You are here: Foswiki>FLAO Web>TowerLab>LabTest20100810 (10 Aug 2010, AlessandroGherardi)Edit Attach
-- AlessandroGherardi - 10 Aug 2010

CCD47 alimentato con alimetatori da banco
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.98 - RON (e-): 9.99 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.52 - RON (e-): 6.26 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.83 - RON (e-): 3.91 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.95 - RON (e-): 4.47 (estimated)
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.65 - RON (e-): 10.32 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.78 - RON (e-): 6.39 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.71 - RON (e-): 3.85 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.46 - RON (e-): 4.23 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.82 - RON (e-): 9.91 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.81 - RON (e-): 6.40 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.70 - RON (e-): 3.85 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.50 - RON (e-): 4.25 (estimated)

Restart
CCD47OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.28 - RON (e-): 10.14 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.66 - RON (e-): 6.33 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.73 - RON (e-): 3.86 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.68 - RON (e-): 4.34 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.80 - RON (e-): 9.90 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.71 - RON (e-): 6.36 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.70 - RON (e-): 3.85 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.48 - RON (e-): 4.24 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.05 - RON (e-): 10.02 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.72 - RON (e-): 6.36 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.69 - RON (e-): 3.85 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.45 - RON (e-): 4.23 (estimated)

CCD47ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.88 - RON (e-): 9.94 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.72 - RON (e-): 6.36 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.69 - RON (e-): 3.84 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.22 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.98 - RON (e-): 9.99 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.72 - RON (e-): 6.36 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.68 - RON (e-): 3.84 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.22 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.82 - RON (e-): 9.91 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.70 - RON (e-): 6.35 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.67 - RON (e-): 3.83 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.49 - RON (e-): 4.25 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.08 - RON (e-): 10.04 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.71 - RON (e-): 6.35 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.67 - RON (e-): 3.83 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.44 - RON (e-): 4.22 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.91 - RON (e-): 9.95 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.72 - RON (e-): 6.36 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.67 - RON (e-): 3.83 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.41 - RON (e-): 4.21 (estimated)
Accesi i motori
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.54 - RON (e-): 9.77 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.31 - RON (e-): 6.16 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.65 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.44 - RON (e-): 4.22 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.71 - RON (e-): 9.85 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.34 - RON (e-): 6.17 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.64 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.40 - RON (e-): 4.20 (estimated)
Accese filter wheels e adc
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.95 - RON (e-): 9.98 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.38 - RON (e-): 6.19 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.83 - RON (e-): 3.92 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.15 - RON (e-): 4.58 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.87 - RON (e-): 9.93 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.42 - RON (e-): 6.21 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.83 - RON (e-): 3.92 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.95 - RON (e-): 4.48 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.04 - RON (e-): 10.02 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.42 - RON (e-): 6.21 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.79 - RON (e-): 3.90 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.91 - RON (e-): 4.45 (estimated)
Topic revision: r1 - 10 Aug 2010, AlessandroGherardi
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback