You are here: Foswiki>FLAO Web>TowerLab>LabTest20100804 (04 Aug 2010, AlessandroGherardi)Edit Attach
-- AlessandroGherardi - 04 Aug 2010

CCD47OFF, wall plug
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 25.37 - RON (e-): 12.69 (estimated)
Speed: 890 kpix/sec - RON (ADU): 15.13 - RON (e-): 7.56 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.18 - RON (e-): 4.09 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.90 - RON (e-): 4.45 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.48 - RON (e-): 13.24 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.98 - RON (e-): 7.49 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.16 - RON (e-): 4.08 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.55 - RON (e-): 4.27 (estimated)
CCD47ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 40.59 - RON (e-): 20.29 (estimated)
Speed: 890 kpix/sec - RON (ADU): 19.63 - RON (e-): 9.81 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.71 - RON (e-): 5.35 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.22 - RON (e-): 6.11 (estimated)
CCD47 OFF, head47 outside W1
NOTE: the CCD47 and CCD39 heads (chassy) are connected by means the clocks and video signal cables)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.70 - RON (e-): 13.35 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.28 - RON (e-): 7.14 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.61 - RON (e-): 4.30 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.17 - RON (e-): 5.08 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 27.37 - RON (e-): 13.69 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.13 - RON (e-): 7.07 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.33 - RON (e-): 4.17 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.67 - RON (e-): 4.33 (estimated)
CCD47 ON, head47 outside che W1
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 40.17 - RON (e-): 20.08 (estimated)
Speed: 890 kpix/sec - RON (ADU): 16.79 - RON (e-): 8.39 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.57 - RON (e-): 5.29 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.54 - RON (e-): 6.27 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 39.95 - RON (e-): 19.98 (estimated)
Speed: 890 kpix/sec - RON (ADU): 16.98 - RON (e-): 8.49 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.37 - RON (e-): 5.18 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.00 - RON (e-): 5.50 (estimated)
CCD47OFF, BOX3 closed
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.28 - RON (e-): 13.14 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.27 - RON (e-): 7.14 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.29 - RON (e-): 4.14 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.72 - RON (e-): 4.36 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.78 - RON (e-): 13.39 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.20 - RON (e-): 7.10 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.28 - RON (e-): 4.14 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.65 - RON (e-): 4.32 (estimated)
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 39.11 - RON (e-): 19.55 (estimated)
Speed: 890 kpix/sec - RON (ADU): 18.55 - RON (e-): 9.28 (estimated)
Speed: 400 kpix/sec - RON (ADU): 11.60 - RON (e-): 5.80 (estimated)
Speed: 150 kpix/sec - RON (ADU): 14.05 - RON (e-): 7.03 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 40.36 - RON (e-): 20.18 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.06 - RON (e-): 8.53 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.75 - RON (e-): 5.38 (estimated)
Speed: 150 kpix/sec - RON (ADU): 14.95 - RON (e-): 7.47 (estimated)
Stages on, CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 40.42 - RON (e-): 20.21 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.40 - RON (e-): 8.70 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.77 - RON (e-): 5.38 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.04 - RON (e-): 6.02 (estimated)
Stages ON, CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 27.65 - RON (e-): 13.82 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.11 - RON (e-): 7.06 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.33 - RON (e-): 4.17 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.65 - RON (e-): 4.33 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 27.49 - RON (e-): 13.75 (estimated)
Speed: 890 kpix/sec - RON (ADU): 14.13 - RON (e-): 7.07 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.32 - RON (e-): 4.16 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.68 - RON (e-): 4.34 (estimated)
CCD47 OFF, cavi 47 distesi
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 31.95 - RON (e-): 15.97 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.00 - RON (e-): 8.50 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.85 - RON (e-): 4.43 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.23 - RON (e-): 4.61 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 33.49 - RON (e-): 16.74 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.17 - RON (e-): 8.58 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.85 - RON (e-): 4.43 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.21 - RON (e-): 4.60 (estimated)
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 47.18 - RON (e-): 23.59 (estimated)
Speed: 890 kpix/sec - RON (ADU): 22.07 - RON (e-): 11.03 (estimated)
Speed: 400 kpix/sec - RON (ADU): 11.09 - RON (e-): 5.54 (estimated)
Speed: 150 kpix/sec - RON (ADU): 13.87 - RON (e-): 6.93 (estimated)
Spostata massa Peltier dalla testa CCD39 alla testa CCD47.
CCD47 OFF, cavi 47 arrotolati come prima
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 28.19 - RON (e-): 14.09 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.73 - RON (e-): 8.86 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.72 - RON (e-): 4.36 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.64 - RON (e-): 4.82 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 28.70 - RON (e-): 14.35 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.49 - RON (e-): 8.75 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.65 - RON (e-): 4.33 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.24 - RON (e-): 4.62 (estimated)
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 40.38 - RON (e-): 20.19 (estimated)
Speed: 890 kpix/sec - RON (ADU): 21.50 - RON (e-): 10.75 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.95 - RON (e-): 5.48 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.46 - RON (e-): 6.23 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 39.94 - RON (e-): 19.97 (estimated)
Speed: 890 kpix/sec - RON (ADU): 21.86 - RON (e-): 10.93 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.75 - RON (e-): 5.38 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.66 - RON (e-): 6.33 (estimated)
Posta una lamina a GND tra testa39 e testa47
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 39.48 - RON (e-): 19.74 (estimated)
Speed: 890 kpix/sec - RON (ADU): 22.35 - RON (e-): 11.18 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.80 - RON (e-): 5.40 (estimated)
Speed: 150 kpix/sec - RON (ADU): 12.16 - RON (e-): 6.08 (estimated)
Posta una lamina a GND tra testa39 box39
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 39.05 - RON (e-): 19.53 (estimated)
Speed: 890 kpix/sec - RON (ADU): 21.51 - RON (e-): 10.75 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.58 - RON (e-): 5.29 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.66 - RON (e-): 5.83 (estimated)
Connesse le teste dei CCD con un cavo
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 47.13 - RON (e-): 23.56 (estimated)
Speed: 890 kpix/sec - RON (ADU): 22.41 - RON (e-): 11.21 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.78 - RON (e-): 5.39 (estimated)
Speed: 150 kpix/sec - RON (ADU): 16.56 - RON (e-): 8.28 (estimated)
Connesse le teste dei CCD con un cavo
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 36.83 - RON (e-): 18.42 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.35 - RON (e-): 8.68 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.78 - RON (e-): 4.39 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.61 - RON (e-): 4.81 (estimated)
Disconnesse le teste dei CCD
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 28.15 - RON (e-): 14.07 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.82 - RON (e-): 8.91 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.73 - RON (e-): 4.37 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.15 - RON (e-): 4.58 (estimated)
LJ fans OFF
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 28.35 - RON (e-): 14.17 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.89 - RON (e-): 8.94 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.74 - RON (e-): 4.37 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.19 - RON (e-): 4.60 (estimated)
LJ fans OFF, BOX fan OFF
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.45 - RON (e-): 13.22 (estimated)
Speed: 890 kpix/sec - RON (ADU): 16.77 - RON (e-): 8.39 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.43 - RON (e-): 4.22 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.89 - RON (e-): 4.44 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.86 - RON (e-): 13.43 (estimated)
Speed: 890 kpix/sec - RON (ADU): 17.01 - RON (e-): 8.51 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.44 - RON (e-): 4.22 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.86 - RON (e-): 4.43 (estimated)
LJ fans ON, BOX fan OFF
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 27.43 - RON (e-): 13.72 (estimated)
Speed: 890 kpix/sec - RON (ADU): 16.89 - RON (e-): 8.44 (estimated)
Speed: 400 kpix/sec - RON (ADU): 8.45 - RON (e-): 4.23 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.90 - RON (e-): 4.45 (estimated)
LJ fans OFF, BOX fan OFF
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 44.82 - RON (e-): 22.41 (estimated)
Speed: 890 kpix/sec - RON (ADU): 19.34 - RON (e-): 9.67 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.22 - RON (e-): 5.11 (estimated)
Speed: 150 kpix/sec - RON (ADU): 10.97 - RON (e-): 5.49 (estimated)
Spostati cavi video e clocks (in mano)
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 41.04 - RON (e-): 20.52 (estimated)
Speed: 890 kpix/sec - RON (ADU): 26.21 - RON (e-): 13.10 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.18 - RON (e-): 5.09 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.43 - RON (e-): 5.71 (estimated)
Spostati cavi video e clocks
CCD47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 41.95 - RON (e-): 20.97 (estimated)
Speed: 890 kpix/sec - RON (ADU): 25.80 - RON (e-): 12.90 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.52 - RON (e-): 5.26 (estimated)
Speed: 150 kpix/sec - RON (ADU): 11.61 - RON (e-): 5.81 (estimated)
Spostati cavi video e clocks
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 26.18 - RON (e-): 13.09 (estimated)
Speed: 890 kpix/sec - RON (ADU): 21.03 - RON (e-): 10.51 (estimated)
Speed: 400 kpix/sec - RON (ADU): 9.00 - RON (e-): 4.50 (estimated)
Speed: 150 kpix/sec - RON (ADU): 8.89 - RON (e-): 4.44 (estimated)
Fascettati cavi video e clocks (come prima)
CCD47 OFF
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 32.09 - RON (e-): 16.04 (estimated)
Speed: 890 kpix/sec - RON (ADU): 24.52 - RON (e-): 12.26 (estimated)
Speed: 400 kpix/sec - RON (ADU): 9.73 - RON (e-): 4.87 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.31 - RON (e-): 4.66 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 29.44 - RON (e-): 14.72 (estimated)
Speed: 890 kpix/sec - RON (ADU): 21.80 - RON (e-): 10.90 (estimated)
Speed: 400 kpix/sec - RON (ADU): 9.27 - RON (e-): 4.64 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.17 - RON (e-): 4.59 (estimated)
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 29.66 - RON (e-): 14.83 (estimated)
Speed: 890 kpix/sec - RON (ADU): 21.86 - RON (e-): 10.93 (estimated)
Speed: 400 kpix/sec - RON (ADU): 9.27 - RON (e-): 4.63 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.18 - RON (e-): 4.59 (estimated)
CCD 47 ON
------------------------------ SUMMARY ------------------------------------
ccd39 - binning 1
Speed: 2500 kpix/sec - RON (ADU): 45.37 - RON (e-): 22.68 (estimated)
Speed: 890 kpix/sec - RON (ADU): 26.95 - RON (e-): 13.47 (estimated)
Speed: 400 kpix/sec - RON (ADU): 10.08 - RON (e-): 5.04 (estimated)
Speed: 150 kpix/sec - RON (ADU): 9.83 - RON (e-): 4.91 (estimated)

CCD47OFF


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 29.30 - RON (e-): 14.65 (estimated)

Speed: 890 kpix/sec - RON (ADU): 17.77 - RON (e-): 8.88 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.88 - RON (e-): 4.44 (estimated)

Speed: 150 kpix/sec - RON (ADU): 9.61 - RON (e-): 4.81 (estimated)

BOX39 shielded, CCD47off


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 34.43 - RON (e-): 17.22 (estimated)

Speed: 890 kpix/sec - RON (ADU): 18.44 - RON (e-): 9.22 (estimated)

Speed: 400 kpix/sec - RON (ADU): 9.01 - RON (e-): 4.50 (estimated)

Speed: 150 kpix/sec - RON (ADU): 9.21 - RON (e-): 4.61 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 21.26 - RON (e-): 10.63 (estimated)

Speed: 890 kpix/sec - RON (ADU): 17.63 - RON (e-): 8.82 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.44 - RON (e-): 4.22 (estimated)

Speed: 150 kpix/sec - RON (ADU): 9.17 - RON (e-): 4.58 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 25.78 - RON (e-): 12.89 (estimated)

Speed: 890 kpix/sec - RON (ADU): 15.77 - RON (e-): 7.89 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.42 - RON (e-): 4.21 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.92 - RON (e-): 4.46 (estimated)

No shield, CCD47 OFF


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 24.65 - RON (e-): 12.32 (estimated)

Speed: 890 kpix/sec - RON (ADU): 15.57 - RON (e-): 7.78 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.31 - RON (e-): 4.15 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.85 - RON (e-): 4.43 (estimated)

BOX47 disconnected

nota: nonostante la box47 sia sconnessa, il SW non da "offline"


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 21.62 - RON (e-): 10.81 (estimated)

Speed: 890 kpix/sec - RON (ADU): 14.16 - RON (e-): 7.08 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.17 - RON (e-): 4.09 (estimated)

Speed: 150 kpix/sec - RON (ADU): 9.07 - RON (e-): 4.53 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 22.86 - RON (e-): 11.43 (estimated)

Speed: 890 kpix/sec - RON (ADU): 14.63 - RON (e-): 7.32 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.15 - RON (e-): 4.07 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.77 - RON (e-): 4.39 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 22.67 - RON (e-): 11.33 (estimated)

Speed: 890 kpix/sec - RON (ADU): 14.81 - RON (e-): 7.40 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.08 - RON (e-): 4.04 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.58 - RON (e-): 4.29 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 22.51 - RON (e-): 11.26 (estimated)

Speed: 890 kpix/sec - RON (ADU): 14.76 - RON (e-): 7.38 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.07 - RON (e-): 4.03 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.63 - RON (e-): 4.32 (estimated)

ATTENZIONE:

TOCCANDO I CONNETTORI DI CLOCK E BIASES SUL LJ39 APPARE O

SCOMPARE IL PATTERN RUMOROSO

anello connesso a terra, box 47 disconnected


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 19.10 - RON (e-): 9.55 (estimated)

Speed: 890 kpix/sec - RON (ADU): 12.47 - RON (e-): 6.23 (estimated)

Speed: 400 kpix/sec - RON (ADU): 7.68 - RON (e-): 3.84 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.40 - RON (e-): 4.20 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 19.01 - RON (e-): 9.51 (estimated)

Speed: 890 kpix/sec - RON (ADU): 12.34 - RON (e-): 6.17 (estimated)

Speed: 400 kpix/sec - RON (ADU): 7.60 - RON (e-): 3.80 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.30 - RON (e-): 4.15 (estimated)


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 19.03 - RON (e-): 9.51 (estimated)

Speed: 890 kpix/sec - RON (ADU): 12.40 - RON (e-): 6.20 (estimated)

Speed: 400 kpix/sec - RON (ADU): 7.62 - RON (e-): 3.81 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.30 - RON (e-): 4.15 (estimated)

Anello sconnesso da terra.


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 19.17 - RON (e-): 9.58 (estimated)

Speed: 890 kpix/sec - RON (ADU): 12.42 - RON (e-): 6.21 (estimated)

Speed: 400 kpix/sec - RON (ADU): 7.62 - RON (e-): 3.81 (estimated)

Speed: 150 kpix/sec - RON (ADU): 8.37 - RON (e-): 4.18 (estimated)

Tocco i connettori di clock e bias e riappare il pattern


SUMMARY ------------------------------------

ccd39 - binning 1

Speed: 2500 kpix/sec - RON (ADU): 27.67 - RON (e-): 13.84 (estimated)

Speed: 890 kpix/sec - RON (ADU): 18.76 - RON (e-): 9.38 (estimated)

Speed: 400 kpix/sec - RON (ADU): 8.87 - RON (e-): 4.43 (estimated)

Speed: 150 kpix/sec - RON (ADU): 9.06 - RON (e-): 4.53 (estimated)