-- RunaBriguglio - 01 Mar 2010

power_on

fsm_startup

fsm_set_flat(/FF)

errore dsp#0-->167 dist_th
fsm_set_flat(/FF) again

temp_analisys su #220, to verifiy #220 behaviour. during FF acquisition it revealed large residuals
#220 mean=5.334e-5 stdev=2.4e-9
#220 stdev=2.4e-9
#220 stdev=2.3E-9
#220 stdev=2.2e-9
#220 mean=5.335e-5 stdev=2.4e-9

55 sec integr time
#220 mean=5.335e-5 stdev=2.8e-9 small drift noticeable

100 sec integr time
#220 mean=5.336e-5 stdev=4.2e-9 drift noticeable

data saved in
/home/aoacct/meas/adsec_calib/temp_analisys/20100301_gainFF.....

Miniskirt and shroud mounted

/home/aoacct/AO/current/conf/adsec/672a/TS3/gain_60um.txt modified

p_clist float 0.13

fsm_load_program ERROR

Fastlink test failed on DSP#: 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167

Fsm_reset

Fsm_load_program ok

2010-03-01 19:08:22.764388

Fsm_set_flat

Il commando non va Avanti, la AdSecMir gui segna zero

 =IDLCTRL00         |ALW|      684|2010-03-01 19:12:12.909250|            MAIN > EndCmd= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > % Compiled module: FSM_SET_FLAT.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Compiled module: UPDATE_STATUS.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Compiled module: READ_SEQ_CH.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Compiled module: CHECK_CHANNEL.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Compiled module: GET_AVE_PC.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Compiled module: SET_ACC.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Compiled module: START_ACC.= 
 =IDLCTRL00         |ALW|      685|2010-03-01 19:12:12.909299|            MAIN > . % Interrupt encountered.= 
<span>IDLCTRL00          |ALW|       685|2010-03-01 19:12:12.909299|             MAIN >  . % Interrupted at: GET_AVE_PC       </span> 

Terminal è fermo

Chiuso idlctrl e rilanciato

Continua a dare errore su get ave pc

Ao_stop e ao_start

Print, fsm_load_program

Set_flat Funziona

Maximum mode index used to apply flat -> # 378

passo 3

Param Selector: 000061C0

COMM MAX [m] 7.54430e-05

FORCE MAX [N] 0.192553

.r test_noise dati in /local/aolog/current/'_lin_13pgain_LBT_0

Temporal analisys saved in /home/aoacct/meas/adsec_calib/temp_analisys/20100301_13pgain.sav

the noise structure and value is close to that of EM AccTest (noise =4nm). the oscillation disappeared

Fsm_rip()

Fsm_set_flat

Maximum mode index used to apply flat -> # 380

passo 3

Param Selector: 000041C0

COMM MAX [m] 7.54415e-05

FORCE MAX [N] 0.218797

Salvataggio di ancuni status_save per comparazione dati arcetri e LBTO

Dati salvati in /home/aoacct/meas/adsec_calib /LBT-lab_status_# 10 salvataggi

Print, fsm_rip

Print, fsm_set_flat(/ff)

Maximum mode index used to apply flat -> # 395

passo 3

Param Selector: 000061C0

COMM MAX [m] 7.54164e-05

FORCE MAX [N] 0.213544

.r ff_acquire_v2, 5 cicli; .r ff_reduce_v2

Risultato: l'attuatore #244 mostra ancora residui importanti, occorre verificare con la step resp.

Fsm_rip; Fsm_set_flat (per togliere le impostazioni del FF)

Maximum mode index used to apply flat -> # 396

passo 3

Param Selector: 000061C0

COMM MAX [m] 7.54478e-05

FORCE MAX [N] 0.216200

Configurato il file step_default.txt

!!! per lanciare lo step_script occorre che XADSEC sia su con update panels

Rip della shell perché si è piantato lo step script (non trovava il file di configurazione perché ancava la sub directory::: sistemato)

FASTDGN00 |WAR| 222105|2010-03-01 21:52:26.552974| FUNCTWARNING > FunctWarning CHDISTAVERAGE-0000 0.00012965

FASTDGN00 |WAR| 222106|2010-03-01 21:52:26.553137| FUNCTEMERGENCYST > CHDISTAVERAGE-0112 = 0.000538805 [/home/aoacct/src/DiagnApp/Funct.h:258]

FASTDGN00 |INF| 222107|2010-03-01 21:52:26.553151| ADAM > Adam: disabling coils...

FASTDGN00 |DEB| 222108|2010-03-01 21:52:26.553176| ADAM > Command to Adam: #011300

FASTDGN00 |DEB| 222109|2010-03-01 21:52:26.553835| ADAM > Answer from Adam: >01^M#011300

FASTDGN00 |ERR| 222110|2010-03-01 21:52:26.553857| ADAM > Adam: coils disabling failed

FASTDGN00 |ERR| 222111|2010-03-01 21:52:26.553882| FUNCTEMERGENCYST > ADAM CAN'T DISABLE COILS [/home/aoacct/src/DiagnApp/Funct.h:263]

FASTDGN00 |WAR| 222112|2010-03-01 21:52:26.576457| FUNCTWARNING > FunctWarning CHINTCONTROLCURRENT-0133 -0.307014

FASTDGN00 |ERR| 222113|2010-03-01 21:52:26.577604| FUNCTALARM > FunctAlarm CHDISTRMS-0000 3.73585e-06

!!! vedere i log 22:00:52.587202 perché dà problemi il set flat

Allora fsm_load_program

Problema con #455:

FASTDGN00 |WAR| 422374|2010-03-01 22:03:26.883922| FUNCTEMERGENCYST > CHDISTAVERAGE-0455 = 1.74354 [/home/aoacct/src/DiagnApp/Funct.h:258]

FASTDGN00 |INF| 422375|2010-03-01 22:03:26.883949| ADAM > Adam: disabling coils...

FASTDGN00 |DEB| 422376|2010-03-01 22:03:26.883968| ADAM > Command to Adam: #011300

FASTDGN00 |DEB| 422377|2010-03-01 22:03:26.884623| ADAM > Answer from Adam: >01^M#011300

FASTDGN00 |ERR| 422378|2010-03-01 22:03:26.884636| ADAM > Adam: coils disabling failed

FASTDGN00 |ERR| 422379|2010-03-01 22:03:26.884660| FUNCTEMERGENCYST > ADAM CAN'T DISABLE COILS [/home/aoacct/src/DiagnApp/Funct.h:263]

Il 455 si trova qui ed è il 646 mirror

Power_off()

Check on #455 flat cable. Look at log book 2010_02_27 same error on #455

#455 flat cable re-inserted into its dist-board connector. now the act is working.

fsm_set_flat

xadsec per lo step; .r step_script

FASTDGN00 |WAR| 222106|2

error on #112: same error as before (accumulators??)

010-03-01 21:52:26.553137| FUNCTEMERGENCYST > CHDISTAVERAGE-0112 = 0.000538805 [/home/aoacct/src/DiagnApp/Funct.h:258]
FASTDGN00 |INF| 222107|2010-03-01 21:52:26.553151| ADAM > Adam: disabling coils...
FASTDGN00 |DEB| 222108|2010-03-01 21:52:26.553176| ADAM > Command to Adam: #011300
FASTDGN00 |DEB| 222109|2010-03-01 21:52:26.553835| ADAM > Answer from Adam: >01^M#011300
FASTDGN00 |ERR| 222110|2010-03-01 21:52:26.553857| ADAM > Adam: coils disabling failed
FASTDGN00 |ERR| 222111|2010-03-01 21:52:26.553882| FUNCTEMERGENCYST > ADAM CAN'T DISABLE

COILS [/home/aoacct/src

FSM_SET_FLAT