Mattina: montato AGW#1 in torre

Pomeriggio: montato il barrone con lo specchio a 15 deg

Acceso W, functionality check OK
Ripristinato SW W sulla workstation della torre

Appaiono delle linee orizzonatli o verticali a seconda della velocita sul CCD39 nelquadrante in alto a DX.

Tutto acceso tranne CCD47
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 29.07 - RON (e-): 14.54 (estimated)
Speed: 2500 kpix/sec - RON (ADU): 29.07 - RON (e-): 14.54 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.25 - RON (e-): 8.63 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.25 - RON (e-): 8.63 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.67 - RON (e-): 5.33 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.67 - RON (e-): 5.33 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.99 - RON (e-): 6.00 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.99 - RON (e-): 6.00 (estimated)

Hexapod position:

hexmoveto([114.1, 1984.8, 1790.0, 41.0, 86.6, 0])

con questa posizione siha lospot sul 47 e luce su una pup

Le pupille sono vignettate nella direzione ortogonale al rotatore del cubo, quindi dovremo tiltare la pennina o lo specchio del barrone

BUON NATALE!!!