You are here: Foswiki>FLAO Web>TowerLab>LabTest20110203 (04 Feb 2011, EnricoPinna)Edit Attach

Bin4

Pup = 20110203-104227
Centri =
Diam =
Dist =
TTMod = 6V / 200 Hz
Thr1 = 0.15
Thr2 = 0.30
Note = non ottimizzate. La forma sembra buona

Pup = 20110203-104743
Centri =
Diam =
Dist =
TTMod = 6V / 200 Hz
Nsubaps = 36
Thr1 = 0.15
Thr2 = 0.30
Note = Hottizzate. La forma fa schifo.

Pup = 20110203-105150
Centri =
Diam =
Dist =
TTMod = 6V / 200 Hz
Nsubaps = 38
Thr1 = 0.15
Thr2 = 0.30
Note = Hottizzate. Appena un pochino meglio, ma variando la soglia non cambia molto. Teniamo queste.

REC a 10 modi bin 4

INTMATACQ 20110203_105902
Pup = 20110203-105150
Dist = 20110131_174710
Iteration = 4
Loop speed = 600
bin = 4
Modul = 6 l/D
Rec.Close loop = NO

REC 20110203_110218
Modal Basis: KL_20101207
Modes = 10
intmatacq = 20110203_105902
n modes = 10
TT removal = ON
cutted = 0
Note:

REC 36 modi bin 4

DISTURBOIM 20110203_111810
Modal basis = KL_20101207
type = Push-pull
Frames per mov = 3
Cycle = 18
#Modes = 36
Amp = 600_optimized_6l
Note =

INTMATACQ 20110203_112010
Pup = 20110203-105150
Dist = 20110203_111810
Iteration = 4
Loop speed = 600
bin = 4
Modul = 6 l/D
Rec.Close loop = 20110203_110218
Gain = 0.05

REC 20110203_112130
Modal Basis: KL_20101207
Modes = 36
intmatacq = 20110203_112010
n modes = 36
TT removal = ON
cutted = 0
Note: chiude correttamente fino a TT = 0.5 HO = 0.5

ND installate nella AUX FW motorizzata thorlabs (quella nuova con mini USB)

Posizione Filtro/i T
1 NE05B + NE05B 32%*32%=10%
2 NE10B + NE10B 10%*10%=1%
3 NE30B 0.1%
4    
5    
6    

-- AlfioPuglisi - 03 Feb 2011
Topic revision: r2 - 04 Feb 2011, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.