Extended object test

set_up OK con oggetto esteso 10:00

dimensioni oggetto con fibra da 300 microni 60 pixiel di IRTC ovvero 600 mas

Partiamo con oggetto esteso, no focus

=======================================

DISTURBOIM 20110221_100457
Modal basis = KL20101207
type = Push-pull
Frames per mov = 3
Cycle = 1
#Modes = 400
Amp = 600opt_6l
Note = oggetto esteso fibra da 300 micron o oggetto da 0.60 arcsec (60 pixel IRTC)

=======================================

INTMATACQ 20110221_102036
Pup = 20110127_123653
Dist = 20110221_100457
Iteration = 4
Loop speed = 600
bin = 1
Modul = 10 lambda/D diam 400 mas
Rec.Close loop = 20110128_111658
Gain vector close loop = 0.05

ampiezza dei seganli misurati piuttosto bassa. Si aumen ta la ampioezza del disturbo.

=======================================

DISTURBOIM 20110221_103214
Modal basis = KL20101207
type = Push-pull
Frames per mov = 3
Cycle = 1
#Modes = 400
Amp = 672_08.fits
Note = oggetto esteso fibra da 300 micron o oggetto da 0.60 arcsec (60 pixel IRTC)

=======================================

INTMATACQ 20110221_103700
Pup = 20110127_123653
Dist = 20110221_103214
Iteration = 4
Loop speed = 600
bin = 1
Modul = 10 lambda/D diam 400 mas
Rec.Close loop = 20110128_143833
Gain vector close loop = 0.05
Note: ora i modi bassi sono troppo potenti

=======================================

Ampiezze per disturbo -> 400modes_10l_extendedobj.fits
Partiamo da questa intmat per generarsi un nuovo file di ampiezze per avere segnale 0.25 rms nei modi troppo ampi.
I modi alti sono limitati dalla possibilita' dello specchio

=======================================

DISTURBOIM 20110221_104518
Modal basis = KL20101207
type = Push-pull
Frames per mov = 3
Cycle = 1
#Modes =400
Amp = 400modes_10l_extendedobj.fits
Note =

=======================================

INTMATACQ 20110221_105220
Pup = 20110127_123653
Dist = 20110221_104518
Iteration = 4
Loop speed = 600
bin = 1
Modul = 10 lambda/D diam 400 mas
Rec.Close loop = 20110128_143833
Gain vector close loop = 0.05
Note:

=======================================

REC 20110221_105406
Modal Basis: KL20101207
intmatacq = 20110221_105220
n modes = 400
TT removal = OFF
cutted = 0
Note: Oscilla con gain 0.15 vicino all'ostruzione centrale

=======================================

Riaggiustiamo le ampiezze, infatti alcuni modi erano ancora un po' saturati

Quindi iteriamo la normalizzazione delle ampiezze per ottenre il sengale di 0.25

Ampiezze per disturbo -> 400modes_10l_extendedobj2.fits

=======================================

DISTURBOIM 20110221_110448
Modal basis = KL20101207
type = Push-pull
Frames per mov = 3
Cycle = 1
#Modes =400
Amp = 400modes_10l_extendedobj2.fits
Note =

=======================================

INTMATACQ 20110221_105220
Pup = 20110127_123653
Dist = 20110221_110448
Iteration = 4
Loop speed = 600
bin = 1
Modul = 10 lambda/D diam 400 mas
Rec.Close loop = 20110128_143833
Gain vector close loop = 0.05
Note:

=======================================

REC 20110221_110630
Modal Basis: KL20101207
intmatacq = 20110221_105220
n modes = 400
TT removal = OFF
cutted = 0
Note: Ora gli rms dei segnali sono veramente attorno a 0.2, stesso problema al centro , ora regge 0.20 fi gsin

=======================================

REC 20110221_110759
Modal Basis: KL20101207
intmatacq = 20110221_105220
n modes = 400
TT removal = ON
cutted = 0
Note: Ora gli rms dei segnali sono veramente attorno a 0.25

=======================================

Proviamo col disturbo atmosferico (NO LGS simulation con fuoco)
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_111549 ??? 6.6 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 600 0.3 0.7 10 2314 ON -1.0 ? -1 -1 ON  
Ci sono problemi di oscillazioni sui modi alti anche con il disturbo atmosferico, si vede anche dal modal plot.

andiamo a 800 e riottimizzato il gain. Ora meglio...

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_112313 ??? 6.6 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.6 10 1750 ON -1.0 ? -1 -1 ON  
20110221_112816 ??? 6.6 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.3 0.6 10 1750 ON -1.0 ? -1 -1 ON  
20110221_112950 ??? 6.6 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.3 0.6 10 1750 ON -1.0 ? -1 -1 ON  
Ora abbassiamo la lampada

Andiamo a 200phot/sub/frame

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_113456 ??? 9.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.4 10 177 ON -1.0 ? -1 -1 ON  
20110221_113551 ??? 9.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.4 10 175 ON -1.0 ? -1 -1 ON  
20110221_113623 ??? 9.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.4 10 174 ON -1.0 ? -1 -1 ON  

Andiamo a 70phot/sub/frame

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_114301 ??? 10.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.4 10 69 ON -1.0 ? -1 -1 ON  
20110221_114429 ??? 10.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.4 10 70 ON -1.0 ? -1 -1 ON  
20110221_114458 ??? 10.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 800 0.2 0.4 10 68 ON -1.0 ? -1 -1 ON  

  • extd_obj_vs_NGSdl.png:
    extd_obj_vs_NGSdl.png

Extedned obj alto flusso a 1kHz

Facciamo alto flusso a 1 kHz per vedere se ad alto flusso raggiungiamo le performances del DL
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_120844 ??? 7.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 955 ON -1.0 ? -1 -1 ON  
20110221_120940 ??? 7.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 954 ON -1.0 ? -1 -1 ON  
20110221_121013 ??? 7.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 957 ON -1.0 ? -1 -1 ON  

Ora prendiamo un po' di PSF dell'ext obj a loop chiuso bright end

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_121305 ??? 7.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 960 ON 0.7 H 200 100 ON SR may be in error
20110221_121344 ??? 7.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 958 ON 0.7 H 200 100 ON SR may be in error
20110221_121422 ??? 7.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 958 ON 0.7 H 200 100 ON SR may be in error

Lasciamo lo stesso flusso e facciamo OL

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_121623 ??? 7.2 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.0 0.0 10 832 ON 0.3 H 200 100 ON SR may be in error
20110221_121701 ??? 7.2 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.0 0.0 10 876 ON 0.3 H 200 100 ON SR may be in error
20110221_121739 ??? 7.2 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.0 0.0 10 897 ON 0.3 H 200 100 ON SR may be in error
oggi=obj_new('aodataset',from='20110221_121305',to='20110221_121900')

Facciamo una PSF CL a regime astronomico

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_122259 ??? 9.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 145 ON 0.7 H 200 100 ON SR may be in error
20110221_122338 ??? 9.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 149 ON 0.7 H 200 100 ON SR may be in error
20110221_122416 ??? 9.1 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 950 0.2 0.3 10 145 ON 0.7 H 200 100 ON SR may be in error

  • ol_clf_clb.png:
    ol_clf_clb.png

Extended Obj + Focus

ATTENZIONE QUI LE MAG VANNO RISCALATE PERCHE' SI VA A 100Hz (con oversampling x focus) ma con atmo come da 1kHz

Qui usiamo il REC dell'ext obj Bright 1500 phot
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_123851 ??? 9.0 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.5 1.2 10 1544 ON -1.0 ? -1 -1 ON  
20110221_123957 ??? 9.0 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.5 1.2 10 1544 ON -1.0 ? -1 -1 ON  
20110221_124438 ??? 9.0 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.2 0.3 10 1544 ON -1.0 ? -1 -1 ON  

Faint
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_124942 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 150 ON -1.0 ? -1 -1 ON  
20110221_125043 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 153 ON -1.0 ? -1 -1 ON  
20110221_125145 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 153 ON -1.0 ? -1 -1 ON  
20110221_125533 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 156 ON -1.0 ? -1 -1 ON  

Abbiamo una correzione sui modi bassi un po' deboli Proviamo ad usare il REC del focus con 20 l/D
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_130533 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 131034 1 350 100 0.4 0.7 20 155 ON -1.0 ? -1 -1 ON  
20110221_130630 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 131034 1 350 100 0.4 0.7 20 155 ON -1.0 ? -1 -1 ON  
20110221_130735 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 131034 1 350 100 0.4 0.7 20 153 ON -1.0 ? -1 -1 ON  
Il modal plot sembra non accorgersi del diverso REC e della mod che e' raddoppiata

Prendiamo la intmat con Ext obj e focus

=======================================

INTMATACQ 20110221_143922
Pup = 20110127_123653
Dist = 20110217_114846 (con fuoco)
Iteration = 4
Loop speed = 100
bin = 1
Modul = 20 l/D
Rec.Close loop = 20110217_131034
Gain vector close loop = 0.05

=======================================

REC 20110221_145039
Modal Basis:
intmatacq = 20110221_143922
n modes = 349
TT removal = OFF
cutted = #
Note: FOCUS REMOVAL ON

=======================================

AO> dl_focus=getaoelab('20110218_112538')
AO> sl_focus=getaoelab('20110221_145904')
AO> sl_nofocus=getaoelab('20110221_112313')
AO> dl_nofocus_80=getaoelab('20110209_121110')
AO> dl_nofocus_66=getaoelab('20110216_175218')

AO> sl_nofocus_800Hz=getaoelab('20110221_112313')

AO> sl_nofocus_100Hz=getaoelab('20110221_155034')

  • LGSextended_vs_NGS.png: dl_nofocus_80 (bianco); dl_nofocus_66 (bianco); sl_focus (giallo)
    LGSextended_vs_NGS.png
Il giallo (LGS estended) mostra un'ottima correzione ai modi alti (come NGS bright), questo puo' essere un vantaggio legato al delay. Qui infatti stiamo andando a 100Hz con una atm piu' veloce. Quindi stiamo regalalndo empo allo specchio,

Questa depressione infatti non si vede con l'extended object senz afuoco che andava a 800.

Ora facciamo il crosscheck: andiamo lenti (100Hz) con il solo extended per vedere se anche lui viene avvantaggiato

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_155034 ??? 9.0 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 1566 ON -1.0 ? -1 -1 ON  
20110221_155243 ??? 9.0 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 1568 ON -1.0 ? -1 -1 ON  
20110221_155351 ??? 9.0 -2147483648 -2147483648 -1.00 -1.00 110759 1 400 100 0.4 0.6 10 1568 ON -1.0 ? -1 -1 ON  
effettivamente avviene cio' che ci aspettiamo: i modi alti a 100Hz vengono avvantaggiati e scendono piu' bassi dell0 LGS bright (dl_nofocus_80)

La cosa che rimane non spiegata e' perche' il dl_focus faccia peggio del sl_focus.

FOCUS + EXTENDED OBJ + mag 9

[ricordiamo che andiamo a 100Hz con atm a 800Hz e la mag va ricalcolata, quella del logtwiki non funziona]
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_160330 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 145039 1 350 100 0.5 0.7 20 159 ON -1.0 ? -1 -1 ON  
20110221_160842 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 145039 1 350 100 0.5 0.7 20 158 ON -1.0 ? -1 -1 ON  
20110221_161011 ??? 11.5 -2147483648 -2147483648 -1.00 -1.00 145039 1 350 100 0.5 0.7 20 158 ON -1.0 ? -1 -1 ON  

Andiamo ancora piu' faint (70phot per subap)
TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR band exp #frames disturb notes
20110221_161616 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 145039 1 350 100 0.3 0.5 20 61 ON -1.0 ? -1 -1 ON  
20110221_161719 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 145039 1 350 100 0.3 0.5 20 63 ON -1.0 ? -1 -1 ON  
20110221_161858 ??? 12.5 -2147483648 -2147483648 -1.00 -1.00 145039 1 350 100 0.3 0.5 20 61 ON -1.0 ? -1 -1 ON  

-- EnricoPinna - 21 Feb 2011

AO> dl_focus=getaoelab('20110218_112538')
AO> sl_focus=getaoelab('20110221_145904')
AO> sl_nofocus=getaoelab('20110221_112313')
AO> dl_nofocus_80=getaoelab('20110209_121110')
AO> dl_nofocus_66=getaoelab('20110216_175218')

AO> sl_nofocus_800Hz=getaoelab('20110221_112313')

AO> sl_nofocus_100Hz=getaoelab('20110221_155034')

AO>