Cambiato cavo di alimentazione del PW lineare: quello a bordo del W mancava del cavo di terra che collega la terra del 110V allo chassis dell'AGW

ON: ccd39, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.70 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.71 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.72 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.77 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.51 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.49 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 7.02 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.76 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.77 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.78 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.03 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.86 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.44 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.64 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.71 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.61 (estimated)

Proviamo a staccare il cavo di terra dallo chassis del lineare.

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.74 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.77 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.67 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.64 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.78 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.02 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.37 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.44 (estimated)

Due elettroni meglio di ieri!


Rimettiamo l'alimentazione nella situazione di ieri (cavo interno all'AGW)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.68 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.87 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.31 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.29 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.73 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.36 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 7.33 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.08 (estimated)

Aspettiamo qualche minuto e riproviamo

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.75 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.93 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.52 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.39 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.43 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.87 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.90 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.71 (estimated)

Ricollegata la terra (senza modificare l'alimentazione)

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.71 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.81 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.99 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.00 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.50 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.75 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.94 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.00 (estimated)


Rivisti tutti i collegamenti a 110V:

- lineare collegato direttamente alla morsettiera dell'AGW
- eliminata derivazione prima della scatola39, accorciati un po' i cavi

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.57 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.83 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.35 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.28 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.75 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.18 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 7.26 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.20 (estimated)

Timestamp: Fri Apr 1 11:54:09

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.60 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.82 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.13 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.07 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.37 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.84 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.94 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.85 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.65 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.68 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.23 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.27 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.72 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.98 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.97 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.01 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.52 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.56 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.06 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.91 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.23 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.62 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.56 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.54 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.60 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.67 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.52 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.14 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.44 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.52 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.61 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.69 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.56 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.48 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.05 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.02 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.18 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.26 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.21 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.28 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.57 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.56 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.69 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.69 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.87 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.04 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.14 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.03 (estimated)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.61 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.65 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.18 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.30 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.54 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.93 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.92 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.82 (estimated)



Isolato il PW lineare dall'AGW - pare che sia ancora a terra passando da qualche altra parte (cavi?)

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.68 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.76 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.22 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.08 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.40 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.68 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.88 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.96 (estimated)

Stacchiamo il cavetto di terra che arriva al PW lineare.

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.63 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.77 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.75 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.89 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.19 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.55 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.51 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.56 (estimated)

Rimettiamo il cavetto per vedere se era lui oppure l'AGW che si stava termalizzando

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.61 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.82 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.84 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.65 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.82 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.69 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.48 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.50 (estimated)

Ne concludiamo che isolare il PW lineare non serve a molto.

Sconesso l'ADC dalla stella di massa

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.74 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.87 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.94 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.85 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.21 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.17 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.60 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.17 (estimated)

Scollegata la stella di massa dallo shield che la portava via. Adesso la board non e' piu' a massa con l'AGW ma ci sono 50 Ohm

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.34 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.87 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.45 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.50 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.44 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.64 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.65 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.02 (estimated)

Testa e board staccati dalla terra. Shield dei motori staccati dalla board.

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.48 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.84 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.79 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.82 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.40 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.86 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.83 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.35 (estimated)

Testa del ccd messa a massa direttamente con il lineare. Cavi di massa dai motori verso la stella scollegati.

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.47 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.77 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.79 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.59 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.23 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.91 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.61 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.09 (estimated)

Ripristinata la stella di massa dei motori. La testa e' collegata alla stella e la stella e' collegata allo chassis nero dell'AGW.

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.43 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.82 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.75 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.31 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.11 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.86 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.66 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.69 (estimated)


Testa ccd39 messa a massa sull'anello dell'AGW. Stella di massa completamente scollegata.

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.83 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.52 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.14 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.19 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.26 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.38 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.42 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.44 (estimated)

Rivestiti i cavi dell'ADC con carta argentata

ON: ccd39, ventole ccd, ventole scatole, bayside stages (no motori), ccd47, FW, ADC, rerotatore

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.85 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.83 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.68 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.53 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.30 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.61 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.99 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.36 (estimated)


Appunto: prima c'erano 50-60 ohm tra AGW frame e scatola39/ccd39.
Ora che e' spento invece non c'e' piu' questa resistenza. Tutto e' in corto (testa, scatola, frame, stella). La misura di resistenza da acceso viene evidentemente da delle correnti che scorrono. Questo spiega anche le misure di ieri quando si trovavano resistenze negative.

Verifica del collegamento tra lo 0V del pw 24V e lo chassis:

staccando fioriera, terza scatola, e connettore board (cioe' tutti i devices serviti dal 24V), lo 0V del PW 24V e lo chassis dell'AGW sono in corto (e anche la testa del ccd39). Quindi il corto e' dentro la scatola 39.

Facciamo le seguenti scoperte:

- staccando il cavo 110V tra le due scatole, il ts8sx rimane spento come previsto. Quindi il ts8dx e' alimentato da un PW dentro la scatola di destra (47)
- ma se in questa condizione accendo il rirotatore (con cui non si puo' comunicare senza ts8), il motore viene alimentato (diventa duro a girarlo a mano). Quindi il rirotatore e' alimentato da un 12V nella scatola di sinistra (39)
- nella scatola 39 ci sono due diversi 12V: quello di servizio per il ts8, e quello che viene dal lineare. Stacchiamo il cavo 110V che va alla scatola 39, in modo da spegnere il 12V di servizio. In questo modo perdiamo tutte le comunicazioni, ma il lineare rimane acceso, la 39 anche, e lo switch di accensione del rotatore rimane attivo. Risultato: il rotatore diventa morbido, cioe' il controller si spenge. Quindi il rotatore e' alimentato dal 12V di servizio.


Intervento 24V

Apriamo la scatola e troviamo un cavetto che unisce il ground del PW 24V e di quello 12V. La ragione di questo collegamento e' sconosciuta, ma supponiamo che sia stato un tentativo di "stella di massa".

Scolleghiamo il cavetto (lasciando il suo connettore vicino al PW, in modo che si possa ricollegarlo in caso di problemi), e rimontiamo la scatola.

Risultato: mezzo W non funziona piu'. Sono morte: le filterwheel (24V), i motori ADC (24V), il rirotatore (12V), e il cube rotator (24V). Il cube stage (12V) funziona ancora. I due ts8 e le due bcu funzionano.

Ricolleghiamo il cavetto e tutto funziona nuovamente (fiuuu....)

Deduzione: i devices a 24V (e forse anche qualcos'altro visto che pure il rirotatore era morto) hanno un cablaggio completamente sbagliato, e la loro alimentazione si richiude passando per chissa' dove (chassis?), invece di passare per i fili che arrivano al GND 24V.

-- AlfioPuglisi - 01 Apr 2011
Topic revision: r3 - 04 Apr 2011, AlfioPuglisi