AGW rimontato sulla torre

Lucifer window montata.
IRTC no.

Functional test completo OK.

Tutto acceso tranne LAMP e MOTORI SLITTE

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.04 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 8.13 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 11.84 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 10.76 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 12.95 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 13.78 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 13.70 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 13.67 (estimated)

Si nota un forte rombo anche a bin1, dove solitamente il rumore non ha strutture. Numericamente non sembra influire molto....

Tutto acceso tranne LAMP, MOTORI SLITTE e CUBO.

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.72 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 7.16 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 8.14 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 8.33 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 9.12 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 9.98 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 10.17 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 9.93 (estimated)

A bin1 non c'e' piu' il rombo.

Off: FW, ADC, REROT, CUBO, LAMP, MOTORI SLITTE.

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.55 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.47 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.28 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.28 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.43 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.57 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.52 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.65 (estimated)

Timestamp: 2011 03 25 14 47 00

Off: FW, ADC, CUBO, LAMP, MOTORI SLITTE.

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.51 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.56 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.07 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.09 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.29 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.53 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.52 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.43 (estimated)

Off: FW, CUBO, LAMP, MOTORI SLITTE.

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.54 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.92 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.70 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.45 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 7.53 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.85 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 8.04 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.77 (estimated)

LUCE VERSO IL SECONDARIO

Riaccendiamo adsec ed esapodo e applichiamo il boardsetup "board16".

Si vedono le pupille ma c'e' una riga nera nel mezzo che potrebbe essere:

1) un filo fuori dalla board (con la luce diretta nel W non si vede)
2) il bordo del cubo come era già successo

Spostando l'esapodo (tilt) e inseguendo con la board il filo si muove leggermente.

Questo è lo screenshot nell'ultima posizione presa da TowerConfigurations (31/01/2011)

print,hexmoveto([385, 2014, 1850.0, 52.5, 137.5, 0])
  • ccd39_filo.png:
    ccd39_filo.png

Off: FW, ADC, CUBO, LAMP, MOTORI SLITTE.

Rimuoviamo la coma e troviamo un'altra posizione esapodo:

print,hexmoveto([585, 1814, 1850.0, 64, 157, 0])

Slitte: X=-83.24 Y=39.38 Z=41.74
Topic revision: r1 - 25 Mar 2011, AlfioPuglisi