Final RON test

Everything ON except for the whole flowerpot:

Bin: 1 - Framerate 990 Hz (2500 kpix/sec) - RON (e-): 9.44 (estimated)
Bin: 2 - Framerate 990 Hz ( 800 kpix/sec) - RON (e-): 6.39 (estimated)
Bin: 2 - Framerate 625 Hz ( 380 kpix/sec) - RON (e-): 3.87 (estimated)
Bin: 2 - Framerate 400 Hz ( 380 kpix/sec) - RON (e-): 3.89 (estimated)
Bin: 3 - Framerate 200 Hz ( 380 kpix/sec) - RON (e-): 3.97 (estimated)
Bin: 4 - Framerate 300 Hz ( 335 kpix/sec) - RON (e-): 4.09 (estimated)
Bin: 4 - Framerate 200 Hz ( 335 kpix/sec) - RON (e-): 4.17 (estimated)
Bin: 4 - Framerate 105 Hz ( 335 kpix/sec) - RON (e-): 4.27 (estimated)

Installazione e messa a fuoco FINZ

LBT-VisAO_FINZ.zmx: LBT-VisAO_FINZ.zmx

-- AlfioPuglisi - 30 Jan 2012 *