Mattina installazione AGW#2 sul RFBG con il barrone installato

h15:00 il W pinga

h20:15 Functional check @RFBG

Pupille vignettate dalla carta di protezione della telecentrica

Cube stage -> OK
Cube rot -> OK
LAMP-> OK

Stage XYZ -> OK
ADC-> OK
FW#1-> OK
FW#2 -> OK
CCD47-> OK
TT-> 3 assi provati indipendentemente OK
Rerot -> OK (pupille ruotano su CCD39)
Camera lens stage -> OK
CCD39 -> OK

RON TEST - Bayside stage frenate


SUMMARY ------------------------------------
Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 8.38 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 5.27 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 3.90 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 3.91 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 3.91 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 3.89 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 3.93 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.09 (estimated)

-- EnricoPinna - 31 Oct 2014