W#2 al telescopio RON TEST

Tutto acceso tranne slitte frenate e LJ fans


SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.70 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.12 (estimated)
Bin: 2 - Framerate 625 Hz ( 800 kpix/sec) - RON (e-): 6.07 (estimated)
Bin: 2 - Framerate 400 Hz ( 800 kpix/sec) - RON (e-): 6.09 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.41 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.41 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.50 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.56 (estimated)


SUMMARY ------------------------------------

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.41 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 5.99 (estimated)
Bin: 2 - Framerate 625 Hz ( 800 kpix/sec) - RON (e-): 5.97 (estimated)
Bin: 2 - Framerate 400 Hz ( 800 kpix/sec) - RON (e-): 6.00 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.39 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.39 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.51 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.63 (estimated)

Spenti: LJ 47 + Cube + Lamp (pic flowerpot = ON), stages


SUMMARY W#2------------------------------------

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.28 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 5.88 (estimated)
Bin: 2 - Framerate 625 Hz ( 800 kpix/sec) - RON (e-): 5.89 (estimated)
Bin: 2 - Framerate 400 Hz ( 800 kpix/sec) - RON (e-): 5.94 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.23 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.26 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.36 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.39 (estimated)

Temp LJ39
case = 38C
Head < -40C
Water = -4.6C

Per confronto riportiamo la migliore misura ottenuta ieri con il W1 nelle stesse condizioni
Rirot, ADC e testa ccd39 su stella di massa sul ccd39 Peltier (W#1 in torre)


SUMMARY W#1------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.63 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.19 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.27 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.35 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 4.49 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.75 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.91 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.87 (estimated)

Qui pero' la temp dell'acuqa era 15.0C, quindi la testa era a 20C in piu' rispetto al telescopio

Test W#1 in torre

- rimesso PW 24V interno
- Cambiata scheda input n. 1 ccd39

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.53 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 9.47 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 4.99 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 4.96 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 5.24 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 5.44 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 5.65 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 5.78 (estimated)

Il bin2 1Khz va malissimo

-- EnricoPinna - 24 Feb 2011
Topic revision: r2 - 25 Feb 2011, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.