CCCD47 OFF con cavi 47 scollegati
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 30.71 - RON (e-): 15.36 (estimated)
Speed: 890 kpix/sec - RON (ADU): 13.34 - RON (e-): 6.67 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.88 - RON (e-): 3.94 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.53 - RON (e-): 4.26 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 27.12 - RON (e-): 13.56 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.10 - RON (e-): 7.05 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.01 - RON (e-): 4.00 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.66 - RON (e-): 4.33 (estimated)
toccando i connettori clck e bias LJ39
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 20.08 - RON (e-): 10.04 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.50 - RON (e-): 6.25 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.68 - RON (e-): 3.84 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.40 - RON (e-): 4.20 (estimated)
CCD47 ON cavi 47 scollegati
toccando i connettori clck e bias LJ39
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.07 - RON (e-): 12.04 (estimated)
Speed: 890 kpix/sec - RON (ADU): 15.69 - RON (e-): 7.85 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.91 - RON (e-): 3.96 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.86 - RON (e-): 4.93 (estimated)
senza toccare i connettori
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.22 - RON (e-): 12.11 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.71 - RON (e-): 6.36 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.98 - RON (e-): 3.99 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.87 - RON (e-): 5.43 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.78 - RON (e-): 12.39 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.44 - RON (e-): 6.22 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.84 - RON (e-): 3.92 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.94 - RON (e-): 4.47 (estimated)
---------------------------------------------------------------------------
POMERIGGIO
---------------------------------------------------------------------------
CCD47 OFF cavi 47 scollegati
Shield dei connettori clocks e bias cortocircuitato con cavo
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 22.58 - RON (e-): 11.29 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.37 - RON (e-): 6.19 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.72 - RON (e-): 3.86 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.59 - RON (e-): 4.30 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.93 - RON (e-): 9.96 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.43 - RON (e-): 6.22 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.64 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.41 - RON (e-): 4.21 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.27 - RON (e-): 9.64 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.43 - RON (e-): 6.21 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.63 - RON (e-): 3.81 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.42 - RON (e-): 4.21 (estimated)
CCD47 ON cavi 47 scollegati
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.17 - RON (e-): 11.59 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.28 - RON (e-): 6.14 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.80 - RON (e-): 3.90 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.73 - RON (e-): 4.37 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.39 - RON (e-): 11.70 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.27 - RON (e-): 6.13 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.14 - RON (e-): 4.07 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.74 - RON (e-): 4.37 (estimated)
Shield dei connettori connessi a GND
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.18 - RON (e-): 11.59 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.21 - RON (e-): 6.10 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.87 - RON (e-): 3.93 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.90 - RON (e-): 4.45 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.07 - RON (e-): 11.53 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.24 - RON (e-): 6.12 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.86 - RON (e-): 3.93 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.65 - RON (e-): 4.32 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.61 - RON (e-): 11.80 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.36 - RON (e-): 6.18 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.84 - RON (e-): 3.92 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.12 - RON (e-): 4.56 (estimated)
toccando i connettori
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.69 - RON (e-): 11.85 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.25 - RON (e-): 6.13 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.95 - RON (e-): 3.97 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.11 - RON (e-): 4.56 (estimated)
CCD47OFF cavi 47 scollegati
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.12 - RON (e-): 9.56 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.39 - RON (e-): 6.19 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.64 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.22 (estimated)
toccando cavi e connettori
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.36 - RON (e-): 9.68 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.40 - RON (e-): 6.20 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.65 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.42 - RON (e-): 4.21 (estimated)
sconnessi da GND
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.19 - RON (e-): 9.60 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.38 - RON (e-): 6.19 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.63 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.22 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.20 - RON (e-): 9.60 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.38 - RON (e-): 6.19 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.62 - RON (e-): 3.81 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.21 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.21 - RON (e-): 9.60 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.32 - RON (e-): 6.16 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.62 - RON (e-): 3.81 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.44 - RON (e-): 4.22 (estimated)
sconnessi tra di loro
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.34 - RON (e-): 9.67 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.30 - RON (e-): 6.15 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.63 - RON (e-): 3.82 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.41 - RON (e-): 4.20 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.36 - RON (e-): 9.68 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.31 - RON (e-): 6.15 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.60 - RON (e-): 3.80 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.38 - RON (e-): 4.19 (estimated)
CCD47 ON cavi 47 scollegati
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.57 - RON (e-): 12.28 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.42 - RON (e-): 6.21 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.85 - RON (e-): 3.93 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.23 - RON (e-): 4.61 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.47 - RON (e-): 12.23 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.34 - RON (e-): 6.17 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.22 - RON (e-): 4.11 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.45 - RON (e-): 4.72 (estimated)
shield dei connettori connessi tra loro
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.52 - RON (e-): 12.26 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.36 - RON (e-): 6.18 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.13 - RON (e-): 4.06 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.16 - RON (e-): 5.58 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.31 - RON (e-): 12.16 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.29 - RON (e-): 6.14 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.80 - RON (e-): 3.90 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.12 - RON (e-): 4.56 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 24.44 - RON (e-): 12.22 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.27 - RON (e-): 6.13 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.80 - RON (e-): 3.90 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.98 - RON (e-): 5.49 (estimated)
shield connessi tra loro e a GND
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.19 - RON (e-): 11.60 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.21 - RON (e-): 6.10 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.80 - RON (e-): 3.90 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.15 - RON (e-): 5.07 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.45 - RON (e-): 11.72 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.22 - RON (e-): 6.11 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.83 - RON (e-): 3.91 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.41 - RON (e-): 4.70 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.50 - RON (e-): 11.75 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.21 - RON (e-): 6.10 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.84 - RON (e-): 3.92 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.50 - RON (e-): 5.25 (estimated)
sconnessi da GND
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.27 - RON (e-): 11.64 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.23 - RON (e-): 6.12 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.89 - RON (e-): 3.94 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.94 - RON (e-): 4.47 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.01 - RON (e-): 11.51 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.18 - RON (e-): 6.09 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.89 - RON (e-): 3.95 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.94 - RON (e-): 4.47 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 23.56 - RON (e-): 11.78 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.20 - RON (e-): 6.10 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.87 - RON (e-): 3.93 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.66 - RON (e-): 4.83 (estimated)
>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>
CCD47OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 19.04 - RON (e-): 9.52 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.29 - RON (e-): 6.15 (estimated)
Speed: 400 kpix/sec - RON (ADU): 7.79 - RON (e-): 3.90 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.90 - RON (e-): 4.45 (estimated)
CCD47On con cavi 47 collegati
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 35.92 - RON (e-): 17.96 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.81 - RON (e-): 6.41 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.23 - RON (e-): 5.12 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.29 - RON (e-): 5.65 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 34.76 - RON (e-): 17.38 (estimated)
Speed: 890 kpix/sec - RON (ADU): 12.57 - RON (e-): 6.28 (estimated)
Speed: 400 kpix/sec - RON (ADU): 9.75 - RON (e-): 4.88 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.70 - RON (e-): 6.35 (estimated)