliato
20110411_113225 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 46 ON 16.5 H 300 50 ON fuoco IRTC sbagliato
20110411_113332 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 40 ON 14.5 H 300 50 ON SR may be in error fuoco IRTC sbagliato

| 20110411_113403 | ??? | 14.6 | -2147483648 | -2147483648 | -1.00 -1.00 | 171410 | 3 | 66 | 200 | 0.8 1.4 | 6 | 43 | ON
15.3 H 300 50 ON fuoco IRTC sbagliato
20110411_113433 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 44 ON 16.5 H 300 50 ON fuoco IRTC sbagliato
LOAD SHAPE ripetiamo le misure

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_113754 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 46 ON 18.8 H 300 50 ON fuoco IRTC sbagliato
20110411_113825 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 45 ON 19.5 H 300 50 ON fuoco IRTC sbagliato
20110411_113855 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 45 ON 19.1 H 300 50 ON fuoco IRTC sbagliato
20110411_114002 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 40 ON 16.6 H 300 50 ON fuoco IRTC sbagliato
20110411_114032 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 43 ON 18.4 H 300 50 ON fuoco IRTC sbagliato
20110411_114102 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.4 6 42 ON 17.6 H 300 50 ON fuoco IRTC sbagliato
Proviamo ad abbassare il gain dei modi alti, cosi' come l'aveca trovato ven, 0.11

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_114206 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 46 ON 19.7 H 300 50 ON fuoco IRTC sbagliato
20110411_114236 ??? 14.4 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 49 ON 19.2 H 300 50 ON fuoco IRTC sbagliato
20110411_114307 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 46 ON 18.4 H 300 50 ON fuoco IRTC sbagliato
20110411_114414 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 41 ON 17.8 H 300 50 ON SR may be in error fuoco IRTC sbagliato
20110411_114444 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 39 ON 18.3 H 300 50 ON fuoco IRTC sbagliato
20110411_114514 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 47 ON 18.8 H 300 50 ON fuoco IRTC sbagliato
Ci accorgiamo che avevamo il fuoco IRTC sbagliato di 1 mm per il la ND di pos2, LO CORREGGIAMO

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_114618 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 45 ON 19.8 H 300 50 ON  
20110411_114649 ??? 14.4 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 48 ON 21.7 H 300 50 ON  
20110411_114719 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 45 ON 21.3 H 300 50 ON  
20110411_114824 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 42 ON 19.8 H 300 50 ON SR may be in error
20110411_114855 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 45 ON 20.6 H 300 50 ON SR may be in error
20110411_114925 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 44 ON 20.0 H 300 50 ON SR may be in error
20110411_115026 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 46 ON 21.0 H 300 50 ON SR may be in error
20110411_115056 ??? 14.4 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 48 ON 20.5 H 300 50 ON  
20110411_115127 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 45 ON 21.2 H 300 50 ON  
20110411_115232 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 43 ON 19.4 H 300 50 ON  
20110411_115303 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 44 ON 20.5 H 300 50 ON SR may be in error
20110411_115333 ??? 14.5 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 47 ON 21.1 H 300 50 ON  

Bin 3 Mag 14.8

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_120301 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 36 ON 15.1 H 400 50 ON  
20110411_120332 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 31 ON 14.9 H 400 50 ON  
20110411_120403 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 38 ON 14.7 H 400 50 ON  
20110411_120513 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 32 ON 12.9 H 400 50 ON  
20110411_120544 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 33 ON 13.8 H 400 50 ON  
20110411_120615 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.1 6 35 ON 13.3 H 400 50 ON  
Abbassiamo ancora la lampada

Bin 3 Mag 15.0

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_120954 ??? 15.0 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.0 6 28 ON 7.6 H 400 50 ON  
20110411_121025 ??? 15.1 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.0 6 26 ON 7.2 H 400 50 ON SR may be in error
20110411_121056 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.0 6 30 ON 7.6 H 400 50 ON  
20110411_121207 ??? 15.3 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.0 6 23 ON 5.9 H 400 50 ON SR may be in error
20110411_121237 ??? 15.0 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.0 6 30 ON 7.3 H 400 50 ON  
20110411_121308 ??? 15.1 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 200 0.8 1.0 6 27 ON 7.0 H 400 50 ON SR may be in error

Bin 3 mag 12.8 @625Hz

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_122757 ??? 12.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 625 0.5 0.6 6 70 ON 35.9 H 150 100 ON  
20110411_122830 ??? 12.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 625 0.5 0.6 6 68 ON 35.4 H 150 100 ON  
20110411_122903 ??? 12.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 625 0.5 0.6 6 68 ON 36.0 H 150 100 ON  
20110411_123017 ??? 12.9 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 625 0.5 0.6 6 66 ON 36.2 H 150 100 ON  
20110411_123050 ??? 12.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 625 0.5 0.6 6 68 ON 36.0 H 150 100 ON  
20110411_123123 ??? 12.8 -2147483648 -2147483648 -1.00 -1.00 171410 3 66 625 0.5 0.6 6 67 ON 36.4 H 150 100 ON  
INTMAT con disturbo atmosferico

Prendiamo i residui (comandi) del loop 20110411_115232

=======================================

DISTURBOIM 20110411_124510
Modal basis = KL_20110311
type = Push-pull
Frames per mov = 3
Cycle = 9
#Modes = 66
Amp = 600_optimized-6l
Note = sommato ai comandi residui del loop 20110411_115232

=======================================

INTMATACQ # 20110411_124522
Pup = # solite
Dist = # 20110411_124510
Iteration = # 4
Loop speed = # 626 Hz
bin = # 3
Modul = # 6lambda/D
Rec.Close loop = # 171410
Gain vector close loop = #0.05

=======================================

REC # 20110411_124522
Modal Basis: KL_20110311
intmatacq = # 20110411_124522
n modes = # 66
TT removal = ON
cutted = # 0
Note: ricostryttore fatto in condizioni di loop chiuso cioe muovendo lo psecchio con i comandi residui registrati durante un CL a BIN3

=======================================

test del nuovo rec (con turbolenza residua) MAG 14.7

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes   20110411_125849 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 41 ON 17.3 H 300 50 ON     20110411_125920 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 38 ON 16.9 H 300 50 ON     20110411_125951 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 42 ON 16.2 H 300 50 ON     20110411_130058 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 35 ON 14.8 H 300 50 ON     20110411_130128 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 39 ON 15.7 H 300 50 ON     20110411_130200 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 40 ON 16.2 H 300 50 ON  

Riapplicato il loadshape

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_130311 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 42 ON 18.0 H 300 50 ON  
20110411_130341 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 40 ON 18.5 H 300 50 ON  
20110411_130411 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 41 ON 18.6 H 300 50 ON  
20110411_130517 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 40 ON 16.8 H 300 50 ON  
20110411_130547 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 40 ON 18.2 H 300 50 ON  
20110411_130617 ??? 14.6 -2147483648 -2147483648 -1.00 -1.00 124642 3 66 200 0.8 1.2 6 41 ON 16.8 H 300 50 ON SR may be in error

PRANZO


Flattening specchio a BIN1 con 500 modi

20110411_500modes.sav

ci puo' essere del defocus: slitta non in posizione nominale


Intmat BIN 4

=======================================

DISTURBOIM 20110411_150416
Modal basis = KL_20110311
type = Push-pull
Frames per mov = 3
Cycle = 16
#Modes = 36
Amp = 600_optimized_6l
Note =

=======================================

INTMATACQ 20110411_152812
Pup = 20110203-105150
Dist = 20110411_150416
Iteration = 4
Loop speed = 701
bin = 4
Modul = 6
Rec.Close loop = None
Gain vector close loop = None

=======================================

REC 20110411_152928
Modal Basis: KL_20110311
intmatacq = 20110411_152812
n modes = 36
TT removal = ON
cutted = 0
Note: ...............

=======================================

BIN4 mag 13.7

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_153629 ??? 13.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.2 6 185 ON 20.7 H 300 50 ON  
20110411_153700 ??? 13.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.2 6 186 ON 21.1 H 300 50 ON  
20110411_153730 ??? 13.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.2 6 184 ON 21.2 H 300 50 ON  
20110411_153836 ??? 13.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.2 6 176 ON 20.5 H 300 50 ON  
20110411_153907 ??? 13.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.2 6 180 ON 0.0 H 300 50 ON Background correction failed
20110411_153938 ??? 13.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.2 1.2 6 185 ON 21.0 H 300 50 ON  
Prima di abbassare la lampada facciamo un RON test perché il display ci sembra un po' rumoroso

OFF: FW, ADC, REROT, stages, LAMP, CUBE

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.79 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.57 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.93 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.97 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.18 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.39 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.52 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.52 (estimated)

OFF: FW, ADC, REROT, stages, LAMP, CUBE, BOXFANS, LJ FANS

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.67 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.47 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.76 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.74 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.95 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.22 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.30 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.31 (estimated)

Il rumore e' cresciuto rispetto a stamattina senza alcun apparente motivo.

BIN4 mag 15.5 RON @5.3

Mettiamo la Aux FW in POS=3 (nominalmente 0.1%, ma dalle misure sembra 0.4%)

Autofocus su IRTC zstages = 53.89mm

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_160817 ??? 15.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 31 ON 8.1 H 400 50 ON  
20110411_160848 ??? 15.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 38 ON 7.3 H 400 50 ON  
20110411_160919 ??? 15.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 27 ON 7.6 H 400 50 ON  
20110411_161029 ??? 15.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 34 ON 6.6 H 400 50 ON  
20110411_161100 ??? 15.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 31 ON 7.0 H 400 50 ON  
20110411_161132 ??? 15.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 31 ON 7.2 H 400 50 ON  
20110411_161244 ??? 15.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 34 ON 7.4 H 400 50 ON  
20110411_161315 ??? 15.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 33 ON 7.2 H 400 50 ON  
20110411_161345 ??? 15.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 29 ON 7.0 H 400 50 ON  
20110411_161455 ??? 15.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 30 ON 6.0 H 400 50 ON  
20110411_161525 ??? 15.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 29 ON 6.3 H 400 50 ON  
20110411_161554 ??? 15.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 0.7 0.7 6 28 ON 6.6 H 400 50 ON  

BIN4 mag 16.5 RON 5.4e-

Mettiamo Aux FW POS=5; teoricamente T=0.001% in pratica meno forte Zstage = 53.60
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_164626 ??? 16.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 27 ON 4.0 H 400 50 ON  
20110411_164715 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 29 ON 4.4 H 400 50 ON  
20110411_164806 ??? 16.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 26 ON 3.9 H 400 50 ON  
20110411_164937 ??? 16.5 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 26 ON 3.6 H 400 50 ON  
20110411_165027 ??? 16.6 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 24 ON 3.7 H 400 50 ON  
20110411_165118 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 29 ON 0.0 H 400 50 ON Background correction failed
20110411_165214 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 28 ON 3.6 H 400 50 ON  
20110411_165304 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 29 ON 3.8 H 400 50 ON  
20110411_165355 ??? 16.3 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 32 ON 3.7 H 400 50 ON  
20110411_165525 ??? 16.1 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 38 ON 3.2 H 400 50 ON  
20110411_165619 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 31 ON 3.4 H 400 50 ON  
20110411_165710 ??? 16.4 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 100 0.7 1.2 6 30 ON 3.6 H 400 50 ON
Proviamo ad abbassare ancora la lampada andando a 20 phot/subap ma fallisce l'auto-gain facendo andare in skipframe lo specchio.

BIN4 mag 14.7

torniamo con aux FW su pos=3

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_170743 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 68 ON 12.1 H 400 50 ON  
20110411_170815 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 71 ON 8.3 H 400 50 ON  
20110411_170858 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 57 ON 10.4 H 400 50 ON SR may be in error
20110411_171710 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 62 ON 11.9 H 400 50 ON  
20110411_171741 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 63 ON 12.1 H 400 50 ON  
20110411_171813 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 61 ON 11.6 H 400 50 ON  
20110411_171924 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 60 ON 11.8 H 400 50 ON  
20110411_171955 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 60 ON 11.4 H 400 50 ON  
20110411_172027 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 200 1.0 1.0 6 63 ON 11.3 H 400 50 ON  

Proviamo a fare lo stesso punto a 400Hz
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_172319 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 401 0.7 0.7 6 34 ON 6.9 H 400 50 ON  
20110411_172349 ??? 14.7 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 401 0.7 0.7 6 34 ON 6.4 H 400 50 ON  
20110411_172419 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 401 0.7 0.7 6 33 ON 6.4 H 400 50 ON  
20110411_172528 ??? 14.9 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 401 0.7 0.7 6 31 ON 5.6 H 400 50 ON  
20110411_172558 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 401 0.7 0.7 6 33 ON 6.3 H 400 50 ON  
20110411_172628 ??? 14.8 -2147483648 -2147483648 -1.00 -1.00 152928 4 36 401 0.7 0.7 6 33 ON 6.7 H 400 50 ON SR may be in error

BIN2 mag13 @500Hz

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110411_174104 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 32 ON 45.4 H 100 150 ON  
20110411_174146 ??? 12.9 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 33 ON 44.2 H 100 150 ON  
20110411_174233 ??? 13.1 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 29 ON 44.6 H 100 150 ON  
20110411_174403 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 32 ON 44.9 H 100 150 ON  
20110411_174445 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 31 ON 45.4 H 100 150 ON  
20110411_174526 ??? 12.9 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 33 ON 45.3 H 100 150 ON  
20110411_174658 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 31 ON 46.4 H 100 150 ON  
20110411_174740 ??? 12.9 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 33 ON 47.2 H 100 150 ON  
20110411_174821 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 31 ON 47.2 H 100 150 ON  
20110411_174952 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 31 ON 45.4 H 100 150 ON  
20110411_175033 ??? 13.0 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 31 ON 45.3 H 100 150 ON  
20110411_175115 ??? 12.9 -2147483648 -2147483648 -1.00 -1.00 171951 2 153 500 0.5 0.7 3 33 ON 46.6 H 100 150 ON  

-- AlfioPuglisi - 11 Apr 2011

bin1_mag9_6 = ['20110407_142718','20110407_143346','20110407_143428','20110407_144422','20110407_144459','20110407_144537','20110407_144659','20110407_144737','20110407_144815']  
bin1_mag10_1= ['20110407_150114','20110407_150603','20110407_150648','20110407_150814','20110407_150953']
bin1_mag10_7= ['20110408_161948','20110408_162026','20110408_162103','20110408_162225','20110408_162302','20110408_162340','20110408_162503','20110408_162540','20110408_162619','20110408_162741','20110408_162818','20110408_162856']
bin1_mag11_2= ['20110408_163954','20110408_164036','20110408_164118','20110408_164250','20110408_164334','20110408_164416','20110408_164533','20110408_164615','20110408_164657','20110408_164828','20110408_164910','20110408_164951']
flao2_bin1 = obj_new('aodataset',[bin1_mag9_6,bin1_mag10_1,bin1_mag10_7, bin1_mag11_2])
sr_b1 = flao2_bin1->value('irtc.sr_se')
mag_b1 = flao2_bin1->value('mag')



bin2_mag10_7= ['20110407_173951','20110407_174018','20110407_174150','20110407_174220']
bin2_mag11_0= ['20110407_174626','20110407_174653','20110407_174758','20110407_174825','20110407_174857']
bin2_mag11_8= ['20110407_180943','20110407_181011','20110407_181038','20110407_181143','20110407_181238','20110407_181358','20110407_181428','20110407_181532','20110407_181600','20110407_181627']
bin2_mag12_5= ['20110408_151120','20110408_151152','20110408_151225','20110408_151339','20110408_151412','20110408_151445','20110408_151608','20110408_151641','20110408_151714','20110408_151829','20110408_151902','20110408_151934']
bin2_mag13_0=['20110411_174104','20110411_174146','20110411_174233','20110411_174403','20110411_174445','20110411_174526','20110411_174658','20110411_174740','20110411_174821','20110411_174952','20110411_175033','20110411_175115']
bin2_mag13_5= ['20110408_155152','20110408_155219','20110408_155248','20110408_155352','20110408_155419','20110408_155447','20110408_155559','20110408_155627','20110408_155655','20110408_155800','20110408_155827']
flao2_bin2 = obj_new('aodataset',[bin2_mag10_7,bin2_mag11_0,bin2_mag11_8, bin2_mag12_5, bin2_mag13_0, bin2_mag13_5])
sr_b2 = flao2_bin2->value('irtc.sr_se')
mag_b2 = flao2_bin2->value('mag')

bin3_mag12_8= ['20110411_122757','20110411_122830','20110411_122903','20110411_123017','20110411_123050','20110411_123123']
bin3_mag13_5= ['20110408_174641','20110408_174809','20110408_175424','20110408_175452']
bin3_mag14_5= ['20110411_114618','20110411_114649','20110411_114719','20110411_115056','20110411_115127','20110411_115232']
bin3_mag14_9= ['20110411_120301','20110411_120332','20110411_120403','20110411_120513','20110411_120544','20110411_120615']
bin3_mag15_1= ['20110411_120954','20110411_121056','20110411_121237']
flao2_bin3=obj_new('aodataset',[bin3_mag12_8, bin3_mag13_5, bin3_mag14_5, bin3_mag14_9, bin3_mag15_1])
sr_b3 = flao2_bin3->value('irtc.sr_se')
mag_b3 = flao2_bin3->value('mag')

bin4_mag13_7= ['20110411_153938','20110411_153700','20110411_153730','20110411_153836','20110411_153938']
bin4_mag14_7= ['20110411_170743','20110411_170815','20110411_171710','20110411_171741','20110411_171813','20110411_171924','20110411_171955','20110411_172027']
bin4_mag15_5= ['20110411_160817','20110411_160848','20110411_160919','20110411_161029','20110411_161100','20110411_161132','20110411_161244','20110411_161315','20110411_161345','20110411_161455',']20110411_161525','20110411_161554']
bin4_mag16_5=['20110411_164626','20110411_164715','20110411_164806','20110411_164937','20110411_165027','20110411_165214','20110411_165304','20110411_165355','20110411_165525','20110411_165619','20110411_165710']
flao2_bin4=obj_new('aodataset',[bin4_mag13_7, bin4_mag14_7, bin4_mag15_5, bin4_mag16_5])
sr_b4 = flao2_bin4->value('irtc.sr_se')
mag_b4 = flao2_bin4->value('mag')

; Simulation data (baseline and goal)
simul_table=fltarr(9,3)
simul_table[*,0]=[8.5,9.5,10.5,11.5,12.5,13.5,14.5,15.5,16.5]
simul_table[*,1]=[0.62,0.59,0.50,0.48,0.42,0.27,0.19,0.071,0.026]
simul_table[*,2]=[0.80,0.77,0.62,0.59,0.54,0.37,0.27,0.12,0.052]

plot, simul_table[*,0], simul_table[*,2], title='SR FLAO#2 @ tower testbench in April 2011', xtitle='R+I mag',ytitle='SR value (H band)', charsize=2, yrange=[0,1]
oplot, simul_table[*,0], simul_table[*,2],col='00ff00'x
oplot, simul_table[*,0], simul_table[*,1],col='0000ff'x
oplot, mag_b1, sr_b1, psym=4
oplot, mag_b2, sr_b2, psym=6
oplot, mag_b3, sr_b3, psym=5 
oplot, mag_b4, sr_b4, psym=2 
  • plottone_20110411.png: