TestLED


DEROT TRACKNUM (20110601_)

20 185114
40 185718
60 185757
80 185824
100 185859
120 185937
140 190005
160 190031
180 190057
200 190125
210 190152

Homing hexapod and move to position:

0, 1.73, 3.89, -324, -40, 0

20 193429
40 193502
60 193544
80 193734
100 193801
120 193824
140 193855
160 193922
180 193957
200 194025
210 194057

RECUPERO SORGENTE

Sorgente abilmente recuperata, siamo ora centrati a loop chiuso con 400 modi nel centro del campo

Hexapod:

X=0 Y=1.70, Z=3.79, RX=-308.2, RY=-39.4, RZ=0

After recovering the source we reduced the rotation data done at BIN1. The X axis of the camera lens get saturated so we decided to do a digital shift of the pupil

After the digital shift the camera lens is able to center the pupil back in the full range of the allowed PISCES rotation 20-210 deg of the instrument rotator.

Then we check BIN2 and BIN3 they are both OK in the range 20-210 of the instrument rotator.

During the test of BIN1,BIN2, BIN3 the loop is succcesfully closed with up to 400 modes

We still have to test BIN4

-- AlfioPuglisi - 01 Jun 2011