10:13
-18C (board temp)   OK OK OK OK OK OK X - OK
Y - KO
OK OK OK NA OK Le pupille appaiono spostate in x verso sinistra sul CCD39 (invasione di quadrante). Il rirotatore muove la pupill di 6-7 pix p2v in X.
14:50 -4 C board temp   OK OK OK OK OK OK X-OK Y-KO OK OK OK NA OK  


Tutto acceso OFF = cube, lamp 10:37 ------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.48 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.83 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.60 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.62 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 7.10 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 7.64 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 7.71 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 7.61 (estimated)

Tutto spento OFF: cube , lamp, ADC REROT, FW, boxfans 10:39 ------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.28 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.29 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.43 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.45 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.52 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.52 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.58 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.69 (estimated)

Tutto spento tranne rirot
OFF: cube , lamp, ADC, FW, boxfans
10:42
------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.27 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.45 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.60 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.68 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.85 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.01 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.96 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.06 (estimated)

Tutto spento tranne rirot
OFF: cube , lamp, ADC, FW, boxfans
12:17
------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.30 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.46 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 5.19 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 5.26 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.39 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.62 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.67 (estimated)

Tutto spento
OFF: cube , lamp, boxfans
12:19
------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.39 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.33 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.43 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.50 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.60 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.70 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.66 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.77 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.65 (estimated)

Tutto acceso
OFF: cube , lamp, ADC, Rirot, FW, boxfans
12:19
------------------------------ SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.43 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.63 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 6.41 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 6.43 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 6.27 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 6.49 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 6.55 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 6.47 (estimated)

Muoviamo il cube rotator per togliere il vignettamento delle pupille. Ora il rirotatore fa ca 3 pix p2v ma le pupille sono ancora scentrate rispetto ai quadranti, questo probabilmente perche' la OLC ha un asse fuori uso EnricoPinna - 10 Mar 2011 2011/3/10 12:21: Cambiato setpoint (12 C)

2011/3/10 14:50:10
Temp: ~ -5 C

Board Setup and single devices: Tutto OK (vedere tabellina)

OFF: Cube, Lamp, Stages disabled

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.19 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.26 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.86 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.98 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.99 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.14 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.21 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.18 (estimated)

OFF: cube , lamp, ADC, FW, boxfans

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.16 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.21 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 3.99 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.01 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.02 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.02 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.05 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.08 (estimated)

pup-23.png

Topic attachments
I Attachment Action Size Date Who Comment
pup-20.bmpbmp pup-20.bmp manage 1 MB 10 Mar 2011 - 11:38 UnknownUser  
pup-23.pngpng pup-23.png manage 19 K 11 Mar 2011 - 09:53 UnknownUser  
This topic: FLAO > WebHome > TowerLab > LabTest20110310
Topic revision: 11 Mar 2011, JavierArgomedo
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback