Clocking Rec KL_v7 bin2

Si fanno le 6 prove con il clocking ruotato a passi di 5 gradi

Dist Base Mod Bin Type Fr Cyc #mod AmpFile Note
20101024_002515 KL_v7 3 2 Push-pull 3 1 10 optimized_amano_3l_153.fits  
20101023_070036 KL_v7 3 2 Push-pull 3 1 153 optimized_amano_3l_153.fits  

Ricostruttori a 10 modi

Si fanno tutti e 6 i rec

IM Pup Dist It Freq Bin Mod CL Rec gain notes
20101024_003808 20100528_023248 20101024_002515 4 600 2 3 NO     clock 123.5
20101024_004113 20100528_023248 20101024_002515 4 600 2 3 NO     clock 126.0
20101024_004320 20100528_023248 20101024_002515 4 600 2 3 NO     clock 128.5
20101024_004511 20100528_023248 20101024_002515 4 600 2 3 NO     clock 131.0
20101024_004653 20100528_023248 20101024_002515 4 600 2 3 NO     clock 133.5
20101024_004838 20100528_023248 20101024_002515 4 600 2 3 NO     clock 136.0
20101024_005028 20100528_023248 20101024_002515 4 600 2 3 NO     clock 138.5

REC Base IM bin mod #modes TTrem cutted notes
20101024_004251 KL_v7 20101024_003808 2 3 9 ON 0 clock 123.5
20101024_004657 KL_v7 20101024_004113 2 3 9 ON 0 clock 126.0
20101024_004827 KL_v7 20101024_004320 2 3 9 ON 0 clock 128.5
20101024_004914 KL_v7 20101024_004511 2 3 9 ON 0 clock 131.0
20101024_005016 KL_v7 20101024_004653 2 3 9 ON 0 clock 133.5
20101024_005052 KL_v7 20101024_004838 2 3 9 ON 0 clock 136.0
20101024_005159 KL_v7 20101024_005028 2 3 9 ON 0 clock 138.5

Ricostruttori a 153 modi

IM Pup Dist It Freq Bin Mod CL Rec gain notes
20101024_005342 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_004251 gain_0.05.fits clock 123.5
20101024_010142 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_004657 gain_0.05.fits clock 126.0
20101024_010628 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_004827 gain_0.05.fits clock 128.5
20101024_011114 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_004914 gain_0.05.fits clock 131.0
20101024_011557 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_005016 gain_0.05.fits clock 133.5
20101024_012023 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_005052 gain_0.05.fits clock 136.0
20101024_012455 20100528_023248 20101024_002515 12 600 2 3 SI 20101024_005159 gain_0.05.fits clock 138.5

REC Base IM bin mod #modes TTrem cutted notes
20101024_005845 KL_v7 20101024_005342 2 3 153 ON 0 clock 123.5
20101024_010528 KL_v7 20101024_010142 2 3 153 ON 0 clock 126.0
20101024_011102 KL_v7 20101024_010628 2 3 153 ON 0 clock 128.5
20101024_011554 KL_v7 20101024_011114 2 3 153 ON 0 clock 131.0
20101024_011938 KL_v7 20101024_011557 2 3 153 ON 0 clock 133.5
20101024_012400 KL_v7 20101024_012023 2 3 153 ON 0 clock 136.0
20101024_013129 KL_v7 20101024_012455 2 3 153 ON 0 clock 138.5

Analisi best clocking del rec a bin 2

ampl = readfits('/local/towerdata/adsec_calib/M2C/KL_v7/modesAmp/optimized_amano_3lambda_153.fits', hampl)
ampenv = readfits('/local/towerdata/adsec_calib/M2C/KL_v7/amp_envelope.fits' , hampenv)
ampm = (ampl*ampenv)[0:152]
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_005845.fits_data.sav' & devs0=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_010528.fits_data.sav' & devs1=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_011102.fits_data.sav' & devs2=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_011554.fits_data.sav' & devs3=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_011938.fits_data.sav' & devs4=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_012400.fits_data.sav' & devs5=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101024_013129.fits_data.sav' & devs6=devs
restore, /v, '/local/towerdata/adsec_calib/M2C/KL_v7/RECs/Intmat_20101023_020127.fits_data.sav' & devs7=devs
sumq0 = total((devs0/ampm)^2)
sumq1 = total((devs1/ampm)^2)
sumq2 = total((devs2/ampm)^2)
sumq3 = total((devs3/ampm)^2)
sumq4 = total((devs4/ampm)^2)
sumq5 = total((devs5/ampm)^2)
sumq6 = total((devs6/ampm)^2)
sumq7 = total((devs7/ampm)^2)
sumq=[sumq0,sumq1,sumq2,sumq3,sumq4,sumq5,sumq6,sumq7]
plot, sumq

Rec KL_v7 1lambda/D

Dist Base Mod Bin Type Fr Cyc #mod AmpFile Note
20101024_014947 KL_v7 1 1 Push-pull 3 1 500 optimized_amano_1l_scalettato.fits  

IM Pup Dist It Freq Bin Mod CL Rec gain notes
20101024_015942 20100525_171742 20101024_014947 48 600 1 1 SI 20101023_030820 gain_0.05.fits clock 123.5

REC Base IM bin mod #modes TTrem cutted notes
20101024_021630 KL_v7 20101024_015942 1 1 500 ON 0 clock 123.5

Un po' di casini a chiudere per via di decentraggi di pupilla, ma poi va bene

TrackNo RefStar Mag El Wind Rec bin #mod freq gain mod nph SR band exp #frames disturb notes
20101024_023707 ??? 6.7 90 0 021630 1 500 1000 0.4 0.6 1 1238 78.5 H 20 500 OFF  
20101024_023801 ??? 6.7 90 0 021630 1 500 1000 0.4 0.6 1 1239 81.5 H 20 500 OFF  
20101024_023848 ??? 6.7 90 0 021630 1 500 1000 0.4 0.6 1 1241 81.8 H 20 500 OFF  
20101024_024357 ??? 6.7 90 0 021630 1 500 1000 0.6 0.7 1 1280 76.0 H 20 500 ON  
20101024_024444 ??? 6.7 90 0 021630 1 500 1000 0.6 0.7 1 1278 78.1 H 20 500 ON  
20101024_024531 ??? 6.7 90 0 021630 1 500 1000 0.6 0.7 1 1279 78.4 H 20 500 ON  

Rec KL_v7 0.5lambda/D

Stesso disturbo del 1 lambda/D

IM Pup Dist It FreqSorted ascending Bin Mod CL Rec gain notes
20101024_041112 20100525_171742 20101024_014947 48 600 1 0.5 SI 20101024_021630 gain_0.05.fits clock 123.5

REC Base IM bin mod #modes TTrem cutted notes
20101024_042609 KL_v7 20101024_041112 1 0.5 500 ON 0 clock 123.5

Test rumore

Board setup "Pup_check_internal" con luce direttamente nel W saltando il retroreflector. Le pupille sono troppo grosse e il loop di cameralens non riesce a centrarle, si centrano a mano

Tracknum Bin Photons
  1 1000
  1 5000
  1 10000
  2 1000
  2 5000
  2 10000
  3 1000
  3 5000
  3 10000

Stella AO_04P40M08_0 R=6.8 H=4.7 elev=78°

Il rec a 500 modi ha dei problemi con la posizione delle pupille: quando si va on target inizia a fare skip frame. C'è qualche problema con l'acq pupil perché le pupille sono un po' alte. Mettiamo il 400 mdodi 214405 che sembra funzionare bene.

Autofocus IRTC

Fuoco un po' di difficile da trovare, oscilla tra 50.0 e 50.8 e il picco è un po' introvabile. Alla fine ci fermiamo a 52.35

TrackNo RefStar Mag El Wind DIMM Rec bin #mod freq gain mod nph SR band exp #frames disturb notes
20101024_091445 AO_04P40M08_0 7.5 79 11 -1.00 214405 1 399 1000 0.8 0.8 3 597 86.1 H 5 11 OFF  
20101024_092349 AO_04P40M08_0 7.4 79 11 -1.00 214405 1 399 1000 0.8 0.8 3 619 78.9 H 5 1000 OFF  
20101024_092519 AO_04P40M08_0 7.5 79 11 -1.00 214405 1 399 1000 0.8 0.8 3 609 63.0 H 5 1000 OFF  
20101024_092649 AO_04P40M08_0 7.5 79 11 -1.00 214405 1 399 1000 0.8 0.8 3 604 54.9 H 5 1000 OFF  
20101024_093140 AO_04P40M08_0 7.5 79 10 1.02 214405 1 399 1000 0.8 0.8 3 604 64.8 H 5 1000 OFF  
20101024_093431 AO_04P40M08_0 7.4 79 11 1.35 214405 1 399 1000 0.8 0.8 3 615 81.1 H 5 1000 OFF  
20101024_093639 AO_04P40M08_0 7.5 78 8 -1.00 214405 1 399 1000 0.8 0.8 3 612 76.1 H 5 1000 OFF  
20101024_093819 AO_04P40M08_0 7.5 78 8 -1.00 214405 1 399 1000 0.8 0.8 3 599 65.0 H 5 1000 OFF  
20101024_094241 AO_04P40M08_0 7.5 78 7 1.18 214405 1 399 1000 0.7 0.8 3 575 55.0 H 10 1000 OFF Selected IRTC dark +611h old!
20101024_094428 AO_04P40M08_0 7.5 78 6 1.12 214405 1 399 1000 0.7 0.8 3 602 72.7 H 10 1000 OFF Selected IRTC dark +611h old!
20101024_094601 AO_04P40M08_0 7.5 78 7 1.18 214405 1 399 1000 0.7 0.8 3 613 94.4 H 10 1000 OFF Selected IRTC dark +611h old!
20101024_095037 AO_04P40M08_0 7.5 77 6 1.17 214405 1 399 1000 0.7 0.8 3 609 77.9 H 5 1000 OFF  
20101024_095225 AO_04P40M08_0 7.5 77 7 1.03 214405 1 399 1000 0.7 0.8 3 611 80.6 H 5 1000 OFF  
20101024_095431 AO_04P40M08_0 7.5 77 7 1.16 214405 1 399 1000 0.7 0.8 3 588 56.7 H 5 1000 OFF  
20101024_095610 AO_04P40M08_0 7.5 77 5 1.26 214405 1 399 1000 0.7 0.8 3 604 78.8 H 5 1000 OFF  
20101024_095752 AO_04P40M08_0 7.5 77 7 1.34 214405 1 399 1000 0.7 0.8 3 597 50.9 H 5 1000 OFF  
20101024_095941 AO_04P40M08_0 7.5 76 6 1.24 214405 1 399 1000 0.7 0.8 3 591 37.0 H 5 1000 OFF  
20101024_100117 AO_04P40M08_0 7.5 76 10 1.02 214405 1 399 1000 0.7 0.8 3 599 74.8 H 5 1000 OFF  
20101024_102110 AO_04P40M08_0 8.1 74 11 1.79 092604 1 400 1000 0.7 0.8 3 347 1.9 H 50 379 OFF Selected IRTC dark +613h old!
20101024_104407 AO_04P40M08_0 7.6 70 11 1.26 214405 1 399 1000 0.0 0.0 3 526 0.5 H 50 200 OFF Selected IRTC dark +613h old!
20101024_104456 AO_04P40M08_0 7.8 70 8 1.28 214405 1 399 1000 0.0 0.0 3 456 0.4 H 50 500 OFF Selected IRTC dark +613h old!
20101024_121939 AO_04P40M08_0 10.7 78 7 1.27 071604 2 153 625 0.5 0.8 3 215 25.4 H 200 50 OFF  
20101024_122012 AO_04P40M08_0 10.7 78 8 1.27 071604 2 153 625 0.5 0.8 3 214 17.5 H 200 50 OFF  
20101024_122030 AO_04P40M08_0 10.7 78 7 1.25 071604 2 153 625 0.5 0.8 3 212 18.4 H 200 50 OFF  
20101024_122623 AO_04P40M08_0 10.7 79 9 1.15 071604 2 153 625 0.5 0.8 3 216 20.8 H 200 50 OFF  
20101024_122652 AO_04P40M08_0 10.7 79 9 1.17 071604 2 153 625 0.5 0.8 3 209 13.3 H 200 50 OFF  
20101024_123139 AO_04P40M08_0 10.7 79 7 1.02 071604 2 153 625 0.7 0.7 3 210 16.9 H 200 50 OFF  
20101024_123543 AO_04P40M08_0 10.7 79 8 1.22 071604 2 153 801 0.7 0.7 3 169 29.4 H 200 50 OFF  
20101024_123602 AO_04P40M08_0 10.7 79 7 1.21 071604 2 153 801 0.7 0.7 3 168 29.4 H 200 50 OFF  
20101024_123619 AO_04P40M08_0 10.7 79 9 1.11 071604 2 153 801 0.7 0.7 3 169 32.0 H 200 50 OFF  
20101024_123700 AO_04P40M08_0 10.7 80 9 1.05 071604 2 153 801 0.7 0.7 3 168 30.0 H 200 50 OFF  
20101024_123721 AO_04P40M08_0 10.7 80 9 1.09 071604 2 153 801 0.7 0.7 3 166 21.8 H 200 50 OFF  
Topic revision: r7 - 11 Sep 2012, EnricoPinna
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback