You are here: Foswiki>FLAO Web>TowerLab>LabTest20110328 (29 Mar 2011, EnricoPinna)Edit Attach
Cerchiam o

Rerot =133deg

Dall'IDL controller abbiamo che rotX muove nell'asse Y del CCD39 ae anche coirrispendete alle slitte Y

Muovendo rotX aumentandolo si vede il filo piu' dritto e ad un angolo un po' di verso rispetto a quando si diminuisce RotX

Off: LAMP e MOTORI SLITTE (solo bin3 e bin4)

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.45 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 7.85 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 11.11 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 10.56 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 11.48 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 12.22 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 12.30 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 12.77 (estimated)

Off: LAMP e MOTORI SLITTE

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.22 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 8.20 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 10.31 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 10.13 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 11.94 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 11.96 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 12.23 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 12.07 (estimated)

Off: LAMP e MOTORI SLITTE e CUBO

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.78 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.87 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.37 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 7.03 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 7.16 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 8.51 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 8.14 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.99 (estimated)

Off: LAMP e MOTORI SLITTE e CUBO e FW

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.68 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.84 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.31 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.86 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.47 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.42 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 7.00 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.85 (estimated)

Off: LAMP e MOTORI SLITTE e CUBO e FW e ADC

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.60 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.60 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.89 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.97 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.17 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.52 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.45 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.39 (estimated)

Off: LAMP e MOTORI SLITTE e CUBO e FW e ADC e REROT

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.68 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.53 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.37 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.40 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.53 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.71 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.70 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.77 (estimated)

Off: LAMP e MOTORI SLITTE e CUBO e FW e ADC e REROT e BOXFANS

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.60 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.45 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.24 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.24 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.34 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.50 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.44 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.53 (estimated)

RON TEST tradizionale

Off: LAMP e MOTORI SLITTE e CUBO e FW e ADC e REROT e BOXFANS

ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 18.14 - RON (e-): 9.07 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.16 - RON (e-): 6.58 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.63 - RON (e-): 4.32 (estimated)

Off: LAMP e MOTORI SLITTE e CUBO e BOXFANS

ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 18.23 - RON (e-): 9.11 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.04 - RON (e-): 7.02 (estimated)
Speed: 400 kpix/sec - RON (ADU): 14.67 - RON (e-): 7.34 (estimated)


A caccia del filino

Mettiamo il barrone e il laser del FISBA con la fibra grossa.

Oltre a un sacco di speckle si vede anche il filo sul ritorno dal tubone.

Si vede gia' sul fascio di ritorno fra la finestra del tubone e lo specchio del barrone -> il filo e' nella torre, dalle dismensioni e' probanilmente nel bicchierino.

Rimettiamo la fibra piccola con la lampada e proviamo a spostare il cubo in traslazione per evitare il filo in qualche modo.

Con lo stage a -5.9mm semrba che ci sia una posizione dove non si vede il filo

Nuovo board setup: torre18
Nuove posizioni esapodo: print, hexmoveto([585.053, 1813.99, 1850, 66.997, 152.997, 0.0399213])

Schedula veloce:

mar 29: ricalibrazione bin1
mer 30: cloop faint end

Se le cose vanno bene:

gio 31/ven 01: completamento tabella

Se invece abbiamo problemi:

ven 01:
gio 31
ven 01: carbone?
lun : carbone?

-- EnricoPinna - 28 Mar 2011

Off: LAMP e MOTORI SLITTE e CUBO e FW e ADC e REROT
Topic revision: r4 - 29 Mar 2011, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.